A clinical comparison between a non-invasive blood pressure monitor using high definition oscillometry (Memodiagnostic MD 15/90 Pro) and invasive arterial blood pressure measurement in anaesthetized dogs.

Maja K Rysnik1, Peter Cripps, Isabelle Iff.   

Abstract

OBJECTIVE: To compare high definition oscillometry (HDO) to invasive blood pressure measurement in anaesthetized dogs. STUDY
DESIGN: Prospective, clinical trial. ANIMALS: Fifty dogs weighing 1.95-79 kg (mean 23.5 kg).
MATERIALS AND METHODS: Anaesthetic and peri-anaesthetic management was chosen according to each dog's physical status and anaesthetist's preference. Direct arterial blood pressure measurements were performed using a catheter placed in the dorsal pedal artery and an electronic pressure transducer connected to a multiparameter monitor. Non-invasive blood pressure measurements were performed using an appropriately sized cuff placed around the tail base. Comparisons between the two methods were made using Bland and Altman plots. The data are reported as mean bias (lower, upper limits of agreement). Further analysis was performed after separating the data into the following categories based on invasive mean arterial blood pressure (MAP): high (MAP > 100 mmHg), medium (70 mmHg < MAP < 100 mmHg) and low (MAP < 70 mmHg) blood pressure (BP). The two methods were compared as used clinically.
RESULTS: Eight hundred measurement pairs for invasive and HDO BP readings were compared. Overall, the HDO measured lower values for SAP and DAP but higher for MAP than the invasive method. The lowest bias (upper, lower limits of agreement) were obtained for MAP, -1 (-22, 19) mmHg. The biggest discrepancy between the methods was reflected by a large bias (limits of agreement) 5 (-34, 45) mmHg, was for SAP. The results for DAP were between those for SAP and MAP with a bias (limits of agreement) of 3 (-20, 27) mmHg. When the values were separated into the pressure range categories the HDO measured higher in the high, medium and low BP groups, with the exception of SAP in the low BP group.
CONCLUSIONS: When considering the mean bias, the accuracy of HDO compared well with direct arterial blood pressure, but the precision was poor, as determined by wide limits of agreement. CLINICAL RELEVANCE: Using trends and serial measurements rather than a single measurement for clinical decision making is recommended with both methods, when used as reported here.
© 2013 University of Liverpool. Veterinary Anaesthesia and Analgesia © 2013 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesia and Analgesia.

Entities:  

Keywords:  anaesthesia; arterial blood pressure; dog; high definition oscillometry; non-invasive blood pressure measurement

Mesh:

Year:  2013        PMID: 23565863     DOI: 10.1111/vaa.12035

Source DB:  PubMed          Journal:  Vet Anaesth Analg        ISSN: 1467-2987            Impact factor:   1.648


  8 in total

Review 1.  Non-invasive blood pressure measurement in animals: Part 1 - Techniques for measurement and validation of non-invasive devices.

Authors:  Alicia Skelding; Alexander Valverde
Journal:  Can Vet J       Date:  2020-04       Impact factor: 1.008

Review 2.  Basics of monitoring equipment.

Authors:  Tanya Duke-Novakovski
Journal:  Can Vet J       Date:  2017-11       Impact factor: 1.008

3.  Blood Pressure Reference Intervals for Ketamine-sedated Rhesus Macaques (Macaca mulatta).

Authors:  Rachel D Brownlee; Philip H Kass; Rebecca L Sammak
Journal:  J Am Assoc Lab Anim Sci       Date:  2020-01-02       Impact factor: 1.232

4.  Prevalence and risk factors associated with systemic hypertension in dogs with spontaneous hyperadrenocorticism.

Authors:  Paula García San José; Carolina Arenas Bermejo; Irene Clares Moral; Pedro Cuesta Alvaro; María Dolores Pérez Alenza
Journal:  J Vet Intern Med       Date:  2020-07-02       Impact factor: 3.333

5.  Impact of equipment and handling on systolic blood pressure measurements in conscious dogs in an animal hospital environment.

Authors:  Maria Lyberg; Ingrid Ljungvall; Jens Häggström; Ellinor Ahlund; Lena Pelander
Journal:  J Vet Intern Med       Date:  2021-02-15       Impact factor: 3.333

6.  Agreement of High-Definition Oscillometry (HDO) and Invasive Blood Pressure Measurements at a Metatarsal Artery in Isoflurane-Anaesthetised Horses.

Authors:  Lara Twele; Stephan Neudeck; Julien Delarocque; Nicole Verhaar; Julia Reiners; Mike Noll; Julia Tünsmeyer; Sabine B R Kästner
Journal:  Animals (Basel)       Date:  2022-02-02       Impact factor: 2.752

7.  Indirect arterial blood pressure measurement in healthy anesthetized cats using a device that combines oscillometry with photoplethysmography.

Authors:  Yasuhiro Heishima; Yasutomo Hori; Seishiro Chikazawa; Kazutaka Kanai; Fumio Hoshi; Naoyuki Itoh
Journal:  J Vet Med Sci       Date:  2016-03-18       Impact factor: 1.267

8.  Maternal Left Ventricular Diastolic and Systolic Function during Normal Pregnancy in Saint Bernard Dogs.

Authors:  Luís Lobo; Joana Esteves-Guimarães; Pedro Oliveira; Luís Salazar; André Pereira; Ana Patrícia Fontes-Sousa
Journal:  Vet Sci       Date:  2021-12-04
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.