Francis Bouchet1, Lilli Geworski, Bernd O Knoop, Ludovic Ferrer, Alina Barriolo-Riedinger, Corinne Millardet, Marjolaine Fourcade, Antoine Martineau, Anne Belly-Poinsignon, Francis Djoumessi, Karine Tendero, Laurent Keros, Frederic Montoya, Christel Mesleard, Anne-Laure Martin, Franck Lacoeuille, Olivier Couturier.
Abstract
UNLABELLED: A multi-centre trial using PET requires the analysis of images acquired on different systems We designed a multi-centre trial to estimate the value of 18F-FLT-PET to predict response to neoadjuvant chemotherapy in patients with newly diagnosed breast cancer. A calibration check of each PET-CT and of its peripheral devices was performed to evaluate the reliability of the results.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23516443 PMCID: PMC3596383 DOI: 10.1371/journal.pone.0058152
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Dose calibrator reproducibility and accuracy results.
| Measurement # | ||||||||||||||
| Site # | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | Mean | SD | Standard source activity | Accuracy |
| 1 | 46.15 | 46.20 | 46.18 | 46.13 | 46.16 | 46.16 | 46.16 | 46.14 | 46.11 | 46.12 | 46.15 | 0.06 | 45.48 | 1,47 |
| 2 | 43.30 | 43.30 | 43.30 | 43.30 | 43.30 | 42.90 | 43.30 | 43.30 | 43.30 | 43.30 | 43.26 | 0.29 | 43.66 | −0,91 |
| 3 | 43.70 | 43.70 | 43.70 | 43.70 | 43.70 | 43.70 | 43.70 | 43.70 | 43.70 | 43.70 | 43.70 | 0.00 | 41.37 | 5,63 |
| 4 | 38.20 | 37.90 | 37.80 | 37.80 | 37.90 | 38.00 | 38.00 | 37.90 | 37.80 | 37.80 | 37.91 | 0.34 | 37.92 | −0,04 |
| 5 | 36.98 | 36.89 | 36.86 | 36.89 | 36.82 | 36.87 | 36.82 | 36.86 | 36.88 | 36.81 | 36.87 | 0.13 | 37.25 | −1,03 |
| 6 | 34.42 | 34.34 | 34.37 | 34.39 | 34.42 | 34.38 | 34.39 | 34.37 | 34.33 | 34.35 | 34.38 | 0.09 | 32.86 | 4,61 |
| 7 | 30.77 | 30.81 | 30.76 | 30.77 | 30.8 | 30.79 | 30.75 | 30.79 | 30.76 | 30.76 | 30.78 | 0.07 | 32.28 | −4,65 |
| 8 | 12.50 | 12.50 | 12.68 | 12.53 | 12.52 | 12.54 | 12.52 | 12.51 | 12.49 | 12.50 | 12.53 | 0.44 | 12.46 | 0,59 |
| 9 | 9.90 | 9.80 | 9.90 | 9.90 | 9.90 | 9.90 | 9.80 | 9.80 | 9.90 | 9.90 | 9.87 | 0.49 | 9.87 | 0,03 |
| 10 | 7.60 | 7.60 | 7.60 | 7.60 | 7.60 | 7.60 | 7.60 | 7.60 | 7.60 | 7.60 | 7.60 | 0.00 | 7.84 | −3,02 |
| 11 | 6.90 | 6.91 | 6.92 | 6.94 | 6.92 | 6.94 | 6.91 | 6.93 | 6.94 | 6.91 | 6.92 | 0.21 | 6.97 | −0,64 |
Repeated measurements of certified standard source (68Ge) on the dose calibrators of 11 sites. Standard source activity was calculated by applying decay calculation between the calibration date and the measurements date. Measurements expressed in MBq. Standard deviation (SD) and accuracy are in %.
