Literature DB >> 23466167

Predicting cancellous bone failure during screw insertion.

Karen J Reynolds1, Tammy M Cleek, Aaron A Mohtar, Trevor C Hearn.   

Abstract

Internal fixation of fractures often requires the tightening of bone screws to stabilise fragments. Inadequate application of torque can leave the fracture unstable, while over-tightening results in the stripping of the thread and loss of fixation. The optimal amount of screw torque is specific to each application and in practice is difficult to attain due to the wide variability in bone properties including bone density. The aim of the research presented in this paper is to investigate the relationships between motor torque and screw compression during powered screw insertion, and to evaluate whether the torque during insertion can be used to predict the ultimate failure torque of the bone. A custom test rig was designed and built for bone screw experiments. By inserting cancellous bone screws into synthetic, ovine and human bone specimens, it was established that variations related to bone density could be automatically detected through the effects of the bone on the rotational characteristics of the screw. The torque measured during screw insertion was found to be directly related to bone density and can be used, on its own, as a good predictor of ultimate failure torque of the bone. Crown
Copyright © 2013. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23466167     DOI: 10.1016/j.jbiomech.2013.01.021

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  4 in total

1.  Variable Bone Density of Scaphoid: Importance of Subchondral Screw Placement.

Authors:  Morgan M Swanstrom; Kyle W Morse; Joseph D Lipman; Krystle A Hearns; Michelle G Carlson
Journal:  J Wrist Surg       Date:  2017-08-08

2.  Stripping torques in human bone can be reliably predicted prior to screw insertion with optimum tightness being found between 70% and 80% of the maximum.

Authors:  James W A Fletcher; Ivan Zderic; Boyko Gueorguiev; R Geoff Richards; Harinderjit S Gill; Michael R Whitehouse; Ezio Preatoni
Journal:  Bone Joint Res       Date:  2020-09-03       Impact factor: 5.853

3.  Bond strength evaluation of cyanoacrylate-based adhesives and screws for bone fixation.

Authors:  Juliana de Souza Vieira; Felipe Rychuv Santos; Jessica Vavassori de Freitas; Flares Baratto-Filho; Carla Castiglia Gonzaga; Melissa Rodrigues de Araujo
Journal:  Oral Maxillofac Surg       Date:  2016-01-06

4.  Pullout Strength Predictor: A Machine Learning Approach.

Authors:  Ravi Khatri; Vicky Varghese; Sunil Sharma; Gurunathan Saravana Kumar; Harvinder Singh Chhabra
Journal:  Asian Spine J       Date:  2019-06-03
  4 in total