Microfluidic glucose biosensors and potassium ion selective electrodes were used in an in vivo study to measure the neurochemical effects of spreading depolarizations (SD), which have been shown to be detrimental to the injured human brain. A microdialysis probe implanted in the cortex of rats was connected to a microfluidic PDMS chip containing the sensors. The dialysate was also analyzed using our gold standard, rapid sampling microdialysis (rsMD). The glucose biosensor performance was validated against rsMD with excellent results. The glucose biosensors successfully monitored concentration changes, in response to SD wave induction, in the range of 10-400 μM with a second time-resolution. The data show that during a SD wave, there is a time delay of 62 ± 24.8 s (n = 4) between the onset of the increase in potassium and the decrease in glucose. This delay can be for the first time demonstrated, thanks to the high-temporal resolution of the microfluidic sensors sampling from a single tissue site (the microdialysis probe), and it indicates that the decrease in glucose is due to the high demand of energy required for repolarization.
Microfluidic glucose biosensors and potassium ion selective electrodes were used in an in vivo study to measure the neurochemical effects of spreading depolarizations (SD), which have been shown to be detrimental to the injured human brain. A microdialysis probe implanted in the cortex of rats was connected to a microfluidic PDMS chip containing the sensors. The dialysate was also analyzed using our gold standard, rapid sampling microdialysis (rsMD). The glucose biosensor performance was validated against rsMD with excellent results. The glucose biosensors successfully monitored concentration changes, in response to SD wave induction, in the range of 10-400 μM with a second time-resolution. The data show that during a SD wave, there is a time delay of 62 ± 24.8 s (n = 4) between the onset of the increase in potassium and the decrease in glucose. This delay can be for the first time demonstrated, thanks to the high-temporal resolution of the microfluidic sensors sampling from a single tissue site (the microdialysis probe), and it indicates that the decrease in glucose is due to the high demand of energy required for repolarization.
Traumatic
brain injury (TBI) has been described as a silent epidemic by an editorial
in Lancet Neurology journal.[1] This stark judgment is based upon facts such as the annual United
States incidence of TBI (390–445/100 000 population)
being higher than that of breast cancer (120/100 000 population),[2] and an estimated direct and indirect cost of
$60 billion in 2000.[3]TBI may take
many forms including intracerebral hemorrhage, subarachnoid hemorrhage,
cerebral contusion, and axonal injury. The primary injury refers to
the damage of the initial trauma and often occurs before the paramedics
arrive at the scene. In some cases, this is a small core that becomes
ischemic due to loss of blood flow and is surrounded by a “penumbra”
of tissue where local blood flow is substantially reduced, compromising
neuronal function but not viability. Approximately 40% of TBI patients
deteriorate in the days after the initial injury,[4] due to the onset of secondary injury, at a time when the
patient is being monitored in the intensive therapy unit (ITU). Secondary
injuries can be caused by ischemia, cerebral hypoxia, cytotoxic cerebral
edema, raised intracranial pressure, neurovascular edema, and spreading
depolarization waves (SDs).The spontaneous occurrence of SDs
in the injured human brain were first observed by the Strong/Boutelle
collaboration.[5] The mass depolarization
of cells, first described by Leao in 1944,[6] propagates like a wave over the cortex, leading to temporary failure
of ion homeostasis with a rise in extracellular potassium and intracellular
sodium and calcium and disruption of cortical function. The depression
of the electrical signal coincides with a transient increase in blood
flow and changes in metabolism due to the high demand of energy needed
to repolarize the cells.[7] SDs have been
shown to occur in 56% of TBI patients, often in stereotyped clusters[8] and with a high indicidence in other types of
brain injury.[9] They have been associated
with ischemic damage[10] and the onset of
neurological deficits.[11] A study by Murray
et al. showed that the commonly used clinical covariates, such as
age, pupil reactivity, level of hypoxia, and motor score, only explain
30% of the variance in TBI outcome.[12] Results
obtained by the Co-Operative Study on Brain Injury Depolarisations
(COSBID) group, show that the occurrence of SDs independently predicts
poor patient outcome, and is a better indicator than any of the currently
used factors.[13]In experimental models,
SD waves can be detected electrophysiologically or real-time imaging
of blood flow can be used to map the path of SD waves. High temporal
resolution is required as the SD wave takes a few minutes to pass
any particular point. Real time measurement of cerebral blood flow
(CBF) using laser speckle contrast imaging (LSI) has been used to
study SD propagation under both normal[14,15] and ischemic[16−18] conditions. It has been shown to be a good tracker of SD waves[16,19] and is a semiquantitative measure of dynamic CBF.[20] LSI has high spatial and temporal resolution (30 μm
and 1 s, respectively), providing a high-resolution map of the evolution
of cerebral blood flow.[19] The number of
depolarizations and the clustering of these events are associated
with the growth of the infarct area[21−23] due to ischemic damage
in the penumbra.[16,17] It has also been shown that SD
waves may cycle around the site of injury, possibly explaining the
repetition at regular intervals seen also in some patients.[15]Using MRI and PET imaging studies in patients,
a UCLA group has shown that during a persistent metabolic crisis,
the lactate/pyruvate ratio increases, and this has been shown to be
indicative of tissue atrophy at 6 months with oxidative brain metabolism
playing a central role.[24,25] Studies of TBI patients
cortical glucose levels using classical microdialysis with hourly
samples, have indicated that depletion of brain glucose carries a
poor prognosis.[26] However, this temporal
resolution is insufficient to detect the frequently occurring dynamic
changes that challenge the tissue. The metabolic signature of a transient
ischemia has been detected using implanted biosensors[27] and online rapid sampling microdialysis (rsMD),[28] as a decrease in glucose levels and increase
in lactate levels, typically lasting 2–5 min.[29] Using rsMD in animal studies, such a signature was also
found as an SD wave passes a MD probe,[30] leading to an extended reduction in brain glucose. In human studies,
rsMD could initially only show a correlation between the number of
SD waves and the fall in brain glucose.[31] Improvements in sensitivity of the assay and the development of
algorithms to reduce noise[32] have confirmed
the chemical signature of SD waves in patients and shown that repetition
of the events can drive down the local glucose concentration, to a
level where the tissue is no longer viable.[29] However, with the 1 min temporal resolution coupled to the small
(20–50 μM), rapid decreases in brain glucose, the temporal
resolution of rsMD is at its limit. To study the temporal profile
of SDs further, two important challenges remain: to know the start
of the SD event in the microdialysis data and to detect the metabolism
changes with high temporal resolution. We have addressed the first
with an online potassium ion selective electrode (ISE)[33] and the second with an online glucose biosensor.
For increased time-resolution, the novel glucose biosensor together
with the potassium ISE were integrated in a microfluidic platform
designed to monitor the neurochemical effects of SD waves.
Results and Discussion
Biosensor Characterization
The glucose biosensors have an average sensitivity of 3.9 ±
0.6 × 103 nA/mM/cm–2 (n = 15) and a T90% of 3.1 ± 1.3 s
(n = 15). All results are presented as mean ±
standard deviation. The main sources of noise are interference from
the magnetic stirrer and the injection of known glucose concentration
used to assess the sensor performance during calibrations. The inherently
fast time responses of these sensors allow for measurements with a
temporal resolution on the order of seconds.Although the sensor
is held at a 0.7 V (vs Ag|AgCl) throughout the experiment, interference
from other electroactive species, namely, ascorbic acid and dopamine,
is not an issue due to the protection provided from the poly(phenol)
film used. Using a similar needle electrode, the interference through
this film has been previously assessed as negligible for an ATP sensor
detecting smaller changes.[34]
Biosensor Validation
To ensure that the online biosensors
were capable of detecting biologically relevant concentrations of
glucose in brain dialysate, the results from the glucose biosensor
within the microfluidic chip were compared to the results of rsMD
which analyzed the same dialysate sample. Figure 1A shows that the results of the two analysis techniques overlaid,
after being individually analyzed and time-aligned. This data set
shows levels of glucose recorded in vivo during 2 h. At 60 min, a
blood clot was removed and at 127 min, the animal was euthanized,
and glucose levels fell to near zero. Glucose concentrations are fluctuating
over the course of the experiments, with both monitoring techniques
showing parallel changes, indicating good agreement between the microfluidic
glucose biosensor and rsMD glucose.
Figure 1
Glucose biosensor validation. (A) Comparison
of glucose rsMD to biosensor data over the course of an procedure.
