Feasibility of flat panel detector computed tomography for position assessment of external ventricular drainage.

Philipp von Gottberg1, Marios Psychogios, Gunther Schuetze, Joseph Cohnen, Paul Zwaka, Michael Knauth, Peter Schramm.   

Abstract

BACKGROUND AND
PURPOSE: New angiographic devices with flat panel detectors allow cross-sectional imaging within the angiographic suite. In patients receiving external ventricular drainage (EVD) to manage hydrocephalus following subarachnoid haemorrhage (SAH), these may help evaluating the position of an EVD without moving the patient to a conventional computed tomography (CT) scanner. It could facilitate patients' management in a life-threatening status. This study therefore compares conventional CT with post-interventional flat panel detector angiographic CT (FDCT) referring to the determinability of an accurate EVD position.
MATERIAL AND METHODS: Twenty patients with SAH received FDCT and conventional CT for primary assessment after EVD insertion. Three single-blinded raters compared both modalities and evaluated the image sufficiency for determining the EVD position, EVD tip, intracranial course and whether a contorted drainage tube could be detected.
RESULTS: FDCT was sufficient to detect a correct EVD position in 82.5% of the cases vs. 100% in conventional CT. Regarding the EVD tip, FDCT delivered at least 'good' results in 82.5% vs. 95% in conventional CT data. Determining the EVD intracranial course, FDCT provided at least 'good' data in 92.5% vs. 100% in conventional CT. For detecting tube contortion, FDCT provided at least 'good' results in 70% vs. 98% in conventional CT.
CONCLUSIONS: FDCT is a promising method to determine the correct position of an EVD in patients with SAH. Following a neuroradiological intervention, it facilitates the patients' management and renders additional transfers to conventional CT unnecessary in the majority of cases.

Entities:  

Mesh:

Year:  2013        PMID: 23487292     DOI: 10.5114/ninp.2013.32996

Source DB:  PubMed          Journal:  Neurol Neurochir Pol        ISSN: 0028-3843            Impact factor:   1.621


  2 in total

1.  Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology.

Authors:  Tobias Struffert; Michael Hauer; Rosemarie Banckwitz; Christoph Köhler; Kevin Royalty; Arnd Doerfler
Journal:  Eur Radiol       Date:  2014-04-02       Impact factor: 5.315

Review 2.  Toward the era of a one-stop imaging service using an angiography suite for neurovascular disorders.

Authors:  Sheng-Che Hung; Chung-Jung Lin; Wan-Yuo Guo; Feng-Chi Chang; Chao-Bao Luo; Michael Mu-Huo Teng; Cheng-Yen Chang
Journal:  Biomed Res Int       Date:  2013-05-14       Impact factor: 3.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.