Comparison of automated and partial coherence keratometry and resulting choice of toric IOL.

Jit B Ale Magar1, Frank Cunningham, Garry Brian.   

Abstract

PURPOSE: Consistent astigmatism correction with implantation of a toric intraocular lens (IOL) requires accurate preoperative keratometry. This article compares corneal astigmatism determined by an autokeratometer (Topcon KR-7100) and a partial coherence interferometry keratometer (IOLMaster 500) and considers if any discrepancy ultimately alters in final cylindrical power of the toric IOL for implantation.
METHODS: Keratometry of 235 eyes was performed using both instruments. Corneal astigmatism was transformed into vector components J0 and J45 and cylindrical power at the IOL plane calculated. Comparisons were made using paired t test and correlation and Bland-Altman analyses.
RESULTS: Although interinstrument differences for J0 (p = 0.013), J45 (p = 0.012), and toric IOL cylindrical power (p < 0.001) were statistically significant, a high correlation for these (R = 0.96, 0.90, and 0.90, respectively) was observed. IOLMaster tended to overestimate corneal astigmatism by 0.13 (±0.31) diopters and toric IOL cylinder by 0.11 (±0.18) diopters. Difference in calculated toric IOL cylindrical power correlated poorly with corneal curvature (R = 0.007) and astigmatism (R = -0.004).
CONCLUSIONS: The two keratometers were generally concordant in measuring corneal astigmatism. However, the resultant choice of toric IOL cylinder power differed appreciably in 40% of eyes examined. Therefore, postoperative visual outcome with toric IOL implantation may be optimized by a thorough analysis of biometry data before IOL selection, paying special attention to any difference in corneal astigmatism as measured by more than one instrument.

Entities:  

Mesh:

Year:  2013        PMID: 23435222     DOI: 10.1097/OPX.0b013e318287345b

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  4 in total

1.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

2.  Comparison of Total Corneal Astigmatism between IOLMaster and Pentacam.

Authors:  Xiaochun Li; Xiaoguang Cao; Yongzhen Bao
Journal:  Biomed Res Int       Date:  2022-07-08       Impact factor: 3.246

3.  Corneal Astigmatism Measurements Comparison among Ray-Tracing Aberrometry, Partial Coherence Interferometry, and Scheimpflug Imaging System.

Authors:  Yaqin Zhang; Jing Dong; Suhua Zhang; Bin Sun; Xiaoliang Wang; Maolong Tang; Xiaogang Wang
Journal:  J Ophthalmol       Date:  2020-04-01       Impact factor: 1.909

4.  The differences of corneal astigmatism in different populations and its quantitative analysis.

Authors:  Ying Zhang; Hongmei Chen; Yingchun Zhang; Yanzhen Yang
Journal:  Afr Health Sci       Date:  2020-06       Impact factor: 0.927

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.