Katsushige Uranishi1, Hirokazu Yamashita, Akiko Okayama, Keigo Yamamoto.
Abstract
Supercritical fluid extraction (SFE) was applied to extraction of pesticides from cereals and pulses. Residues were extracted from homogenized samples mixed with water-absorbent polymer and supercritical carbon dioxide in a stainless steel tube, followed by elution with acetonitrile. Co-extractives were removed by means of mini-column clean-up. Measurement was performed by GC-MS/MS. Calibration was achieved by preparing matrix-matched calibration standards to counteract matrix effects. With the Japanese method validation guideline for pesticide residues as a reference, the method was assessed in 5 agricultural products spiked with 334 pesticides at 0.01 and 0.1 µg/g. Compounds at each level were extracted from 2 samples on 5 separate days. The trueness of the method for 137 pesticides in all samples was 70-120%, and the repeatability and within-run reproducibility were also consistent with the guideline. The trueness of the method for the other 101 pesticides was in the range of 50-70%, though the repeatability and within-run reproducibility were satisfactory. This method is available as a multiresidue analysis method for cereals and pulses.Entities: