Literature DB >> 2347730

Selection of techniques for orthovoltage radiation therapy.

E C McCullough1.   

Abstract

Selection of appropriate orthovoltage radiation techniques (peak kilovoltage plus total filtration) may be based on the following factors: (a) central axis percent depth dose, (b) exposure rate, (c) sensitivity of backscatter factor to size of a blocked field, (d) differential bone energy absorption, (e) penumbra characteristics, and (f) field uniformity. An analysis of these factors as a function of beam quality (i.e. half-value layer) has been performed and a recommendation of a minimal number of techniques (consistent with widely available techniques) is presented.

Mesh:

Year:  1990        PMID: 2347730     DOI: 10.1016/0360-3016(90)90463-t

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  1 in total

1.  Effect of the bone heterogeneity on the dose prescription in orthovoltage radiotherapy: A Monte Carlo study.

Authors:  James C L Chow; Grigor N Grigorov
Journal:  Rep Pract Oncol Radiother       Date:  2011-11-15
  1 in total