The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling.

Juliana Ribeiro Pala Jorge1, Valentim Adelino Ricardo Barao, Juliana Aparecida Delben, Wirley Goncalves Assuncao.   

Abstract

PURPOSE: This study aimed to evaluate the role of the implant/abutment system on torque maintenance of titanium retention screws and the vertical misfit of screw-retained implant-supported crowns before and after mechanical cycling.
MATERIALS AND METHODS: Three groups were studied: morse taper implants with conical abutments (MTC group), external-hexagon implants with conical abutments (EHC group), and external-hexagon implants with UCLA abutments (EHU group). Metallic crowns casted in cobalt-chromium alloy were used (n = 10). Retention screws received insertion torque and, after 3 minutes, initial detorque was measured. Crowns were retightened and submitted to cyclic loading testing under oblique loading (30 degrees) of 130 ± 10 N at 2 Hz of frequency, totaling 1 × 106 cycles. After cycling, final detorque was measured. Vertical misfit was measured using a stereomicroscope. Data were analyzed by analysis of variance, Tukey test, and Pearson correlation test (P < .05).
RESULTS: All detorque values were lower than the insertion torque both before and after mechanical cycling. No statistically significant difference was observed among groups before mechanical cycling. After mechanical cycling, a statistically significantly lower loss of detorque was verified in the MTC group in comparison to the EHC group. Significantly lower vertical misfit values were noted after mechanical cycling but there was no difference among groups. There was no significant correlation between detorque values and vertical misfit.
CONCLUSIONS: All groups presented a significant decrease of torque before and after mechanical cycling. The morse taper connection promoted the highest torque maintenance. Mechanical cycling reduced the vertical misfit of all groups, although no significant correlation between vertical misfit and torque loss was found.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23527343     DOI: 10.11607/jomi.2727

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  12 in total

1.  Wear of Morse taper and external hexagon implant joints after abutment removal.

Authors:  Abraão M Prado; Jorge Pereira; Filipe S Silva; Bruno Henriques; Rubens M Nascimento; Cesar A M Benfatti; José López-López; Júlio C M Souza
Journal:  J Mater Sci Mater Med       Date:  2017-03-20       Impact factor: 3.896

2.  In Vitro Study of Preload Loss in Different Implant Abutment Connection Designs.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Marta Relvas; Ana Cristina Braga; Blanca Ríos-Carrasco; Javier Gil; José Vicente Rios-Santos; Ana Fernández-Palacín; Mariano Herrero-Climent
Journal:  Materials (Basel)       Date:  2022-02-14       Impact factor: 3.623

Review 3.  Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic review.

Authors:  Marcelo Coelho Goiato; Eduardo Piza Pellizzer; Emily Vivianne Freitas da Silva; Liliane da Rocha Bonatto; Daniela Micheline dos Santos
Journal:  Oral Maxillofac Surg       Date:  2015-04-25

Review 4.  Preload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review.

Authors:  Beatriz Pardal-Peláez; Javier Montero
Journal:  J Clin Exp Dent       Date:  2017-11-01

5.  Marginal Vertical Fit along the Implant-Abutment Interface: A Microscope Qualitative Analysis.

Authors:  Nicola Mobilio; Alberto Fasiol; Giulio Franceschetti; Santo Catapano
Journal:  Dent J (Basel)       Date:  2016-09-06

6.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

7.  Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force.

Authors:  Yuseung Yi; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

Review 8.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

9.  Analysis of Torque Maintenance and Fracture Resistance after Fatigue in Retention Screws Made of Different Metals for Screw-Retained Implant-Borne Prosthesis Joints.

Authors:  Maria Beatriz Bello Taborda; Gabriela Sumie Yaguinuma Gonçalves; Cecília Alves de Sousa; Wirley Gonçalves Assunção
Journal:  Int J Dent       Date:  2021-11-18

10.  Influence of Manufacturing Methods of Implant-Supported Crowns on External and Internal Marginal Fit: A Micro-CT Analysis.

Authors:  Izabela C M Moris; Silas Borges Monteiro; Raíssa Martins; Ricardo Faria Ribeiro; Erica A Gomes
Journal:  Biomed Res Int       Date:  2018-01-23       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.