We present the delivery of high energy microsecond pulses through a hollow-core negative-curvature fiber at 2.94 µm. The energy densities delivered far exceed those required for biological tissue manipulation and are of the order of 2300 J/cm(2). Tissue ablation was demonstrated on hard and soft tissue in dry and aqueous conditions with no detrimental effects to the fiber or catastrophic damage to the end facets. The energy is guided in a well confined single mode allowing for a small and controllable focused spot delivered flexibly to the point of operation. Hence, a mechanically and chemically robust alternative to the existing Er:YAG delivery systems is proposed which paves the way for new routes for minimally invasive surgical laser procedures.
We present the delivery of high energy microsecond pulses through a hollow-core negative-curvature fiber at 2.94 µm. The energy densities delivered far exceed those required for biological tissue manipulation and are of the order of 2300 J/cm(2). Tissue ablation was demonstrated on hard and soft tissue in dry and aqueous conditions with no detrimental effects to the fiber or catastrophic damage to the end facets. The energy is guided in a well confined single mode allowing for a small and controllable focused spot delivered flexibly to the point of operation. Hence, a mechanically and chemically robust alternative to the existing Er:YAG delivery systems is proposed which paves the way for new routes for minimally invasive surgical laser procedures.
Entities:
Keywords:
(060.2270) Fiber characterization; (060.2430) Fibers, single-mode; (060.5295) Photonic crystal fibers; (170.1020) Ablation of tissue; (170.3890) Medical optics instrumentation
Er:YAG lasers emitting at a wavelength of 2.94 µm are widely used in medicine. The wide
application of this laser in medical procedures is due to the high absorption coefficient of water
contained in biological tissue of around 12000 cm−1 at this wavelength [1]. This leads to the unique capability of a small penetration depth
and therefore high ablation rates, with high precision and a minimal heat-affected zone. This is key
to reducing collateral damage, or cell death in surrounding tissue, during surgical procedures.
Additional generic advantages of laser based procedures are that no pressure is applied, reducing
the pain for the patient, e.g., in dental drilling [2], and
the cut geometry is not limited by the drill/scalpel geometry but is dictated by the focused spot
size which generally can be significantly smaller than traditional surgical tools.In order to harness the advantages of lasers for medical procedures the laser light should
ideally be delivered via a flexible guide from the laser source to the patient, allowing complete
freedom for the surgeon. Several delivery systems are already in use to deliver the necessary powers
to manipulate (i.e., cut and drill) biological/human tissue. The two most common ones are
articulated arms and large core multimode optical fibers. The articulated arm is a system of rigid
tubes connected by movable joints with built in mirrors and an interchangeable hand piece. The hand
piece contains optics to generate a particular beam profile on the tissue to be processed. There has
been continued improvement of these systems and early issues, like beam wandering when the arm was
moved, have been largely overcome. However, such arms can never be completely flexible in 3D space,
restricting the surgeon. Generally, the size of the hand piece prevents the use of the articulated
arm in minimally invasive surgery or in combination with endoscopy, although by attaching a short
fiber to the hand piece some of these space restrictions are eliminated.The alternative to the articulated arm would be to use a robust and flexible fiber delivery
system. The main drive for fiber delivery is the flexibility it gives the users and the small
physical size and weight which would therefore drastically increase the usability of these surgical
lasers. There are a number of solid core fibers operating at the wavelength of 2.94 µm, that
have been investigated for this application, based on chalcogenides [3,4], GeO2 [4] or sapphire [5]. All these fiber types were
demonstrated to deliver high power multimode laser beams.In solid core fibers the laser induced damage threshold (LIDT) of the fiber material imposes a
limit on the power handling capabilities. In order to circumvent such issues hollow core fibers have
been developed. The main types of these fibers guide by Bragg reflection, or by internal reflection
at a dielectric coated metallic interface in the case of the leaky tube waveguide [6-8]. However, these
types of waveguides are limited in length due to the fabrication process. Hollow core photonic
crystal fibers (HC-PCF) have been shown to guide in this wavelength region with power handling
capabilities suited to laser surgery [9]. Since in these
waveguide designs the light is mainly guided inside the hollow core of the fiber, they typically
have a higher damage threshold than solid core fibers [10-13].In this paper the application of a new fiber design based on the principles of hollow core
microstructured fibers to surgery is described. Due to the shape of the central hollow core
geometry, as shown Fig. 1
[14], this fiber is referred to as a Negative
Curvature Fiber (NCF). The fiber is fabricated [14] from
fused silica using a conventional stack and draw technique similar to that used for HC-PCF. The
material is Suprasil F300 which has a bulk attenuation of ~50 dB/m at 2.94 µm (see Fig. 2
) [9]. However, as light is mainly confined to the
hollow core the influence of this high absorption is significantly reduced allowing low-loss fibers
to be made in this wavelength region [10,15]. Silica is a desirable material for medical applications as it
has several advantages, e.g., it can withstand high temperatures, it is bio-inert, mechanically and
chemically robust and it has been extensively characterized for fiber drawing which makes it easy to
handle. The broadband guidance of the fiber has been previously described, with low attenuation
ranges from 2 µm to 2.5 µm and from 2.8 µm to 3.8 µm [14]. The lowest attenuation achieved was 34 dB/km at 3.05
µm. In this paper the fiber presented in [14] was used
to demonstrate the high energy laser pulse delivery of 2.94 µm radiation for use as a
flexible delivery system for surgical lasers.
