Literature DB >> 23462677

Effect of MRI acoustic noise on cerebral fludeoxyglucose uptake in simultaneous MR-PET imaging.

Daniel B Chonde1, Nasreddin Abolmaali, Grae Arabasz, Alexander R Guimaraes, Ciprian Catana.   

Abstract

UNLABELLED: Integrated scanners capable of simultaneous positron emission tomography (PET) and magnetic resonance imaging (MRI) data acquisition are now available for human use. Although the scanners' manufacturers have made substantial efforts to understand and minimize the mutual electromagnetic interference between the 2 modalities, the potential physiological inference has not been evaluated. In this study, we have studied the influence of the acoustic noise produced by the magnetic resonance (MR) gradients on brain fludeoxyglucose (FDG) uptake in the Siemens MR-BrainPET prototype. Although particular attention was paid to the primary auditory cortex (PAC), a brain-wide analysis was also performed.
METHODS: The effects of the MR on the PET count rate and image quantification were first investigated in phantoms. Next, 10 healthy volunteers underwent 2 simultaneous FDG-PET/MR scans in the supine position with the FDG injection occurring inside the MR-BrainPET, alternating between a "quiet" (control) environment in which no MR sequences were run during the FDG uptake phase (the first 40 minutes after radiotracer administration) and a "noisy" (test) environment in which MR sequences were run for the entire time. Cortical and subcortical regions of interest were derived from the high-resolution morphological MR data using FreeSurfer. The changes in the FDG uptake in the FreeSurfer-derived regions of interest between the 2 conditions were analyzed from parametric and static PET images, and on a voxel-by-voxel basis using SPM8 and FreeSurfer.
RESULTS: Only minimal to no electromagnetic interference was observed for most of the MR sequences tested, with a maximum drop in count rate of 1.5% and a maximum change in the measured activity of 1.1% in the corresponding images. The region of interest-based analysis showed statistically significant increases in the right PAC in both the parametric (9.13% [4.73%]) and static (4.18% [2.87%]) images. The SPM8 analysis showed no statistically significant clusters in any images when a P < 0.05 (corrected) was used; however, a P < 0.001 (uncorrected) resolved bilateral statistically significant clusters of increased FDG uptake in the area of the PAC for the parametric image (left, 8.37% [1.55%]; right, 8.20% [1.17%]) but only unilateral increase in the static image (left, 8.68% [3.89%]).
CONCLUSIONS: Although the operation of the BrainPET prototype is virtually unaffected by the MR scanner, the acoustic noise produced by the MR gradients causes a focal increase in the FDG uptake in the PAC, which could affect the interpretation of pathological (or brain-activation-related) changes in the FDG uptake in this region if the expected effects are of comparable amplitude.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23462677      PMCID: PMC3775482          DOI: 10.1097/RLI.0b013e3182839fbc

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  33 in total

1.  Event-related fMRI of the auditory cortex.

Authors:  P Belin; R J Zatorre; R Hoge; A C Evans; B Pike
Journal:  Neuroimage       Date:  1999-10       Impact factor: 6.556

2.  Auditory temporal processing: responses to sinusoidally amplitude-modulated tones in the inferior colliculus.

Authors:  B S Krishna; M N Semple
Journal:  J Neurophysiol       Date:  2000-07       Impact factor: 2.714

3.  Continuous transcranial magnetic stimulation during positron emission tomography: a suitable tool for imaging regional excitability of the human cortex.

Authors:  H R Siebner; B Takano; A Peinemann; M Schwaiger; B Conrad; A Drzezga
Journal:  Neuroimage       Date:  2001-10       Impact factor: 6.556

4.  Whole brain segmentation: automated labeling of neuroanatomical structures in the human brain.

Authors:  Bruce Fischl; David H Salat; Evelina Busa; Marilyn Albert; Megan Dieterich; Christian Haselgrove; Andre van der Kouwe; Ron Killiany; David Kennedy; Shuna Klaveness; Albert Montillo; Nikos Makris; Bruce Rosen; Anders M Dale
Journal:  Neuron       Date:  2002-01-31       Impact factor: 17.173

Review 5.  The descending auditory pathway and acousticomotor systems: connections with the inferior colliculus.

