A novel, simplified ex vivo method for measuring water exchange performance of heat and moisture exchangers for tracheostomy application.

Cindy van den Boer1, Sara H Muller, Andrew D Vincent, Klaus Züchner, Michiel W M van den Brekel, Frans J M Hilgers.   

Abstract

BACKGROUND: Breathing through a tracheostomy results in insufficient warming and humidification of inspired air. This loss of air-conditioning can be partially compensated for with the application of a heat and moisture exchanger (HME) over the tracheostomy. In vitro (International Organization for Standardization [ISO] standard 9360-2:2001) and in vivo measurements of the effects of an HME are complex and technically challenging. The aim of this study was to develop a simple method to measure the ex vivo HME performance comparable with previous in vitro and in vivo results.
METHODS: HMEs were weighed at the end of inspiration and at the end of expiration at different breathing volumes. Four HMEs (Atos Medical, Hörby, Sweden) with known in vivo humidity and in vitro water loss values were tested. The associations between weight change, volume, and absolute humidity were determined using both linear and non-linear mixed effects models.
RESULTS: The rating between the 4 HMEs by weighing correlated with previous intra-tracheal measurements (R(2) = 0.98), and the ISO standard (R(2) = 0.77).
CONCLUSIONS: Assessment of the weight change between end of inhalation and end of exhalation is a valid and simple method of measuring the water exchange performance of an HME.

Entities:  

Keywords:  heat and moisture exchanger; humidity; respiration; total laryngectomy; tracheostomy

Mesh:

Year:  2013        PMID: 23481438     DOI: 10.4187/respcare.02369

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

1.  Ex vivo water exchange performance and short-term clinical feasibility assessment of newly developed heat and moisture exchangers for pulmonary rehabilitation after total laryngectomy.

Authors:  Cindy van den Boer; Sara H Muller; Andrew D Vincent; Klaus Züchner; Michiel W M van den Brekel; Frans J M Hilgers
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-05-01       Impact factor: 2.503

2.  Adjustable breathing resistance for laryngectomized patients: Proof of principle in a novel heat and moisture exchanger cassette.

Authors:  Maartje Leemans; Sara H Muller; Maarten J A van Alphen; Wim Vallenduuk; Richard Dirven; Michiel W M van den Brekel
Journal:  Head Neck       Date:  2020-12-08       Impact factor: 3.147

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.