Literature DB >> 31754314

Comparison Of Digital Manometer And Water Column Manometer Pressures Measurements During Lumbar Puncture.

Nitin K Sekhri1,2, Shalvi Parikh1, Garret M Weber1,2.   

Abstract

BACKGROUND: Cerebrospinal fluid (CSF) pressure measurement is routinely performed via a conventional water column manometer. There is increasing interest in using a digital manometer in measuring CSF pressures. The aim of this study is to compare column and digital manometers, in addition to measuring time to acquire the pressure readings. RESEARCH DESIGN AND METHODS: This prospective study included 27 patients who were referred for a fluoroscopically guided lumbar puncture. Opening pressure and closing pressure measurements were done with a digital manometer and then a traditional water column manometer. The time to obtain each pressure measurement was also recorded and compared.
RESULTS: Mean time to obtain pressure reading was significantly lower in the digital manometer group when compared to the water column manometer group (8.1 seconds vs. 42.2 seconds, P<0.05 for opening pressure and 8.92 seconds vs. 45.15 seconds, P<0.05 for closing pressure). Correlation between the opening pressure measurements (Pearson coefficient r= 0.98) and closing pressure (Pearson coefficient r= 0.89) was strong. However, the digital manometer reading consistently read higher.
CONCLUSION: Digital manometry during an LP yielded is faster however there might be a clinical difference between the devices. Clinicians must be careful in using the device across all cases.
© 2019 Sekhri et al.

Entities:  

Keywords:  IIH; closing pressure; digital manometer; idiopathic intracranial hypertension; lumbar puncture; normal pressure hydrocephalus; opening pressure

Year:  2019        PMID: 31754314      PMCID: PMC6825514          DOI: 10.2147/MDER.S225757

Source DB:  PubMed          Journal:  Med Devices (Auckl)        ISSN: 1179-1470


Introduction

The lumbar puncture (LP) was first described in the late 1890s by Dr. Heinrich Irenäus Quincke and continues to be an essential procedure for both diagnostic and therapeutic purposes.1–3 The indications for LP vary but include investigation to diagnose subarachnoid hemorrhage, meningitis, neurological disorders which include demyelinating disorders such as multiple sclerosis, as well as managing and diagnosing disorders of intracranial pressure, such as idiopathic intracranial hypertension (IIH) and normal pressure hydrocephalus.2,3 The opening pressure (OP) in particular has been proven to correlate and serve as a reliable marker for intracranial pressure, thus, can be critical in the diagnoses of many diseases.4 In some instances, such as with idiopathic intracranial hypertension and NPH, the diagnosis can be made or confirmed by OP alone.5–7 As such, the OP measurement is a crucial part of the LP, and accuracy of the OP measurement is paramount.8 Additionally, often for diagnostic and therapeutic purposes, the closing pressure (CP) is taken immediately after the necessary CSF is collected and prior to removal of the needle from the intrathecal space.9 Traditionally, a water column manometer is used for pressure measurement during the LP procedure. However, using a standard water column manometer carries some disadvantages: first, pressure reading can be time-consuming; second, these manometers are bulky in size, therefore require manual dexterity for stability and sometimes one may require two people during measurement; third measurements may easily be altered by the position of the manometer. As such, a disposable digital pressure transducer, the Compass device (Centurion Medical Products, Williamston, MI), is available for direct measurement and monitoring of pressures during LP.10 For central venous catheter placement, a similar compact digital pressure-measuring device was shown to be safe, simple with improved provider satisfaction and less cumbersome than traditional manometry methods (including water column manometry) to confirm venous access.11 However, the authors were not able to identify any research in the literature regarding its utilization during LP. After using the Compass device for several pressure readings during LPs, the authors reasoned that the digital device allows for CSF readings with significant procedural time reduction. However, there is a scarcity of data on the use of the product for LP pressure measurement. The purpose of this study compared the digital device with the standard of care water column manometry.