Data was individually time-aligned and analyzed using calibrations
of sensors before being overlaid. The levels vary over the duration
of the procedure and finally fall to near zero after euthanasia at
107 min. (B) Comparison of glucose biosensor and rsMD data. (C) The
mean difference in data measurement as a function of the rsMD glucose
value. From these data, on average across the concentration range,
the concentration determined by the biosensor was marginally, but
significantly lower than that determined by rsMD (−16.6 ±
44 μM, p = 0.0001, n = 356).
Glucose biosensor validation. (A) Comparison
of glucose rsMD to biosensor data over the course of an procedure.
Data was individually time-aligned and analyzed using calibrations
of sensors before being overlaid. The levels vary over the duration
of the procedure and finally fall to near zero after euthanasia at
107 min. (B) Comparison of glucose biosensor and rsMD data. (C) The
mean difference in data measurement as a function of the rsMD glucose
value. From these data, on average across the concentration range,
the concentration determined by the biosensor was marginally, but
significantly lower than that determined by rsMD (−16.6 ±
44 μM, p = 0.0001, n = 356).Figure 1B shows the comparison of three online biosensors to the equivalent
rsMD data. If the glucose measurement was that of blood, it would
be appropriate to use a Clarke error grid.[35] However, for continuous glucose monitoring, the relative timing
of paired results during a dynamic change is very important.[36] As both monitoring systems used in the current
study analyze the same dialysate stream from the same extracellular
tissue space, the mean difference between the two techniques provides
a better estimation of sensor performance. This is shown in Figure 1C. On average, there is a slight bias for the biosensor
to report lower values compared to that of rsMD (−16.6 ±
44 μM, p = 0.0001, n = 356).
However, analysis shows that the least squared regression slope is
not significantly different from zero.
Multimodal
Approach to SD
A multimodal approach was taken to investigate
spreading depolarizations and fully test the microfluidic platform.
All the recordings were individually calibrated and time aligned to
t0, the time when the SD wave reaches the MD probe. Figure 2 shows the experimental setup used. The t0, t1, and t2 values were determined experimentally as explained
in the Methods section.
Figure 2
Translational setup.
Not to scale. t0 is the time the event
hits the MD membrane. t1 is the time delay
before the dialysate, representative of the event, reaches the sensors
(on average 4 min). t2 is the dialysate
transfer time from the sensors to rsMD (on average 15 min). A potassium
ISE and a glucose biosensor are placed within a chamber of 68 nL volume.
Dialysate is passed to rsMD for flow injection analysis of glucose
and lactate.
Translational setup.
Not to scale. t0 is the time the event
hits the MD membrane. t1 is the time delay
before the dialysate, representative of the event, reaches the sensors
(on average 4 min). t2 is the dialysate
transfer time from the sensors to rsMD (on average 15 min). A potassium
ISE and a glucose biosensor are placed within a chamber of 68 nL volume.
Dialysate is passed to rsMD for flow injection analysis of glucose
and lactate.An overview of all the
measurements taken is shown in Figure 3. Successful
induction of the SD wave was confirmed by a DC negative shift and
a transient suppression of the ECoG activity (Figure 3B), which are characteristic of cellular depolarization. The
propagation of the SD wave could be tracked by LSI; the passage of
a hyperemic wave at a speed of 4.4 mm/min passed the parenchymal microelectrode,
and the MD probe was verified (Figure 3A).
By using such a multimodal approach, we can be sure that changes in
concentrations given by the microfluidic sensors are a response to
a true event taking place in the brain tissue sampled by the MD probe,
which gives confidence in the interpretation of the online MD sensor
data.
Figure 3
Multimodal response to an SD wave. (A) Blood flow at three different
ROIs in progressing distance (red, green, black) from the site of
the needle prick. (B) DC potential (light blue) and ECoG (dark blue).
(C) Microdialysate potassium. (D) Glucose readings from rsMD (red)
and dialysate biosensor (yellow). (E) rsMD lactate.
Multimodal response to an SD wave. (A) Blood flow at three different
ROIs in progressing distance (red, green, black) from the site of
the needle prick. (B) DC potential (light blue) and ECoG (dark blue).
