Literature DB >> 23456030

An alternative position for the BIS-Vista montage in frontal approach neurosurgical cases.

Priscilla Nelson1, Jonas A Nelson, Alexander J Chen, W Andrew Kofke.   

Abstract

BACKGROUND: Appropriate placement of the bispectral index (BIS)-vista montage for frontal approach neurosurgical procedures is a neuromonitoring challenge. The standard bifrontal application interferes with the operative field; yet to date, no other placements have demonstrated good agreement. The purpose of our study was to compare the standard BIS montage with an alternate BIS montage across the nasal dorsum for neuromonitoring.
MATERIALS AND METHODS: The authors performed a prospective study, enrolling patients and performing neuromonitoring using both the standard and the alternative montage on each patient. Data from the 2 placements were compared and analyzed using a Bland-Altman analysis, a Scatter plot analysis, and a matched-pair analysis.
RESULTS: Overall, 2567 minutes of data from each montage was collected on 28 subjects. Comparing the overall difference in score, the alternate BIS montage score was, on average, 2.0 (6.2) greater than the standard BIS montage score (P<0.0001). The Bland-Altman analysis revealed a difference in score of -2.0 (95% confidence interval, -14.1, 10.1), with 108/2567 (4.2%) of the values lying outside of the limit of agreement. The scatter plot analysis overall produced a trend line with the equation y=0.94x+0.82, with an R coefficient of 0.82.
CONCLUSIONS: We determined that the nasal montage produces values that have slightly more variability compared with that ideally desired, but the variability is not clinically significant. In cases where the standard BIS-vista montage would interfere with the operative field, an alternative positioning of the BIS montage across the nasal bridge and under the eye can be used.

Entities:  

Mesh:

Year:  2013        PMID: 23456030     DOI: 10.1097/ANA.0b013e31826ca3a0

Source DB:  PubMed          Journal:  J Neurosurg Anesthesiol        ISSN: 0898-4921            Impact factor:   3.956


  6 in total

1.  Intraoperative auditory evoked potential recordings are more reliable at signal detection from different sensor sites on the forehead compared to bispectral index.

Authors:  Yoshiya Ishioka; Shigekazu Sugino; Tomo Hayase; Piotr K Janicki
Journal:  J Clin Monit Comput       Date:  2015-12-12       Impact factor: 2.502

2.  Anesthesia depth monitoring using alternative placement of entropy sensors: a prospective study.

Authors:  David I Vachnadze; Boris A Akselrod; Denis A Guskov; Alevtina V Goncharova
Journal:  J Clin Monit Comput       Date:  2018-12-19       Impact factor: 2.502

3.  Appearance of an electrocardiogram tracing on a bispectral index monitor with plausible BIS values.

Authors:  Takashi Suzuki; Shin-Ya Kimura; Taketo Shinoda
Journal:  J Clin Monit Comput       Date:  2015-03-10       Impact factor: 2.502

4.  Comparison of bispectral index scores from the standard frontal sensor position with those from an alternative mandibular position.

Authors:  Shin Young Lee; Young Sung Kim; Byung Gun Lim; Heezoo Kim; Myoung-Hoon Kong; Il-Ok Lee
Journal:  Korean J Anesthesiol       Date:  2014-04-28

5.  Could this be another story of montage?

Authors:  Sung Uk Choi
Journal:  Korean J Anesthesiol       Date:  2014-04

6.  Intraoperative Frontal Alpha-Band Power Correlates with Preoperative Neurocognitive Function in Older Adults.

Authors:  Charles M Giattino; Jacob E Gardner; Faris M Sbahi; Kenneth C Roberts; Mary Cooter; Eugene Moretti; Jeffrey N Browndyke; Joseph P Mathew; Marty G Woldorff; Miles Berger
Journal:  Front Syst Neurosci       Date:  2017-05-08
  6 in total