A comparison of noninvasive bioreactance with oesophageal Doppler estimation of stroke volume during open abdominal surgery: an observational study.

Daniel H Conway1, Osman A Hussain, Iain Gall.   

Abstract

CONTEXT: The anaesthetist must maintain tissue perfusion by ensuring optimal perioperative fluid balance. This can be achieved using less invasive cardiac output monitors such as oesophageal Doppler monitoring (ODM). Other less invasive cardiac output monitors using bio-impedence technology (noninvasive cardiac output monitoring, NICOM) may have a role in monitoring the circulation and informing fluid management decisions.
OBJECTIVE: To compare estimates of stroke volume from ODM with those from NICOM, a noninvasive monitor using bioreactance, a modification of transthoracic bio-impedence.
DESIGN: An observational study.
SETTING: Manchester Royal Infirmary, UK. Data collected in 2011 and 2012. PARTICIPANTS: Twenty-two patients scheduled for major, open abdominal surgery. Reasons for noninclusion: atrial fibrillation; heart failure; oesophageal disease; lack of capacity; and known sensitivity to colloid. INTERVENTION: All patients had oesophageal Doppler cardiac output monitoring as a standard element of anaesthesia care. We placed NICOM Bioreactance electrodes and recorded stroke volume estimates from both devices. Fluid challenges were given by the anaesthetist and the haemodynamic responses were recorded. MAIN OUTCOME MEASURE: Stroke volume during surgery. The Bland-Altman method was used to compare bias and limits of agreement for stroke volume and cardiac output. Fluid responders were defined as patients who increased stroke volume by at least 10% after fluid loading. The precision of each device was calculated during periods of haemodynamic stability.
RESULTS: We made 788 acceptable measurements of cardiac output. The bias was -6.9 ml and the limits of agreement were -22.9 to 36.8 ml. The percentage error was 57%. Average precision for both the ODM and NICOM were similar, 8.5% (SD 5.4%) and 8.7% (SD 3.2%). The concordance for the stroke volume change following fluid challenge was 90.5%. Both devices produced unacceptable readings with electrical diathermy.
CONCLUSION: Simultaneous stroke volume estimations made by noninvasive Bioreactance (NICOM) and oesophageal Doppler showed bias and limits of agreement that are not clinically acceptable. The measurements made by these two devices cannot be regarded as interchangeable.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23549128     DOI: 10.1097/EJA.0b013e3283603250

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  10 in total

1.  The effect of head up tilting on bioreactance cardiac output and stroke volume readings using suprasternal transcutaneous Doppler as a control in healthy young adults.

Authors:  Jie Zhang; Lester A H Critchley; Daniel C W Lee; Kim S Khaw; Shara W Y Lee
Journal:  J Clin Monit Comput       Date:  2016-02-12       Impact factor: 2.502

Review 2.  Passive leg raising for predicting fluid responsiveness: a systematic review and meta-analysis.

Authors:  Xavier Monnet; Paul Marik; Jean-Louis Teboul
Journal:  Intensive Care Med       Date:  2016-01-29       Impact factor: 17.440

Review 3.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

4.  Comparison of stroke volume measurement between non-invasive bioreactance and esophageal Doppler in patients undergoing major abdominal-pelvic surgery.

Authors:  Gennaro De Pascale; Mervyn Singer; David Brealey
Journal:  J Anesth       Date:  2017-04-08       Impact factor: 2.078

5.  The ability of stroke volume variation measured by a noninvasive cardiac output monitor to predict fluid responsiveness in mechanically ventilated children.

Authors:  Ji Yeon Lee; Ji Young Kim; Chang Hyu Choi; Hong Soon Kim; Kyung Cheon Lee; Hyun Jeong Kwak
Journal:  Pediatr Cardiol       Date:  2013-08-21       Impact factor: 1.655

6.  Lithotomy versus jack-knife position on haemodynamic parameters assessed by impedance cardiography during anorectal surgery under low dose spinal anaesthesia: a randomized controlled trial.

Authors:  Jurgita Borodiciene; Jurate Gudaityte; Andrius Macas
Journal:  BMC Anesthesiol       Date:  2015-05-06       Impact factor: 2.217

7.  Bioreactance Is Not Interchangeable with Thermodilution for Measuring Cardiac Output during Adult Liver Transplantation.

Authors:  Sangbin Han; Jong Hwan Lee; Gaabsoo Kim; Justin Sangwook Ko; Soo Joo Choi; Ji Hae Kwon; Burn Young Heo; Mi Sook Gwak
Journal:  PLoS One       Date:  2015-05-27       Impact factor: 3.240

8.  Haemodynamic Optimization by Oesophageal Doppler and Pulse Power Wave Analysis in Liver Surgery: A Randomised Controlled Trial.

Authors:  Aarne Feldheiser; Velizara Pavlova; Karin Weimann; Oliver Hunsicker; Martin Stockmann; Mandy Koch; Alexander Giebels; Klaus-Dieter Wernecke; Claudia D Spies
Journal:  PLoS One       Date:  2015-07-17       Impact factor: 3.240

Review 9.  Intravascular volume therapy in adults: Guidelines from the Association of the Scientific Medical Societies in Germany.

Authors:  Gernot Marx; Achim W Schindler; Christoph Mosch; Joerg Albers; Michael Bauer; Irmela Gnass; Carsten Hobohm; Uwe Janssens; Stefan Kluge; Peter Kranke; Tobias Maurer; Waltraut Merz; Edmund Neugebauer; Michael Quintel; Norbert Senninger; Hans-Joachim Trampisch; Christian Waydhas; Rene Wildenauer; Kai Zacharowski; Michaela Eikermann
Journal:  Eur J Anaesthesiol       Date:  2016-07       Impact factor: 4.330

10.  Emergency department non-invasive cardiac output study (EDNICO): an accuracy study.

Authors:  David McGregor; Shrey Sharma; Saksham Gupta; Shanaz Ahmed; Tim Harris
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2020-01-31       Impact factor: 2.953

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.