(C) Microdialysate potassium. (D) Glucose readings from rsMD (red)
and dialysate biosensor (yellow). (E) rsMD lactate.The on-chip MD potassium ISE response is shown
in Figure 3C. An increase in extracellular
potassium is typical for an SD wave and can be here observed in the
microdialysate, given the limitations of microdialysis sampling and
dispersion. A detailed study on MD online potassium measurement can
be found in ref (33). The glucose levels, as recorded by both the microfluidic biosensor
and rsMD, are compared in Figure 3D. The concentrations
correlate very well before and during the SD. After the SD, the rsMD
trace is slightly higher, though the difference is within the error
calculated for Figure 1C. During the SD, the
concentration of glucose decreases by 12 μM in both data sets.
The rsMD lactate data is shown in Figure 3E,
where levels increase as the SD wave passes, as observed previously
in experimental and clinical monitoring of SD with rsMD.[14,29]
Microfluidic Sensor Data: Neurometabolic Coupling
To better appreciate the advantage of the finer time-resolution
of the glucose biosensor (as opposed to the minute time-resolution
of the rsMD), the responses of the microfluidic sensors and of the
near real-time CBF measurements are shown in Figure 4. Four sections of data are shown, each being the response
to a different needle prick, from two rats, one under propofol (Figure 4B–D) and the other under isoflurane (Figure 4E). In all cases, blood flow, on-chip potassium
and glucose biosensor responses have been time-aligned (see Methods section). A propagating hyperaemia is seen
in the laser speckle data with each needle prick, indicating a SD
wave passing through the area where the MD probe was implanted at
a speed of 4.4 mm/min, typical of first SD wave speeds.[37] This indicates that the presence of the MD probe
did not alter the propagation direction or speed of the SD wave. Furthermore,
the hyperaemic peak coincides with the onset of increase in extracellular
potassium, as detected by the microfluidic ISE. Once the hyperemia
has passed, the potassium levels begin to fall to near baseline, indicating
repolarization of the neuronal cells. In each case the glucose biosensors
show a transient decrease in the local concentration. This is true
for all four NP data sets as seen in Figure 4. Interestingly, in our multi-needle-prick model, brain glucose and
CBF had all recovered to their baseline value before the changes induced
by the SD wave (baseline [glucose] ≈ 0.47 mM under propofol,
0.105 mM under isoflurane, and CBF ≈ 50% under propofol and
100% under isoflurane).
Figure 4
Microfluidic sensors. Panel (A) is a laser speckle
image showing the left hemisphere through a thinned skull. The sites
for needle prick stimuli are indicated (NP1,2,3) and the positions
of the intraparenchymal microelectrode (DC) and microdialysis probe
(MD) are also marked. Three circular ROIs were used to analyze the
blood flow response around the implanted measurement sites. The color
of the circle indicates the trace seen in the blood flow graphs. The
dotted line indicates the wavefront of the SD wave elicited by the
second needle prick at the time of the still. Panels (B–D)
show the response of the microfluidic sensors of needle pricks conducted
in the same rat under propofol (NP1,2,3). Panel (E) shows the responses
from a different rat under isoflurane (NP1). The gray dotted lines
in (B–E) show the moment the SD wave hits the MD probe as indicated
by the change in potassium.
Microfluidic sensors. Panel (A) is a laser speckle
image showing the left hemisphere through a thinned skull. The sites
for needle prick stimuli are indicated (NP1,2,3) and the positions
of the intraparenchymal microelectrode (DC) and microdialysis probe
(MD) are also marked. Three circular ROIs were used to analyze the
blood flow response around the implanted measurement sites. The color
of the circle indicates the trace seen in the blood flow graphs. The
dotted line indicates the wavefront of the SD wave elicited by the
second needle prick at the time of the still. Panels (B–D)
show the response of the microfluidic sensors of needle pricks conducted
in the same rat under propofol (NP1,2,3). Panel (E) shows the responses
from a different rat under isoflurane (NP1). The gray dotted lines
in (B–E) show the moment the SD wave hits the MD probe as indicated
by the change in potassium.Given microdialysis recovery (estimated in vitro as 32–33%),
these baseline values are in accordance with interstitial glucose
concentrations found in the literature to be in the order of 2–3
mM in anaesthetised animals.[28] Although
the higher cerebral blood flow under isoflurane would lower the in
vivo recovery of a species not present in the blood,[38] for glucose this would not be the case as the blood flow
is the source of brain glucose. The extracellular level of glucose
is a balance between this source and the local utilization. Our lower
baseline dialysate glucose concentration under isoflurane compared
to propofol is mainly due to the anesthetic, since propofol increases
brain glucose by 50% compared to isoflurane,[39] and partly due to a slight difference in the in vivo recovery of
the MD probe. Given the small sample of animals used in this study,
we cannot conclude on a significant quantitative effect of anesthesia
on brain glucose concentrations. The temporal relationship between
microdialysis potassium and glucose is however not influenced by the
anesthetics.In one animal, after the second needle prick, there
was a transient drop in blood pressure (down to 48 mmHg), which triggered
the spontaneous propagation of a second SD wave 6 min after the induced
SD (shown in Figure 5). Here, in contrast to
the SD initiated via the needle prick, the glucose concentration decreases
immediately as the SD passes the MD probe.
