A multicenter evaluation of a compact, sterile, single-use pressure transducer for central venous catheter placement.

Kei Togashi1, Koichiro Nandate, Carli Hoaglan, Benjamin Sherman, Andrew Bowdle.   

Abstract

BACKGROUND: Inadvertent arterial placement of a large-bore catheter during attempted placement of a central venous catheter (CVC) occurs at a rate of 0.1% to 1.0% and may result in hemorrhage, pseudoaneurysm, stroke, or death. Ultrasound guidance or observation of color and pulsatility of blood are not reliable methods for avoiding this serious complication. Measurement of pressure in the needle or short plastic catheter before insertion of the guidewire has been shown to be highly reliable; however, traditional pressure measurement methodology is cumbersome. Recently a compact, sterile, single-use pressure transducer with an integrated digital display has become available. In this study, we evaluated the performance of this new device (Compass® Vascular Access).
METHODS: In this prospective, observational study at 4 academic medical centers 298 CVCs were placed. Pressure was measured using the Compass transducer before and after guidewire insertion. Other details of the procedure were at the discretion of the clinician. Data describing the CVC placement and any complications were collected.
RESULTS: Trainees placed 279 of 298 CVCs. Ultrasound guidance was used for 286 of 298 CVCs. Seven of the CVC placements occurred in the intensive care unit, with the balance occurring in the operating room. Ten of the CVCs were placed in a subclavian vein, with the balance being internal jugular vein. Two hundred seventy-four of 298 CVCs were placed on the right side. Venous pressure measured before and after guidewire insertion was 7.2 ± 4.3 (SD) and 6.5 ± 4.3 (SD) mm Hg respectively (P = 0.03). The satisfaction score recorded by the physician performing the procedure was 8.0 ± 2.1 (SD; visual analog scale 1-10, 10 being most satisfying). There were 5 inadvertent arterial punctures (1.7%). Ultrasound guidance was used in all 5 cases of arterial puncture. All of the arterial punctures were recognized before guidewire insertion by measurement of arterial pressure with the Compass transducer. No guidewires or CVC catheters were placed in arteries.
CONCLUSION: The Compass pressure transducer for CVC placement performed as intended in 298 cases from 4 academic medical centers. There were 5 inadvertent arterial punctures despite the use of ultrasound guidance, all of which were correctly identified by pressure measurement using the Compass. The device was easily used by trainees, and users expressed a positive level of satisfaction.

Entities:  

Mesh:

Year:  2013        PMID: 23492959     DOI: 10.1213/ANE.0b013e31828a6e53

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  7 in total

Review 1.  Assistive technology for ultrasound-guided central venous catheter placement.

Authors:  Mohammad Ikhsan; Kok Kiong Tan; Andi Sudjana Putra
Journal:  J Med Ultrason (2001)       Date:  2017-04-19       Impact factor: 1.314

2.  Inferior thyroid artery pseudoaneurysm associated with internal jugular vein puncture: a case report.

Authors:  Jinguang Ruan; Cao Zhang; Zhiyou Peng; David Yue Tang; Zhiying Feng
Journal:  BMC Anesthesiol       Date:  2015-05-06       Impact factor: 2.217

3.  Spectral tissue sensing to identify intra- and extravascular needle placement - A randomized single-blind controlled trial.

Authors:  Andrea J R Balthasar; Geert-Jan van Geffen; Marjolein van der Voort; Gerald W Lucassen; Stefan Roggeveen; Ivar J Bruaset; Joergen Bruhn
Journal:  PLoS One       Date:  2017-03-09       Impact factor: 3.240

4.  Novel technique for continuous pressure monitoring of esophageal balloon in balloon tamponade device for acute variceal bleed.

Authors:  Brian H Torres-Gonzalez; Thomas L Powell; Jennifer Yee; Katherine M Hunold
Journal:  J Am Coll Emerg Physicians Open       Date:  2022-04-29

5.  Inferior vena cava/aorta diameter index in the assessment of the body fluid status - a comparative study of measurements performed by experienced and inexperienced examiners in a group of young adults.

Authors:  Kaja Durajska; Emilia Januszkiewicz; Łukasz Szmygel; Wojciech Kosiak
Journal:  J Ultrason       Date:  2014-09-30

6.  Mini-invasive treatment of a large pseudoaneurysm of the neck related to central venous catheter placement: A case report.

Authors:  Chiara Palermo; Angelo Sanfiorenzo; Alessia Testo Giaquinta; Carla Virgilo; Massimiliano Veroux; Pierfrancesco Veroux
Journal:  Medicine (Baltimore)       Date:  2018-07       Impact factor: 1.889

7.  Comparison Of Digital Manometer And Water Column Manometer Pressures Measurements During Lumbar Puncture.

Authors:  Nitin K Sekhri; Shalvi Parikh; Garret M Weber
Journal:  Med Devices (Auckl)       Date:  2019-10-29
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.