In vitro fatigue and fracture strength testing of one-piece zirconia implant abutments and zirconia implant abutments connected to titanium cores.

Michael Stimmelmayr1, Stefanie Sagerer, Kurt Erdelt, Florian Beuer.   

Abstract

PURPOSE: The purpose of this study was to determine the fatigue and fracture strength of zirconia implant abutments connected to titanium cores (ZrTi) and one-piece zirconia implant abutments (Zr).
MATERIALS AND METHODS: Thirty-two implants were secured into epoxy resin blocks and connected with abutments. Eight specimens of each group (Zr with a diameter of 3.75 mm [Zr3.75] or 5.5 mm [Zr5.5] and ZrTi with a diameter of 3.75 mm [ZrTi3.75] or 5.5 mm [ZrTi5.5]) were thermally cycled from 5°C to 55°C and loaded with 100,000 cycles at 120 N at 30 degrees off-axis. All specimens were then tested for fracture resistance using a compressive load at 30 degrees off-axis.
RESULTS: Abutment fracture or screw loosening was not observed during thermal cycling and cyclic loading. The median fracture resistance values and standard deviations were 526 N (± 32 N) for group Zr3.75, 1,241 N (± 269 N) for group ZrTi3.75, 1,894 N (± 137 N) for group Zr5.5, and 2,225 N (± 63 N) for group ZrTi5.5. Statistically significant differences in fracture strength were found between Zr and ZrTi for each implant diameter.
CONCLUSIONS: Zirconia implant abutments connected to titanium cores showed higher fracture strength compared to one-piece zirconia abutments. Hence, they might be preferable for clinical use.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23527351     DOI: 10.11607/jomi.2772

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  10 in total

1.  Mechanical performance of cement- and screw-retained all-ceramic single crowns on dental implants.

Authors:  Matthias Obermeier; Oliver Ristow; Kurt Erdelt; Florian Beuer
Journal:  Clin Oral Investig       Date:  2017-07-15       Impact factor: 3.573

2.  Fracture behavior of zirconia implant abutments is influenced by superstructure-geometry.

Authors:  Frank P Nothdurft; Konrad Neumann; Andreas W Knauber
Journal:  Clin Oral Investig       Date:  2013-09-20       Impact factor: 3.573

3.  Comparative evaluation of the wear resistance of two different implant abutment materials after cyclic loading - An in vitro study.

Authors:  Maniamuthu Ragupathi; Vallabh Mahadevan; N S Azhagarasan; Hariharan Ramakrishnan; S Jayakrishnakumar
Journal:  Contemp Clin Dent       Date:  2020-11-26

4.  Tensile Bond Strength of Three Custom-made Tooth-Colored Implant Superstructures to Titanium Inserts.

Authors:  Mitra Zirak; Mahroo Vojdani; Amir Ali Reza Khaledi; Mitra Farzin
Journal:  J Dent (Shiraz)       Date:  2019-06

5.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

6.  Fracture resistance of different implant supported ceramic abutment/crown systems.

Authors:  Merve Bankoglu Gungor; Secil Karakoca Nemli; Handan Yilmaz; Cemal Aydin
Journal:  Eur Oral Res       Date:  2019-05-01

7.  Fracture strength analysis of titanium insert-reinforced zirconia abutments according to the axial height of the titanium insert with an internal connection.

Authors:  Seung-Rye Song; Kyeong-Mee Park; Bock-Young Jung
Journal:  PLoS One       Date:  2021-04-01       Impact factor: 3.240

8.  Titanium Base Abutments in Implant Prosthodontics: A Literature Review.

Authors:  Ahmad M Al-Thobity
Journal:  Eur J Dent       Date:  2021-11-18

9.  Comparing the Fracture Resistance and Modes of Failure in Different Types of CAD/CAM Zirconia Abutments with Internal Hexagonal Implants: An In Vitro Study.

Authors:  Yu-Tsen Chang; Yu-Ling Wu; Hung-Shyong Chen; Ming-Hsu Tsai; Chia-Chen Chang; Aaron Yu-Jen Wu
Journal:  Materials (Basel)       Date:  2022-04-04       Impact factor: 3.623

10.  Effect of Implant Platform Connection and Abutment Material on Removal Torque and Implant Hexagon Plastic Deformation.

Authors:  Paulo Henrique de Souza Pereira; Marina Amaral; Kusai Baroudi; Rafael Pino Vitti; Mohammad Zakaria Nassani; Laís Regiane da Silva-Concílio
Journal:  Eur J Dent       Date:  2019-12-03
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.