Figure 5
Spontaneous SD wave following
a spontaneous drop in blood pressure. Top graph shows the blood flow;
bottom graph shows the online glucose (orange) and potassium (purple)
data. The first SD was induced from a needle prick. The second SD
wave is a spontaneous wave, likely to be caused by a drop in blood
pressure (gray line indicates when the blood pressure was at its lowest).
Spontaneous SD wave following
a spontaneous drop in blood pressure. Top graph shows the blood flow;
bottom graph shows the online glucose (orange) and potassium (purple)
data. The first SD was induced from a needle prick. The second SD
wave is a spontaneous wave, likely to be caused by a drop in blood
pressure (gray line indicates when the blood pressure was at its lowest).The temporal coupling between
membrane cell repolarization, cerebral blood flow, and glucose utilization
can be seen for the first time thanks to our subminute temporal resolution.
The microfluidic data from the four needle pricks was indeed analyzed
more closely by looking at just the microdialysate potassium and glucose
levels, as seen in Figure 6. Allocating the
point of change in potassium as t = 0, all potassium
and glucose concentration changes were aligned. The background level
of each is also removed and the relative changes (Δ[K+] and Δ[glucose]) and timings can be closer assessed. In the
top graph, all of the potassium changes are overlaid. Clearly, each
SD wave causes a transient increase in MD potassium concentrations,
characteristic of transient extracellular potassium changes observed
with in situ electrodes.[33,40] In the bottom graph
the glucose concentration changes are shown for the same four needle
pricks. The changes are all small and all negative from the baseline.
The magnitude of the changes in glucose seen here are of similar magnitude
to the changes seen in patient data using rsMD.[29] Interestingly, all four needle pricks show some delay in
the decrease of local concentration compared to the start of the response
of potassium. Note that the glucose biosensor was always positioned
upstream of the potassium sensor. In calibrations, the biosensor responded
to the change in concentrations first (data not shown) and this slight
delay was taken into account when time-aligning the two sensors. Therefore
we can rule out this delay being due to mass transport effects. There
is thus a 62 ± 24.8 s (n = 4) delay between
the onset of the potassium and onset of the glucose responses, which
is of physiological origin. This time delay is too short to be seen
using the one minute sampling of rsMD. The decrease in glucose actually
coincides with the reversal of the potassium trend, as seen in Figure 6. Re-establishment of the ionic gradients after
depolarization requires indeed much energy to power the pumps to actively
transport ions across the cell membranes, and therefore the local
demand for glucose increases.
Figure 6
Assessment of glucose and potassium changes.
Time t = 0 is the point where MD potassium levels
start to increase. Potassium and glucose levels were time aligned
to each other using calibration data (see Methods section). Background levels of both potassium and glucose have been
removed. (i)–(iii) are traces (B)–(D) of Figure 5, respectively, and were conducted under propofol.
(iv) is trace (E) of Figure 5 where isoflurane
was the anesthesia.
Assessment of glucose and potassium changes.
Time t = 0 is the point where MD potassium levels
start to increase. Potassium and glucose levels were time aligned
to each other using calibration data (see Methods section). Background levels of both potassium and glucose have been
removed. (i)–(iii) are traces (B)–(D) of Figure 5, respectively, and were conducted under propofol.
