Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in healthy eyes.

Carlos García-Resúa1, Hugo Pena-Verdeal, Mercedes Miñones, M Jesus Giraldez, Eva Yebra-Pimentel.   

Abstract

BACKGROUND: Non-contact Tonopachy NT-530P (Nidek Co., LTD) provides intraocular pressure (IOP) and central corneal thickness (CCT) measurements. This study assesses the reliability and repeatability of its IOP measurements in young healthy adult subjects.
METHODS: IOP was determined in the right eye of 64 healthy patients using Tonopachy followed by the Canon TX-10 non-contact and Goldmann applanation (GAT) tonometers. Tonopachy IOP measurements were corrected (Tonopachy-C) or not (Tonopachy-NC) by the instrument for central corneal thickness. Central corneal thickness measurements provided by Tonopachy were also used to correlate (Pearson's coefficient) central corneal thickness with the GAT and Canon TX-10 IOPs. Repeatability of Tonopachy and GAT was assessed in the right eye of 31 subjects in two separate sessions one week apart. Differences between pairs of instruments and between sessions were determined using Bland-Altman plots. The coefficient of repeatability was calculated as the 95% limits of agreement (LoA) of differences between the two sessions.
RESULTS: Tonopachy-NC, Tonopachy-C, Canon TX-10 and the Goldmann tonometers showed a mean IOP of 14.62, 15.64, 15.02 and 14.68 mmHg, respectively. Tonopachy-NC and Canon TX-10 readings did not differ significantly from the Goldmann (p > 0.05), with close agreement with both tonometers (GAT versus Tonopachy-NC: -3.84 to 3.96 mmHg; Goldmann versus Canon TX-10: -4.75 to 4.07 mmHg). Tonopachy-C readings differed significantly from Goldmann (mean difference -0.96 mmHg, p = 0.001, LoA from -5.09 to 3.17). Coefficients of repeatability were ± 3.70, ± 3.14 and ± 3.33 mmHg for GAT, Tonopachy-NC and Tonopachy-C, respectively. Central corneal thickness measured with Tonopachy was 530.42 ± 34.96 μm. There was a significant correlation between central corneal thickness and IOP for all tonometers except Tonopachy-C.
CONCLUSIONS: Reasonable agreement was observed between the Goldmann and Tonopachy. This instrument provides reliable and repeatable IOP measurements not influenced by central corneal thickness. The Tonopachy-NC and Canon TX-10 offer similar reliability compared to the Goldmann applanation tonometer in a young normotensive population.
© 2013 The Authors. Clinical and Experimental Optometry © 2013 Optometrists Association Australia.

Entities:  

Mesh:

Year:  2013        PMID: 23414260     DOI: 10.1111/j.1444-0938.2012.00818.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  6 in total

1.  Correlation analysis between central corneal thickness and intraocular pressure in juveniles in Northern China: the Jinan city eye study.

Authors:  Wen Wei; Zhaoshan Fan; Lihua Wang; Zhiwei Li; Wanzhen Jiao; Yun Li
Journal:  PLoS One       Date:  2014-08-22       Impact factor: 3.240

2.  Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects.

Authors:  Seung Pil Bang; Chong Eun Lee; Yu Cheol Kim
Journal:  BMC Ophthalmol       Date:  2017-11-02       Impact factor: 2.209

3.  Validity of Tono-pachymetry for Measuring Corrected Intraocular Pressure in Non-surgical and Post-photorefractive Keratectomy Eyes.

Authors:  In Kyun Hahn; Jae Yong Kim; Myoung Joon Kim; Hungwon Tchah; Chan Hee Moon
Journal:  Korean J Ophthalmol       Date:  2017-02-02

4.  Reliable intraocular pressure measurement using automated radio-wave telemetry.

Authors:  Eleftherios I Paschalis; Fabiano Cade; Samir Melki; Louis R Pasquale; Claes H Dohlman; Joseph B Ciolino
Journal:  Clin Ophthalmol       Date:  2014-01-07

5.  Comparison of the corrected intraocular pressure by tonopachy with that of Goldmann applanation tonometry in normal and glaucomatous patients.

Authors:  Girish Velis; Srinivasan Kavitha; Nazlee Zebardast; Sabyasachi Sengupta; Rengaraj Venkatesh
Journal:  Indian J Ophthalmol       Date:  2020-04       Impact factor: 1.848

6.  A Comparative Study of the Clinical Efficacy and Safety of Pneumatic Trabeculoplasty and Selective Laser Trabeculoplasty for Patients with Primary Open-Angle Glaucoma or Ocular Hypertension.

Authors:  Yanling Song; Qiujie Song; Yanhui Qin; Jinfeng Xu; Ping Chen; Hong Li; Wenjie Han
Journal:  Med Sci Monit       Date:  2021-12-27
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.