Literature DB >> 23507528

Integration of fluoroscopy-based guidance in orthopaedic trauma surgery - a prospective cohort study.

Michael Kraus1, Stephanie von dem Berge, Hendrik Schöll, Gert Krischak, Florian Gebhard.   

Abstract

INTRODUCTION: Computer-assisted guidance systems are not used frequently for musculoskeletal injuries unless there are potential advantages. We investigated a novel fluoroscopy-based image guidance system in orthopaedic trauma surgery.
MATERIALS AND METHODS: The study was a prospective, not randomised, single-centre case series at a level I trauma centre. A total of 45 patients with 46 injuries (foot 12, shoulder 10, long bones seven, hand and wrist seven, ankle seven and spine and pelvis four) were included. Different surgical procedures were examined following the basic principles of the Arbeitsgemeinschaft für Osteosynthesefragen/Association for the Study of Internal Fixation (AO/ASIF). Main outcome measurements were the number of trials for implant placement, total surgery time, usability via user questionnaire and system failure rate.
RESULTS: In all cases, the trajectory function was used, inserting a total of 56 guided implants. The system failed when used in pelvic and spinal injuries, resulting in a total failure rate of 6.5% (n=3) of all included cases. The overall usability was rated as good, scoring 84.3%.
CONCLUSION: The novel image-guidance system could be integrated into the surgical workflow and was used successfully in orthopaedic trauma surgery. Expected advantages should be explored in randomised studies.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  C-arm; Computer navigation; Fluoroscopy; Fracture; Image-based guidance; Intra-operative imaging; Trauma

Mesh:

Year:  2013        PMID: 23507528     DOI: 10.1016/j.injury.2013.02.008

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  4 in total

1.  Ultrasound imaging and segmentation of bone surfaces: A review.

Authors:  Ilker Hacihaliloglu
Journal:  Technology (Singap World Sci)       Date:  2017-03-31

2.  Preclinical usability study of multiple augmented reality concepts for K-wire placement.

Authors:  Marius Fischer; Bernhard Fuerst; Sing Chun Lee; Javad Fotouhi; Severine Habert; Simon Weidert; Ekkehard Euler; Greg Osgood; Nassir Navab
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-03-19       Impact factor: 2.924

3.  Influence of multi-angle input of intraoperative fluoroscopic images on the spatial positioning accuracy of the C-arm calibration-based algorithm of a CAOS system.

Authors:  Xiangqian Chen; Yu Wang; Gang Zhu; Weijun Zhang; Gang Zhou; Yubo Fan
Journal:  Med Biol Eng Comput       Date:  2020-01-09       Impact factor: 2.602

4.  Automatic intraoperative stitching of nonoverlapping cone-beam CT acquisitions.

Authors:  Javad Fotouhi; Bernhard Fuerst; Mathias Unberath; Stefan Reichenstein; Sing Chun Lee; Alex A Johnson; Greg M Osgood; Mehran Armand; Nassir Navab
Journal:  Med Phys       Date:  2018-04-10       Impact factor: 4.071

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.