(iv) is trace (E) of Figure 5 where isoflurane
was the anesthesia.This coupling between
restoration of ion homeostasis and increased metabolic demand can
be for the first time unambiguously shown thanks to the very high-temporal
resolution of both the ISE and the glucose biosensor, which analyze
the same dialysate sample. Remarkably, despite a clear hyperemia associated
with each SD wave (see Figure 4B–E),
which presumably increases the supply of glucose and oxygen from the
blood, extracellular glucose drops. This has been previously observed
in cats using rsMD.[14] It can therefore
be concluded that the drop in cerebral glucose, observed here in rats,
and elsewhere in cats and in brain injurypatients, is the consequence of the increased energy demand for cell repolarization,
demand which is not met by local supply from the blood.
Conclusion
This study has shown the development of
microfluidic glucose biosensors to be instrumental in the development
of a clinical monitoring system. In this study, we used an in vivo
needle prick model, as a translational replica of a “mini”
traumatic brain injury to assess the performance of the new online
microdialysis biosensors and investigate neuro-metabolic coupling
in response to an SD wave.The glucose biosensor was validated
against rsMD glucose with excellent results. A potassium sensor was
additionally placed on-chip within the analysis chamber to analyze
the sample of dialysate. Thanks to the fast time responses and continuous
analysis of microdialysate streams, this study has allowed an unambiguous
insight into the timing of the chemical events that occur during an
SD wave, a distinct advantage over the currently used flow injection
system. It was indeed found that the decrease in glucose only occurred
as the cells started repolarizing, as indicated by the coincidence
of the onset of the glucose fall with the onset of the recovery of
potassium. It was concluded that when the cells begin to repolarize
after the passage of an SD wave, the supply of metabolic substrates
by the blood does not match the increased demand for repolarization
and so the local glucose concentration is seen to fall. On-going optimization
of the microfluidic manifold and sensors[41] for clinical use will drive this work forward. The small sample
size for the animal data in the current study restricts our ability
to analyze the mechanism of changes in detail. However, the findings
presented here highlight the potential that the microfluidic system
holds for the monitoring of brain injurypatients by allowing the
analysis of multiple analytes in real-time within the same low-volume
dialysate sample. With the very high temporal resolution this provides,
the cause/consequence questions can begin to be answered. More widely,
this technology could allow the detection and study of transient ischemia
or functional activation.
Methods
Electrode Fabrication
The glucose biosensors in this
paper are based upon the combined needle electrodes,[34] a photograph of which is shown in Figure 7A. Briefly, 50 μm Teflon insulated platinum wire (A-M
Systems Inc.) and 50 μm polyester insulated silver wire (AM
systems) were threaded through a 27G hypodermic needle. The wires
were stripped of their insulator using a lighter to expose the metal
wire. Electrical wire was glued using conductive silver epoxy glue
(RS Components). Epoxy resin (Robnor resins, CY1301 and HY1300) was
used to fill the internal volume of the needle and secure the wires
in place. Once the epoxy had cured, the sharp end of the needle was
cut using a diamond saw (Buehler) to expose the microelectrodes. The
ends were polished using alumina slurries finishing with 0.05 μm.
The silver disc was chloridized by placing in Referencing Solution
(BAS) for 5 s, to create Ag|AgCl reference electrode. Cyclic voltammetry
was used to assess the surface of the electrode. The working electrode
consisted of a 50 μm diameter platinum disc electrode. A scanning
electron microscopy (SEM) image of the polished needle tip can be
seen in Figure 7B.
Figure 7
Sensor design. (A) Photograph
of combined needle electrode. (B) SEM image of the needle tip. The
working electrode is a 50 μm platinum/iridium disc; the reference
electrode is formed by chloridizing the 50 μm silver disc to
get Ag|AgCl; the counter electrode is the stainless steel shaft of
the hypodermic needle. (C) Raw data calibration using a 50 μm
glucose biosensor. Biosensor is held at a constant potential of 0.75
V. Aliquots of glucose standard are added and current response recorded.
Purple arrow indicates the change in glucose concentration from 0
to 100 μM. (D) Calibration indicating current against glucose
concentration using a 50 μm glucose biosensor. Data is fitted
with the Hill equation giving an apparent Km value of 11.35 ± 0.075 mM.
