Accurate continuous drug delivery at low infusion rate with a novel microvolumetric infusion pump (MVIP): pump design, evaluation and comparison to the current standard.

M Weiss1, S Gerber, R M Füchslin, T A Neff.   

Abstract

Infusion devices for continuous and precise drug administration are indispensable tools in anaesthesia and critical care medicine. Problems such as start-up delays, non-continuous flow and susceptibility to hydrostatic pressure changes at low infusion rates resulting in accidental bolus release or prolonged flow interruption are inherent to current infusion technology. In order to improve precise drug delivery, an innovative technical concept has been realised in a novel microvolumetric infusion pump (MVIP) device. The MVIP principle includes repeated filling and emptying of a non-compliant microsyringe without the use of valves. The performance of the MVIP prototype has been evaluated and compared with standard syringe infusion pump assemblies. The novel MVIP concept has thereby proven to eliminate most problems during infusion start-up, steady state flow and vertical pump displacement, and has the potential of revolutionising infusion technology and setting a new dimension in patient safety.

Entities:  

Mesh:

Year:  2004        PMID: 15479325     DOI: 10.1111/j.1365-2044.2004.03898.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  3 in total

1.  Microfluidic Thermal Flowmeters for Drug Injection Monitoring.

Authors:  Il Doh; Daniel Sim; Steve S Kim
Journal:  Sensors (Basel)       Date:  2022-04-20       Impact factor: 3.847

2.  Digital droplet infusion.

Authors:  Zecong Fang; Andrew I Li; Hong Liu; Tingrui Pan
Journal:  Lab Chip       Date:  2020-11-13       Impact factor: 6.799

3.  False selection of syringe-brand compatibility and the method of correction during target-controlled infusion of propofol.

Authors:  Yun-Jeong Chae; Jong-Yeop Kim; Do-Won Kim; Bong-Ki Moon; Sang-Kee Min
Journal:  Korean J Anesthesiol       Date:  2013-03-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.