In-phantom calibration of Selectron-LDR sources.

H Meertens1.   

Abstract

Source strength measurements were performed for cesium-137 spherical sources of nine Selectron-LDR remote afterloading systems in The Netherlands. The mean reference air kerma rate of a set of sources was obtained from air kerma rate measurements in a phantom at a distance of about 5.5 cm from a large number (24 to 30) of sources. The results were compared with the source strengths specified by the manufacturer. Discrepancies between measured and manufacturers data ranging from -2.3% to +3.9% were observed. The frequency distribution of the strength of the sources within a set was measured by the use of a well-type ionisation chamber. The root mean square deviations of the nine source sets ranged from 1.0 to 3.2%. Application of the in-phantom calibration method would improve the precision of source strength measurements and therefore reduce the differences in dose delivery between institutes. The uncertainty of the in-phantom method, which could be estimated by statistical methods, was 0.3% (one standard deviation). That part of the uncertainty, which could not be evaluated by statistical methods, was estimated to be about 1.2% (one effective standard deviation). This latter part in the uncertainty needs further investigation in order to reduce the overall uncertainty.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2343153     DOI: 10.1016/0167-8140(90)90010-t

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  1 in total

Review 1.  Dosimetric audit in brachytherapy.

Authors:  A L Palmer; D A Bradley; A Nisbet
Journal:  Br J Radiol       Date:  2014-05-06       Impact factor: 3.039

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.