| Literature DB >> 23560192 |
Yun-Jeong Chae1, Jong-Yeop Kim, Do-Won Kim, Bong-Ki Moon, Sang-Kee Min.
Abstract
BACKGROUND: We evaluated volumetric differences of syringe brand compatibilities, and investigated the impact of false brand settings on target-controlled infusion (TCI) and their methods of correction.Entities:
Keywords: Propofol; Syringe; Target-controlled infusion; Volumetric differences
Year: 2013 PMID: 23560192 PMCID: PMC3611076 DOI: 10.4097/kjae.2013.64.3.251
Source DB: PubMed Journal: Korean J Anesthesiol ISSN: 2005-6419
Fig. 1The schematic infusion assembly used for the determination of volumetric differences between the syringe brand compatibilities and TCI experiments for the corrections of the false setting of syringe compatibilities.
Percentage Differences (%) between the Nominal and Measured Bolus after Administration of 10 ml Bolus using BD Plastipak and Terumo Compatible Syringes with 7 Different Brand Compatibilities
Values indicate mean ± SD.
Processes of Active Corrections for the BDPTERUMO and TERUMOBDP during Targeting 4.0 µg/ml of Effect-site Concentration of Propofol, at the Time from the Start of Correction
BDPTERUMO: BD Plastipak syringe with false setting to Terumo, TERUMOBDP: Terumo syringe with false setting to BDP, Ceff: effect-site concentration.
*An activation of the function key of the software.
Fig. 2Time-courses of the reproduced actual plasma (solid lines) and the effect-site (dotted lines) concentrations during the target-controlled infusion of propofol using the BD Plastipak syringe with false setting to Terumo. Simple (upper graph) and active (lower graph) corrections were performed at 30 min (Peak A = the test increase, Peak B = the manual fill-up bolus, Peak C = the syringe replacement).
Fig. 3Time-courses of the reproduced actual plasma (solid lines) and the effect-site (dotted lines) concentrations during the target-controlled infusion of propofol using the Terumo syringe with false setting to BD Plastipak. Simple (upper graph) and active (lower graph) corrections were performed at 30 min.
Simple and Active Correction for the BDPTERUMO and TERUMOBDP Targeting Effect-site Concentration of Propofol, and Time to Reach ± 5% of Target
Values indicate mean ± SD. BDPTERUMO: BD Plastipak syringe with false setting to Terumo, TERUMOBDP: Terumo syringe with false setting to BDP, Ceff: effect-site concentration, T±5%target: time to reach ± 5% of target.
Fig. 4The nominal (dotted lines) and reproduced actual (solid lines) predicted plasma concentrations of propofol at the central compartment, C1 (upper), and at two peripheral compartments, C2 (middle), and C3 (lower) targeting 4.0 concentration, and the active corrections were performed for the BDP with false setting to TERUMO (upper graph) and the TERUMO with false setting to BDP (lower graph).