In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants.

F Duran-Sindreu1, S Gomes, E Stöber, M Mercadé, L Jané, M Roig.   

Abstract

AIM: To evaluate in vivo the performance of the iPex and Root ZX electronic apex locators (EALs) in the presence of several irrigant solutions: 2.5% sodium hypochlorite (NaOCl) and 2% chlorhexidine (CHX).
METHODOLOGY: Thirty-two single-rooted human teeth that were scheduled for extraction were selected. Teeth with metallic restorations, fractures, root resorption, pulp necrosis or open apices were not included The working length (WL) was determined electronically for the root canals with the iPex and Root ZX EALs in the presence of two different irrigant solutions, 2.5% NaOCl and 2% CHX. After the teeth had been extracted, a size 10 K-file was used to determine the reference working length (RWL), which was established at 0.5 mm short of the major foramen. In each case, the RWL was subtracted from the electronic measurements. Positive values indicated electronic measurements that exceeded the RWL (long measurements), whereas negative values indicated measurements that were short of the RWL. The values obtained with the different irrigants and EALs were compared using the paired t-test. Significance was set at P < 0.05.
RESULTS: The accuracy of the iPex nor Root ZX EAL was not affected by 2.5% NaOCl or 2% CHX (P > 0.05). However, significant differences were observed between the readings of the iPex and Root ZX, irrespective of whether 2.5% NaOCl or 2% CHX was used as the irrigant (P < 0.05). The iPex was less accurate than the Root ZX in determining the RWL.
CONCLUSIONS: The accuracy of neither the iPex nor Root ZX EAL was affected by the irrigant used. However, the iPex was less accurate than the Root ZX in determining the RWL both for 2.5% NaOCl and for 2% CHX.
© 2013 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  Root ZX; electronic apex locator; iPex; working length

Mesh:

Substances:

Year:  2013        PMID: 23551276     DOI: 10.1111/iej.12057

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  9 in total

1.  Comparison of two electronic apex locators on human cadavers.

Authors:  Luca Marigo; Gian Luca Gervasi; Francesco Somma; Giuseppe Squeo; Raffaella Castagnola
Journal:  Clin Oral Investig       Date:  2015-11-04       Impact factor: 3.573

2.  The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.

Authors:  Carlos G Adorno; Sandra M Solaeche; Ileana E Ferreira; Andrea Pedrozo; Patricia M Escobar; Vicente R Fretes
Journal:  Clin Oral Investig       Date:  2021-02-13       Impact factor: 3.573

3.  A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.

Authors:  Mugem Asli Gurel; Bagdagul Helvacioglu Kivanc; Adil Ekici
Journal:  J Istanb Univ Fac Dent       Date:  2017-01-02

4.  The accuracy of the auto-stop function of different endodontic devices in detecting the apical constriction.

Authors:  David W Christofzik; Andreas Bartols; Mahmoud Khaled; Birte Größner-Schreiber; Christof E Dörfer
Journal:  BMC Oral Health       Date:  2017-11-29       Impact factor: 2.757

5.  Comparative Evaluation of Accuracy of Two Electronic Apex Locators in the Presence of Contemporary Irrigants: An In vitro Study.

Authors:  Queentaj Baruah; Nidhi Sinha; Bijay Singh; P Narendranatha Reddy; Kaveri Baruah; Vaishak Augustine
Journal:  J Int Soc Prev Community Dent       Date:  2018-07-18

6.  Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants.

Authors:  Sakshi Bilaiya; Pallav Mahesh Patni; Pradeep Jain; Sanket Hans Pandey; Swadhin Raghuwanshi; Bhupesh Bagulkar
Journal:  Eur Endod J       Date:  2020-03-19

7.  The Precision of Propex Pixi with Different Instruments and Coronal Preflaring Procedures.

Authors:  Inês Ferreira; Ana Cristina Braga; Irene Pina-Vaz
Journal:  Eur Endod J       Date:  2019-07-10

8.  Accuracy of two electronic apex locators in locating root perforations in curved canals in dry and wet conditions: A comparative in vitro study.

Authors:  Monisha Parshotam Khatri; Sheetal B Ghivari; Madhu Pujar; Reshma Faras; Pallavi Gopeshetti; Amulya Vanti
Journal:  Dent Res J (Isfahan)       Date:  2019-11-12

9.  Influence of Root Canal Curvature on the Accuracy of Root ZX Electronic Foramen Locator: An In Vitro Study.

Authors:  Masoud Saatchi; Shiva Iravani; Mehrdad Akhavan Khaleghi; Amin Mortaheb
Journal:  Iran Endod J       Date:  2017
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.