Authors:  R F Huffman; O W Henson
Journal:  Brain Res Brain Res Rev       Date:  1990 Sep-Dec

6.  Imaging human intra-cerebral connectivity by PET during TMS.

Authors:  P Fox; R Ingham; M S George; H Mayberg; J Ingham; J Roby; C Martin; P Jerabek
Journal:  Neuroreport       Date:  1997-08-18       Impact factor: 1.837

7.  Transcranial magnetic stimulation during positron emission tomography: a new method for studying connectivity of the human cerebral cortex.

Authors:  T Paus; R Jech; C J Thompson; R Comeau; T Peters; A C Evans
Journal:  J Neurosci       Date:  1997-05-01       Impact factor: 6.167

8.  Nonoxidative glucose consumption during focal physiologic neural activity.

Authors:  P T Fox; M E Raichle; M A Mintun; C Dence
Journal:  Science       Date:  1988-07-22       Impact factor: 47.728

9.  Toward implementing an MRI-based PET attenuation-correction method for neurologic studies on the MR-PET brain prototype.

Authors:  Ciprian Catana; Andre van der Kouwe; Thomas Benner; Christian J Michel; Michael Hamm; Matthias Fenchel; Bruce Fischl; Bruce Rosen; Matthias Schmand; A Gregory Sorensen
Journal:  J Nucl Med       Date:  2010-09       Impact factor: 10.057

10.  An interactive technique for three-dimensional image registration: validation for PET, SPECT, MRI and CT brain studies.

Authors:  U Pietrzyk; K Herholz; G Fink; A Jacobs; R Mielke; I Slansky; M Würker; W D Heiss
Journal:  J Nucl Med       Date:  1994-12       Impact factor: 10.057

View more
  7 in total

1.  A 31-channel MR brain array coil compatible with positron emission tomography.

Authors:  Christin Y Sander; Boris Keil; Daniel B Chonde; Bruce R Rosen; Ciprian Catana; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2014-07-07       Impact factor: 4.668

Review 2.  Principles of Simultaneous PET/MR Imaging.

Authors:  Ciprian Catana
Journal:  Magn Reson Imaging Clin N Am       Date:  2017-05       Impact factor: 2.266

3.  Masamune: a tool for automatic dynamic PET data processing, image reconstruction and integrated PET/MRI data analysis.

Authors:  Daniel B Chonde; David Izquierdo-Garcia; Kevin Chen; Spencer L Bowen; Ciprian Catana
Journal:  EJNMMI Phys       Date:  2014-07

4.  Integrated Positron Emission Tomography/Magnetic Resonance Imaging for Resting-State Functional and Metabolic Imaging in Human Brain: What Is Correlated and What Is Impacted.

Authors:  Yi Shan; Zhe Wang; Shuangshuang Song; Qiaoyi Xue; Qi Ge; Hongwei Yang; Bixiao Cui; Miao Zhang; Yun Zhou; Jie Lu
Journal:  Front Neurosci       Date:  2022-03-02       Impact factor: 4.677

5.  Precuneus and Cingulate Cortex Atrophy and Hypometabolism in Patients with Alzheimer's Disease and Mild Cognitive Impairment: MRI and (18)F-FDG PET Quantitative Analysis Using FreeSurfer.

Authors:  Matthieu Bailly; Christophe Destrieux; Caroline Hommet; Karl Mondon; Jean-Philippe Cottier; Emilie Beaufils; Emilie Vierron; Johnny Vercouillie; Méziane Ibazizene; Thierry Voisin; Pierre Payoux; Louisa Barré; Vincent Camus; Denis Guilloteau; Maria-Joao Ribeiro
Journal:  Biomed Res Int       Date:  2015-06-17       Impact factor: 3.411

Review 6.  Hybrid PET/MR Imaging and Brain Connectivity.

Authors:  Marco Aiello; Carlo Cavaliere; Marco Salvatore
Journal:  Front Neurosci       Date:  2016-03-01       Impact factor: 4.677

7.  Visualization of the auditory pathway in rats with 18F-FDG PET activation studies based on different auditory stimuli and reference conditions including cochlea ablation.

Authors:  Martin Mamach; Mariella Kessler; Jens P Bankstahl; Florian Wilke; Lilli Geworski; Frank M Bengel; Simone Kurt; Georg Berding
Journal:  PLoS One       Date:  2018-10-02       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.