Patients And Methods

Approval was obtained and overseen by from New York Medical College institutional review board before the study commenced. All patients provided written informed consent before undergoing study procedures. This study was performed in compliance with the Health Insurance Portability and Accountability Act along with adhering to the principles of Good Clinical Practice and the Declaration of Helsinki. Patients presenting for fluoroscopic-guided LPs were asked to participate in this prospective study. Written consent was obtained for both procedure and for research study. Patients were positioned in the left lateral decubitus position with slight cervical spine flexion and with knees slightly drawn to their chest as per normal practice for obtaining accurate pressures during lumbar puncture.2,5,12–14 Using sterile technique, a 20-gauge Quincke needle was advanced into the subarachnoid space under fluoroscopic guidance. The L4–5 interspace was the first site of attempt, but if based on imaging, was deemed to be difficult, the L3–4 interspace was then used. Four pressure measurements were then obtained during the procedure. First, an OP using the digital device (OPD), then an OP measured using the water column manometer (OPC). CP also obtained in a similar manner, obtaining the CP first for the digital device (CPD) and then the water column manometer (CPC). Measurements obtained using the digital device were in accordance with the manufacturer’s directions when the measurement stabilized on the display screen after successful calibration, removal of luer cap and stabilization of pressure for 2–3 s.15 The pressure measurements with the water column manometer were done in accordance with good practice.16 The measurements were taken by connecting the needle directly (without tubing) to the manometer via a 4-way stopcock while the patient was in the lateral decubitus. The patient was encouraged to relax, breath normally and not Valsalva. The manometer was kept level with the hub, and the pressure was measured when the CSF meniscus stopped rising and respiratory fluctuations were apparent. All pressures were recorded in cmH2O (see Figure 1 for an example of pressure measurements using the two devices).
Figure 1

Images of two forms of pressure measurement used during study in the lateral decubitus position. (A) Water column manometer directly attached to needle via a 4-way stopcock and (B) The digital compass device attached directly to needle.

Images of two forms of pressure measurement used during study in the lateral decubitus position. (A) Water column manometer directly attached to needle via a 4-way stopcock and (B) The digital compass device attached directly to needle. Time (in seconds) to obtain each measurement was also recorded as Time for OPD, Time for OPC, Time for CPD and T for CPC. Basic Demographics of the patients were also recorded, such as gender, height, weight (and thus body mass index) and indication for the procedure. Patients were excluded if they were unable to consent, a minor (age <18), hemodynamically unstable, required mechanical respiratory support, or patients with OP was above the limit of the water column manometer (56 cmH2O). Statistical analysis was done using Excel (2016; Microsoft, Redmond, Washington). The digital times and water column manometer times differences were compared using a two sided-tailed paired t-test. To identify the differences in measurements between the digital and water column manometer, a Bland–Altman plot was generated comparing aggregate pressure (both closing and OPs), isolated OP and isolated CP.

Results

Twenty-seven patients were recruited to participate in the study. All of these patients were able to undergo LP with measurement of the four different pressure measurements as described in the materials and methods. Demographics and indications of each patient were recorded (Table 1). Infectious workup accounted for the most common indication (33%). IIH was a common indication accounting for 5 (19%) of the indications. These patients had a higher aggregate pressure as expected, with a mean of 25 cmH2O compared to 20 cmH2O for the rest of the patients.
Table 1

Demographics And Indications For Referral For The Fluoroscopic LP Along With The Mean Opening Pressure Readings For Each Indication

ROW LabelsNo. PatientsMean OPAverage Of Weight (kg)Average Of AgeAverage Of Height (cm)
Dementia4228066169
IIH52510627166
Infectious9177257170
Inflammatory6229243174
Malignancy32110533159
Mean218747169
Sum27
Demographics And Indications For Referral For The Fluoroscopic LP Along With The Mean Opening Pressure Readings For Each Indication When comparing times to obtain the pressure readings, there was a significant difference seen across all pressures. See Table 2 for time comparisons of the water column manometer and the digital manometer.
Table 2

Mean Pressure Readings (cmH2O) And Time (In Seconds) To Obtain Pressure Reading For Digital Manometer And Water Column Manometer With P-Values

