Literature DB >> 23582763

Reliability and validity of an automated computerized visual acuity and stereoacuity test in children using an interactive video game.

Dae Joong Ma1, Hee Kyung Yang, Jeong-Min Hwang.   

Abstract

PURPOSE: To evaluate the test-retest reliability and validity of the new automated computerized distance visual acuity and stereoacuity test for children, which uses an interactive video game.
DESIGN: Retrospective, observational case series.
METHODS: A total of 102 children aged between 3 and 7 years underwent the Snellen visual acuity test, the Distance Randot Stereotest, and the new automated computerized distance visual acuity and stereoacuity test. The test-retest reliability and validity of the automated computerized tests were assessed and compared with the Snellen visual acuity test and the Distance Randot Stereotest with frequency distributions of the differences, Bland-Altman plots, and Deming regression.
RESULTS: The automated computerized distance visual acuity test had high test-retest reliability (95% limits of agreement ±0.18 logMAR, 90.0% of the differences within 0.2 logMAR) and acceptable validity as compared with the Snellen visual acuity chart (95% limits of agreement ±0.27 logMAR, 81.3% of the differences within 0.2 logMAR). The automated computerized distance stereoacuity test had high test-retest reliability (95% limits of agreement ±0.29 log arc second, 95.1% of the differences within 0.3 log arc second) and acceptable validity as compared with the Distance Randot Stereotest (95% limits of agreement ±0.35 log arc second, 93.9% of the differences within 0.3 log arc second).
CONCLUSIONS: The new automated computerized distance visual acuity and stereoacuity test, which uses an interactive video game, has good reliability and acceptable validity compared with the Snellen visual acuity chart and the Distance Randot Stereotest.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23582763     DOI: 10.1016/j.ajo.2013.02.018

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  9 in total

1.  A new visual acuity test on touchpad for vision screening in children.

Authors:  Elsa Di Foggia; Noémie Stoll; Hélène Meunier; Adam Rimelé; Pascal Ance; Pierre-Henri Moreau; Claude Speeg-Schatz; Arnaud Sauer
Journal:  Int J Ophthalmol       Date:  2020-09-18       Impact factor: 1.779

2.  Effect of chromatic contrast on stereoacuity measurement with computer-aided three-dimensional technology.

Authors:  Lingzhi Zhao; Huang Wu
Journal:  Ann Transl Med       Date:  2019-05

3.  The difference in stereoacuity testing: contour-based and random dot-based graphs at far and near distances.

Authors:  Lingzhi Zhao; Huang Wu
Journal:  Ann Transl Med       Date:  2019-05

4.  ASTEROID: A New Clinical Stereotest on an Autostereo 3D Tablet.

Authors:  Kathleen Vancleef; Ignacio Serrano-Pedraza; Craig Sharp; Gareth Slack; Carla Black; Therese Casanova; Jess Hugill; Sheima Rafiq; James Burridge; Vito Puyat; Josee Ewane Enongue; Henry Gale; Hannah Akotei; Zoe Collier; Helen Haggerty; Kathryn Smart; Christine Powell; Kate Taylor; Michael P Clarke; Graham Morgan; Jenny C A Read
Journal:  Transl Vis Sci Technol       Date:  2019-02-28       Impact factor: 3.283

5.  Effect of Luminance and Contrast Variation on Stereoacuity Measurements Using Smartphone Technology.

Authors:  Lu Liu; Lingxian Xu; Junyue Wang; Huang Wu
Journal:  J Ophthalmol       Date:  2021-12-23       Impact factor: 1.909

6.  Development and validation of a new method for visual acuity assesment on tablet in pediatric population: eMOVA test.

Authors:  Noémie Stoll; Elsa Di Foggia; Claude Speeg-Schatz; Hélène Meunier; Adam Rimele; Pascal Ancé; Pierre-Henri Moreau; Arnaud Sauer
Journal:  BMC Ophthalmol       Date:  2022-04-19       Impact factor: 2.086

7.  Distance stereotesting using vision test charts for intermittent exotropia.

Authors:  Noriko Nishikawa; Satoshi Ishiko; Ikuko Yamaga; Miho Sato; Akitoshi Yoshida
Journal:  Clin Ophthalmol       Date:  2015-08-25

8.  Evaluating stereoacuity with 3D shutter glasses technology.

Authors:  Huang Wu; Han Jin; Ying Sun; Yang Wang; Min Ge; Yang Chen; Yunfeng Chi
Journal:  BMC Ophthalmol       Date:  2016-04-26       Impact factor: 2.209

9.  Screening for Stereopsis Using an Eye-Tracking Glasses-Free Display in Adults: A Pilot Study.

Authors:  Fang Liu; Jing Zhao; Tian Han; Yang Shen; Meng Li; Jingrong Liu; Dong Yang; Yong Fang; Li Yan; Xingtao Zhou
Journal:  Front Med (Lausanne)       Date:  2022-01-17
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.