Mean activity concentration estimated from 15 cm diameter circular Regions of Interest (ROIs) on 15 consecutive slices and accuracy.
| Slice # | Max deviation | Activity in phantom | Accuracy | |||||||||||||||||
| Site # | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 | 15 | Mean | SD | |||
| 1 | 53.23 | 52.89 | 53.08 | 52.90 | 53.26 | 53.07 | 53.20 | 53.13 | 53.44 | 53.45 | 55.16 | 54.81 | 53.93 | 53.77 | 54.17 | 53.57 | 1.28 | 2.97 | 53.83 | −0.49 |
| 3 | 48.20 | 48.30 | 48.40 | 48.44 | 48.41 | 48.29 | 48.13 | 48.11 | 48.44 | 48.61 | 48.55 | 48.63 | 48.60 | 48.44 | 48.57 | 48.41 | 0.36 | −0.62 | 47.32 | 2.30 |
| 4 | 46.26 | 45.95 | 45.30 | 45.46 | 45.39 | 45.31 | 45.58 | 45.85 | 45.91 | 46.07 | 45.98 | 46.70 | 46.21 | 47.12 | 48.61 | 46.11 | 1.86 | 5.42 | 51.22 | −9.97 |
| 7 | 48.54 | 48.57 | 48.53 | 48.53 | 48.55 | 48.59 | 48.55 | 48.57 | 48.63 | 48.52 | 48.64 | 48.77 | 48.72 | 48.63 | 48.48 | 48.59 | 0.16 | 0.37 | 51.35 | −5.38 |
Data acquired on the 4 PET systems used in 2D mode. Activity concentration in images and in phantom expressed in kBq/mL. Standard deviation (SD), accuracy and maximum deviation are in %.
Mean activity concentration estimated from 15 cm diameter circular Regions of Interest (ROIs) on 15 consecutive slices and accuracy.
| Slice # | Max deviation | Activity in phantom | Accuracy | |||||||||||||||||
| Site # | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 | 15 | Mean | SD | |||
| 1 | 27.06 | 26.38 | 25.71 | 25.23 | 25.21 | 25.30 | 25.62 | 25.88 | 26.32 | 26.12 | 25.30 | 23.96 | 23.48 | 23.81 | 24.37 | 25.32 | 4.06 | −7.25 | 23.10 | 9.60 |
| 2 | 17.78 | 17.76 | 17.73 | 17.73 | 17.74 | 17.71 | 17.69 | 17.68 | 17.69 | 17.73 | 17.76 | 17.74 | 17.73 | 17.74 | 17.74 | 17.73 | 0.16 | 0.30 | 18.03 | −1.67 |
| 4 | 19.12 | 18.90 | 18.27 | 17.81 | 17.57 | 17.48 | 17.58 | 17.71 | 17.90 | 18.25 | 18.43 | 18.47 | 17.51 | 17.00 | 16.83 | 17.92 | 3.62 | 6.71 | 16.97 | 5.61 |
| 5 | 15.94 | 15.87 | 15.85 | 15.80 | 15.75 | 15.78 | 15.80 | 15.65 | 15.70 | 15.54 | 15.53 | 15.61 | 15.54 | 15.53 | 15.41 | 15.69 | 0.99 | −1.74 | 15.66 | 0.17 |
| 6 | 18.04 | 18.02 | 18.03 | 18.03 | 18.01 | 17.98 | 18.00 | 18.07 | 18.09 | 18.04 | 17.97 | 17.93 | 17.93 | 17.97 | 18.03 | 18.01 | 0.26 | 0.46 | 18.17 | −0.90 |
| 9 | 17.04 | 16.75 | 16.97 | 16.81 | 16.82 | 16.93 | 16.86 | 16.83 | 16.88 | 16.84 | 16.84 | 16.81 | 16.79 | 16.63 | 16.82 | 16.84 | 0.56 | −1.23 | 17.07 | −1.34 |
| 10 | 18.12 | 18.12 | 18.09 | 18.05 | 18.05 | 18.02 | 18.02 | 17.96 | 17.96 | 17.93 | 17.89 | 17.88 | 17.94 | 17.96 | 18.03 | 18.00 | 0.43 | 0.67 | 17.41 | 3.40 |
| 11 | 19.74 | 19.94 | 20.16 | 20.48 | 20.24 | 20.06 | 19.92 | 19.84 | 19.78 | 19.70 | 19.59 | 19.51 | 19.41 | 19.37 | 19.39 | 19.81 | 1.67 | 3.39 | 20.10 | −1.45 |
Data acquired on the 8 PET systems used in 3D mode. Activity concentration in images and in phantom expressed in kBq/mL. Standard deviation (SD), accuracy and maximum deviation are in %.
Figure 1Observed artefacts on images acquired on uniform phantom.
Display scale is [0; 100%]. A: no artefacts (site 9), B: LOR-RAMLA concentric ring artefacts (site 6), C: FBP concentric ring artefacts (site 4).
Figure 2Central profile on a transverse slice of phantoms.
Data are normalized to the maximum value intensity of the profile. No abnormality was seen.