_P-Digital (cmH2O)P-Column (cmH2O)pTime-Digital (sec)Time-Column (sec)p
Opening Pressure21.719.9<0.0018.142.2<0.001
Closing Pressure13.511.9<0.0038.945.1<0.001
All Pressures Readings (mean)17.615.9<0.0018.543.7<0.001
Mean Pressure Readings (cmH2O) And Time (In Seconds) To Obtain Pressure Reading For Digital Manometer And Water Column Manometer With P-Values When comparing the OPs measured, there was a significant correlation between the two measurements, with a Pearson coefficient of 0.97 (Figure 2A) along with a strong correlation for CPs with a Pearson coefficient of 0.89 (Figure 2B). However, while there was strong correlation between the two devices, the digital device pressure reading was consistently above that measured by the water column manometer. The mean difference between the measurements was 1.85 cmH2O for OPs (Figure 3A), with the digital device commonly measuring above the water column manometer, and the mean difference for CP was 1.56 cmH2O (Figure 3B), again with the digital device reading above the water column manometer.
Figure 2

Linear regression of digital pressure readings of the two different devices with a line of identity (dashed line). (A) Comparison of pressure readings during opening pressure (r= 0.97) (B) Comparison of pressure readings during closing pressure (r=0.89).

Figure 3

Bland–Altman plots showing the difference between the digital manometer reading and concurrent water column manometer reading during LP. The y-axis shows the mean difference of the two readings, with positive mean representing higher pressure by digital device and x-axis being the pressure reading in cmH2O (A) Plot for opening pressure with the mean difference (dashed line) positive indicating higher pressure for the digital device (B) Plot for closing pressures again with a positive mean difference indicating higher pressures with a wider confidence interval (solid line) compared to the opening pressures.

Linear regression of digital pressure readings of the two different devices with a line of identity (dashed line). (A) Comparison of pressure readings during opening pressure (r= 0.97) (B) Comparison of pressure readings during closing pressure (r=0.89). Bland–Altman plots showing the difference between the digital manometer reading and concurrent water column manometer reading during LP. The y-axis shows the mean difference of the two readings, with positive mean representing higher pressure by digital device and x-axis being the pressure reading in cmH2O (A) Plot for opening pressure with the mean difference (dashed line) positive indicating higher pressure for the digital device (B) Plot for closing pressures again with a positive mean difference indicating higher pressures with a wider confidence interval (solid line) compared to the opening pressures.

Discussion

As stated, prior, LPs are critical for the diagnosis and treatment of many disease processes. Pressure monitoring during the procedure is paramount and can be affected by many things such as patient body habitus particularly obesity, mechanical ventilation, pain, anxiety and position. It is very common, as was seen in this study, for patients to receive an LP for likely pressure-related disorders such as idiopathic intracranial hypertension. Accurate OP measurement is imperative in the diagnosis of idiopathic intracranial hypertension, and clinicians treating patients with this disorder must use everything in their armamentarium to ensure accurate diagnosis given the potentially risky treatment options such as CSF shunting. While obtaining an accurate OP or CP is typically straightforward using traditional water column manometry, it is not uncommon for clinicians to experience time constraints due to special populations when trying to obtain pressure readings, such as the pediatric population, comatose patients or patients with delirium.17 Because of these special populations, it is understandable that clinician has sought out a more efficient way to measure pressures during these procedures, hence the growing utilization of digital manometry for the measurement of CSF pressure. As was seen in this study, the use of the digital Compass device can be incredibly efficient, with a reduction in measurement time by approximately 80%. However, clinicians must weigh the time saved to the additional cost of the digital device, which has a list price of approximately $59.95 per device. With over 300,000 LPs being done in the United States annually, the potential yearly cost of $18,000,000 to the already financially burdened United States health-care system must not be taken lightly.18 The use of the water column manometer has been tried and true for decades. The digital device did read higher a majority of the time, with an average of close to 2 cmH2O for OP. While this difference is likely not going to have much clinical impact in most patients, many patients are on the border of normal, as was seen in this study as well with an overall average OP being 21 cmH2O. It is critical that any clinician performing an LP utilizes the most accurate methodology to obtain a pressure; however, this method is not known. The speed of the digital device may be appealing, one must consider the accuracy particularly on borderline pressures since using the device one is straying from common medical practice. The authors do note that the digital device did perform the first pressure for both OPs and CPs. This was done by design as the Compass device would likely utilize the least amount of CSF volume to obtain the pressure, <0.3mL, likely not affecting the water column manometers pressure; however, this may not be entirely true. The pressure may drop with this small decrease in CSF volume, accounting for the difference seen in this study of the pressures. However, one would expect the differences to be close to negligible at the CP measurements where the digital device still reads higher than that of the water column manometer on average making this finding likely a true finding. The pressures were obtained by 2 physicians, and interobserver variability was able to not studied which is an additional limitation to the study. While vast more attention is paid to the OP, the CP remains an important measurement, particularly when treating idiopathic intracranial hypertension.19 When obtaining CPs, air entrapment often occurs, as was seen in every CP performed during this study with a water column manometer, making reading of the CP difficult (see Figure 4 as an example of air entrapped), but is an essential measurement to a therapeutic LP.20 This is likely a significant reason why there is less of a correlation between the two devices during CPs.
Figure 4