Sensor design. (A) Photograph
of combined needle electrode. (B) SEM image of the needle tip. The
working electrode is a 50 μm platinum/iridium disc; the reference
electrode is formed by chloridizing the 50 μm silver disc to
get Ag|AgCl; the counter electrode is the stainless steel shaft of
the hypodermic needle. (C) Raw data calibration using a 50 μm
glucose biosensor. Biosensor is held at a constant potential of 0.75
V. Aliquots of glucose standard are added and current response recorded.
Purple arrow indicates the change in glucose concentration from 0
to 100 μM. (D) Calibration indicating current against glucose
concentration using a 50 μm glucose biosensor. Data is fitted
with the Hill equation giving an apparent Km value of 11.35 ± 0.075 mM.
Glucose Biosensor Fabrication
All
biosensors were controlled via a lab built potentiostat feeding into
a Powerlab 16/35 running LabChart Pro (AD Instruments). The combined
electrode was placed in a solution containing 50 mM Phenol and 4 mg/mL
glucose oxidase. The pH of the solution was buffered to pH 7.2. The
electrode was placed in the enzyme solution for 10 min. Then under
potentiostatic control, the working electrode was held at 0 V for
20 s, polarized to 0.9 V for 15 min for electropolymerization, and
then held at 0 V for 20 s. The biosensor was gently rinsed with deionized
water and stored dry overnight at 4 °C before use. The biosensors
were characterized in a beaker containing phosphate buffer solution
(PBS), which was highly stirred. Aliquots of a concentrated glucose
solution were added to the PBS using a glass syringe (Hamilton), to
create known step changes in concentration in the bulk media. Figure 7C shows the raw data of a 100 μM step change
of glucose. The limit of detection, calculated as 3x the standard
deviation of the baseline signal, is 1.5 μM. Figure 7D shows a calibration within the brain physiological
range. The data has been fitted with the Hill equation using IgorPro.
In Vivo Experiments
All animal procedures
were carried out in accordance with the German Laws for Animal Protection
and institutional guidelines. Two male Wistar rats were anesthetized
with isoflurane (5% induction, 1.5–2% maintenance) in a 70%:30%
nitrous oxide/oxygen mixture during all surgical procedures. Rectal
temperature was maintained at 37 °C using a servo-controlled
heating blanket. The left femoral artery was cannulated for continuous
blood pressure recording and hourly arterial blood gases measurements.
The left frontal and parietal bones were exposed and thinned out to
transparency using a dental drill. Drilling was performed under continuous
saline irrigation to prevent heat injury. The cavity formed covered
most of the left cerebral cortex and provided a field of view of ∼10
× 7 mm for LSI. Two small burr-holes were drilled into the frontal
and parietal cortex for subsequent needle pricks at two different
sites (see Figure 4A). Two small durectomies
were drilled at 3.5 mm posterior and 3.5 mm lateral to Bregma and
4 mm posterior and 3.5 mm lateral to Bregma for the parenchymal electrode
and the microdialysis (MD) probe, respectively. The durectomies were
placed in areas devoid of major blood vessels and at the edge of the
field of view so as to limit artifacts for LSI. The distance between
the microelectrode and the MD probe (≈0.5 mm distance) was
as small as possible to ensure simultaneity of the events at both
measurement sites.After all surgical procedures were completed,
LSI was started and continued uninterrupted during the whole duration
of the experiments. The LSI method was implemented as previously described.[15,19] After baseline imaging, the MD probe (MAB 6.14.2, Microbiotech,
Sweden) was implanted using a piezoelectric motor at a rate of 5 μm/s,
while being perfused with aCSF at a rate of 1.6 μL/min. The
probe was implanted obliquely to full membrane length (2 mm), so as
to ensure maximal sampling from cortical tissue. The MD probe outlet
was connected directly to a microfluidic analysis chip containing
first a glucose biosensor and then a potassium ISE.[33] The dialysate stream was then fed into the rsMD valve for
flow injection analysis of glucose and lactate. A schematic of the
procedure setup is shown in Figure 2. A glass
microelectrode (2.5–3 μm tip diameter) was then implanted
at a depth of 500 μm. Direct current (DC) and electrocorticogram
(ECoG) were measured versus a subcutaneous sintered Ag/AgCl wire electrode
that served as a reference. The animal was earthed to the instrumentation
ground.A waiting period of 45 min followed implantations, during
which all measured parameters (arterial blood pressure, CBF, brain
glucose, lactate and potassium, DC, and ECoG) stabilized to normal
physiological values. One animal was maintained under isoflurane (1.5%),
and the other animal was switched to propofol (38 mg/kg/h infusion
via tail vein), an intravenous anesthetic commonly used in ITU for
TBI patients.[42] The effect of propofol
compared to isoflurane on CBF and SD waves has been discussed elsewhere.[43,44] After stabilization of all parameters, a first needle prick was
performed on the frontal cortex. The needle prick model can be understood
as a mechanically induced focal traumatic injury, causing one single
SD wave.[45,46] After 1 h, when CBF had recovered to its
baseline values, a second SD wave was elicited by needle prick on
the parietal cortex. When the animal conditions allowed it, a third
needle prick was performed 1 h later on the frontal cortex, triggering
a single SD wave. In one animal, after the SD wave caused by the second
needle prick, a transient drop in blood pressure was observed (down
to 48 mmHg), which triggered the spontaneous propagation of a second
SD wave (see Figure 5). At the end of the experiment,
the animals were euthanised by intravenous injection of a high concentration
of potassium chloride (3 M KCl).
