A community-based evaluation of the Vita-Stat automatic blood pressure recorder.

K Salaita1, P K Whelton, A J Seidler.   

Abstract

A random sample of ten community-based Vita-Stat automatic blood pressure recorders (ABPR) was evaluated for accuracy and repeatability. Each of 518 adult volunteers had two randomly ordered systolic and diastolic Vita-Stat blood pressure measurements compared with two corresponding measurements obtained by a trained observer using a Random-Zero sphygmomanometer. Eight of the ten Vita-Stat recorders underestimated systolic blood pressure. In contrast, diastolic blood pressure was overestimated by nine of the ten machines. While the overall differences in average blood pressure were small (mean +/- SEM = -2.4 +/- 0.6 mm Hg for systolic blood pressure and +2.3 +/- 0.4 mm Hg for diastolic blood pressure), the magnitude of the average discrepancy varied considerably by machine (+4.7 to -13.8 mm Hg for systolic and +5.0 to -2.0 for diastolic blood pressure). At every level of systolic and diastolic blood pressure the Vita-Stat ABPR provided a less accurate method of classifying blood pressure among individuals than the human observer. Additional analyses exploring the ability of the Vita-Stat machine to measure an individual's blood pressure within 2, 4, or 6 mm Hg of the corresponding Random-Zero value again suggested that the Vita-Stat ABPR was less accurate than the human observer. Duplicate Vita-Stat readings were less repeatable than corresponding Random-Zero measurements. Based on the findings of this study, the Vita-Stat ABPR appears to be an unsatisfactory tool for community-based self-measurement of blood pressure.

Entities:  

Mesh:

Year:  1990        PMID: 2350476     DOI: 10.1093/ajh/3.5.366

Source DB:  PubMed          Journal:  Am J Hypertens        ISSN: 0895-7061            Impact factor:   2.689


  3 in total

1.  Accuracy of monitors used for blood pressure checks in English retail pharmacies: a cross-sectional observational study.

Authors:  James Hodgkinson; Constantinos Koshiaris; Una Martin; Jonathan Mant; Carl Heneghan; Fd Richard Hobbs; Richard J McManus
Journal:  Br J Gen Pract       Date:  2016-03-29       Impact factor: 5.386

2.  Evaluation of a community-based automated blood pressure measuring device.

Authors:  Jacqueline E Lewis; Eleanor Boyle; Lucy Magharious; Martin G Myers
Journal:  CMAJ       Date:  2002-04-30       Impact factor: 8.262

3.  Hypertensive patients' use of blood pressure monitors stationed in pharmacies and other locations: a cross-sectional mail survey.

Authors:  Anthony J Viera; Lauren W Cohen; C Madeline Mitchell; Philip D Sloane
Journal:  BMC Health Serv Res       Date:  2008-10-22       Impact factor: 2.655