Example of air (stars) being entrapped in water column manometer during closing pressure making pressure reading difficult.

Example of air (stars) being entrapped in water column manometer during closing pressure making pressure reading difficult.

Conclusion

CSF pressure measurement during LP is critical for both diagnostic and therapeutic reasons. Particularly, this measurement can be difficult to obtain in certain populations, such as in pediatric patient populations and patients with altered mental status. Digital measurement with an electronic device can help decrease the amount of time it takes to obtain these measurements. Digital CSF pressure measurement with an electronic device obtains measurements highly correlating with standard water column manometer and can help decrease the amount of time it takes to obtain these measurements. However, further studies to confirm accuracy in a clinical setting along with analysis of the economic impact are needed before it becomes the standard of care.
  17 in total

Review 1.  Fluoroscopically Guided Lumbar Puncture.

Authors:  Keith A Cauley
Journal:  AJR Am J Roentgenol       Date:  2015-10       Impact factor: 3.959

2.  CSF pressure assessed by lumbar puncture agrees with intracranial pressure.

Authors:  N Lenfeldt; L-O D Koskinen; A T Bergenheim; J Malm; A Eklund
Journal:  Neurology       Date:  2007-01-09       Impact factor: 9.910

Review 3.  Cerebrospinal fluid pressure in adults.

Authors:  Sarah C M Lee; Christian J Lueck
Journal:  J Neuroophthalmol       Date:  2014-09       Impact factor: 3.042

Review 4.  Lumbar puncture under fluoroscopy guidance: a technical review for radiologists.

Authors:  Can Özütemiz; Jeffrey B Rykken
Journal:  Diagn Interv Radiol       Date:  2019-03       Impact factor: 2.630

Review 5.  Interpretation of lumbar puncture opening pressure measurements in children.

Authors:  Robert A Avery
Journal:  J Neuroophthalmol       Date:  2014-09       Impact factor: 3.042

Review 6.  Difficult Lumbar Puncture: Pitfalls and Tips from the Trenches.

Authors:  P A Hudgins; A J Fountain; P R Chapman; L M Shah
Journal:  AJNR Am J Neuroradiol       Date:  2017-03-16       Impact factor: 3.825

7.  Idiopathic intracranial hypertension in pediatric patients.

Authors:  Nad'a Jirásková; Pavel Rozsíval
Journal:  Clin Ophthalmol       Date:  2008-12

8.  Revised diagnostic criteria for the pseudotumor cerebri syndrome in adults and children.

Authors:  Deborah I Friedman; Grant T Liu; Kathleen B Digre
Journal:  Neurology       Date:  2013-08-21       Impact factor: 9.910

9.  Epidemiology of lumbar punctures in hospitalized patients in the United States.

Authors:  Adrienne Vickers; John P Donnelly; Justin Xavier Moore; Scott R Barnum; Theresa N Schein; Henry E Wang
Journal:  PLoS One       Date:  2018-12-13       Impact factor: 3.240

10.  The effect of whole body position on lumbar cerebrospinal fluid opening pressure.

Authors:  Pasiri Sithinamsuwan; Nakorn Sithinamsuwan; Sirakarn Tejavanija; Chesda Udommongkol; Samart Nidhinandana
Journal:  Cerebrospinal Fluid Res       Date:  2008-07-02
View more
  1 in total

1.  Digital manometry to measure cerebrospinal fluid pressure during lumbar puncture.

Authors:  Robert Heider; Peter G Kranz; Erin Hope Weant; Linda Gray; Timothy J Amrhein
Journal:  Neuroradiol J       Date:  2021-12-24
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.