Data
Analysis
Speckle contrast images were analyzed for CBF by
placing regions of interest (ROIs) of 1 mm diameter in areas devoid
of major blood vessels. Three ROIs were positioned around the implanted
electrode and MD to capture the SD wave as it approaches and passes
the probes (see Figure 4). CBF values are here
reported relative to baseline CBF under isoflurane taken as 100% CBF.
Electrophysiological data from the parenchymal microelectrode were
digitized at 250 Hz and low-passed filter for DC and high -pass filtered
for ECoG (0.1 Hz cutoff frequency in both cases). The online MD data
were analyzed independently and blindly for rsMD glucose and lactate
in Cologne and for the online potassium and glucose sensors in London.
Data were denoized using our previous published algorithms.[32] Before each in vivo experiment, an in vitro
MD recovery experiment was conducted, where the MD probe was moved
between known solutions at known times. This not only checked the
functionality of the MD probe but also allowed the time difference
between the known responses of the different analysis techniques to
be measured. If t0 is the time at which
a change occurred at the MD probe membrane, t1 is the length of time for this change to be detected at the
microfluidic sensors and t2 is the time
to move the dialysate from the sensors to being injected into the
rsMD valve, as seen in Figure 2. Once these
values are known, all the individual data sets can be aligned to the
same time point, t0. All results presented
in the text have been time-aligned between the different analysis
techniques.
Authors: Delphine Feuerstein; Andrew Manning; Parastoo Hashemi; Robin Bhatia; Martin Fabricius; Christos Tolias; Clemens Pahl; Max Ervine; Anthony J Strong; Martyn G Boutelle Journal: J Cereb Blood Flow Metab Date: 2010-02-10 Impact factor: 6.200
Authors: Yueqiao Xu; David L McArthur; Jeffry R Alger; Maria Etchepare; David A Hovda; Thomas C Glenn; Sungcheng Huang; Ivo Dinov; Paul M Vespa Journal: J Cereb Blood Flow Metab Date: 2009-12-23 Impact factor: 6.200
Authors: Jens P Dreier; Sebastian Major; Andrew Manning; Johannes Woitzik; Chistoph Drenckhahn; Jens Steinbrink; Christos Tolias; Ana I Oliveira-Ferreira; Martin Fabricius; Jed A Hartings; Peter Vajkoczy; Martin Lauritzen; Ulrich Dirnagl; Georg Bohner; Anthony J Strong Journal: Brain Date: 2009-05-06 Impact factor: 13.501
Authors: Hajime Nakamura; Anthony J Strong; Christian Dohmen; Oliver W Sakowitz; Stefan Vollmar; Michael Sué; Lutz Kracht; Parastoo Hashemi; Robin Bhatia; Toshiki Yoshimine; Jens P Dreier; Andrew K Dunn; Rudolf Graf Journal: Brain Date: 2010-05-26 Impact factor: 13.501
Authors: Jing Zhang; Andrea Jaquins-Gerstl; Kathryn M Nesbitt; Sarah C Rutan; Adrian C Michael; Stephen G Weber Journal: Anal Chem Date: 2013-09-24 Impact factor: 6.986
Authors: Erika L Varner; Chi Leng Leong; Andrea Jaquins-Gerstl; Kathryn M Nesbitt; Martyn G Boutelle; Adrian C Michael Journal: ACS Chem Neurosci Date: 2017-05-18 Impact factor: 4.418