Fig. 1
SEM picture of the negative curvature fiber used in these experiments.
Fig. 2
Absorption spectrum of silica (Suprasil F300) in the mid IR [9].
SEM picture of the negative curvature fiber used in these experiments.Absorption spectrum of silica (Suprasil F300) in the mid IR [9].In order to be considered as an alternative solution for flexible delivery of Er:YAG laser light
the fiber must reliably deliver pulses of sufficient energy to ablate a wide range of biological
tissue. The typical thresholds required as reported in the literature are shown in Table 1
. As described in this paper, the energy density transmitted through this fiber far
exceeds the thresholds required for hard and soft biological tissue ablation.
Table 1
Ablation thresholds for different biological tissues
Rep rate [Hz]
Pulse length [µs]
Tissue type
Threshold [J/cm2]
Refs.
2
250
Human dental enamel
35
[16]
7-10
250
Human skin
1.6
[17]
1.7
250
Pig retina
1
[18]
1
100-5000
Human dentine
2.69-3.66
[19,20]
5
NA
Pig skin (vitro)
3.6-5.6
[21]
2
200
Guinea pig skin
0.6-1.5
[22]
2. Laser and optics
The laser used in our experiments was an Impex High Tech ERB 15 laser. The operating wavelength
is 2.937 µm and the pulse-length is 225 µs FWHM (Fig.
3a
), with an M2 of ~2.5 at a repetition rate of ~15 Hz. The spatial profile of the
laser output has a donut shape (Fig. 3b); the low resolution
in the image is a result of the relatively large pixel size of the IR camera used (each pixel is 50
× 50 µm).
Fig. 3
(a) Temporal profile of the laser pulse (b) Spatial beam profile of the laser showing a donut
shaped beam.
(a) Temporal profile of the laser pulse (b) Spatial beam profile of the laser showing a donut
shaped beam.The laser light was coupled into the fiber using a lens of focal length f = 100 mm, giving a
focused spot size diameter of 67 µm and focused cone angle of 70 mrad.This optical arrangement was found to give the best coupling efficiency but due to the mismatch
between the laser mode and the fiber mode field profile (which is a Gaussian-like single mode) the
maximum coupling efficiency achieved was around 35%. This coupling efficiency was independent from
the incident energy level. It is possible to improve the laser beam profile, i.e., achieve a smaller
M2 value by inserting an aperture in the laser cavity but this resulted in a significant
loss of output power and consequently the donut mode was used to excite the fiber for all
experiments. It is assumed that modes that have not been coupled into the core mode couple into
cladding modes. These cladding modes are gradually absorbed over a short length in the outer fused
silica cladding and the protective polyimide jacket on the outside of the fiber. The absorption of
fused silica at this wavelength is around 50 dB/m [9]. Since
the cladding modes are leaking away over a distance, the localized energy deposition in the fused
silica cladding is reduced and no measures had to be taken to cool the fiber, and importantly, no
damage from the absorbed energy was observed.
3. Fiber
The guidance of this fiber can be explained by the Anti-resonant reflecting optical waveguide
(ARROW) principle [23]. As described by Litchinitser et al.
[23], wavelengths which are in resonance with the core wall
cannot be confined in the core but leak away thought the wall, resulting in a high attenuation.
However, frequencies that are anti-resonant with the wall cannot propagate within it and will be
more confined inside the core. The two interfaces of wall and air can be described as a
Fabry-Perot-like resonator. Anti-resonant wavelengths experience a low leakage through the wall and
hence a lower attenuation as a result of destructive interference in the Fabry-Perot resonator.Fabrication of the fiber was performed using a standard stack and draw method, as described in
reference [14]. The core diameter (defined as the shortest
distance inside the core) is 94 µm, and the angle of the full acceptance cone is 60 mrad (NA
of 0.03).
3.1. Attenuation
To measure the attenuation of the fiber at 2.94 µm a cut back measurement was carried out,
using a tunable laser as the optical source. The total attenuation in these NCF fibers is a
combination of confinement and absorptive losses and bend induced losses. We therefore took great
care to ensure the fiber was bent in a known and defined manner. The ends of the fiber were held
straight and the fiber was bent with a diameter of 50 cm. The bent part of the fiber consisted of
5.5 full circles. The initial length of the fiber was 9.55 m and a 3 m long piece was cut back,
which corresponds to two full circles. The attenuation measured for this configuration, where 5.5 m
of the fiber is coiled with a diameter of 0.5 m and 0.5 m straight fiber at either end, was measured
to be 0.183 dB/m ± 0.05 dB/m.Figure 4
shows the additional losses that accrue if the fiber is bent by 180° with diameters
from 50 cm down to 5 cm for 1.23 m long fiber piece. As can be seen there are no significant
additional loses if the bend diameter is >30 cm, which is sufficient for many applications. The
bending losses are most likely from core mode(s) coupling into cladding modes which are strongly
absorbed. From the same graph it is noticeable that the attenuation for shorter wavelengths for a
bend diameter of 20 cm is higher than for 10 cm. The details of this unexpected
behavior—which is presumably due to coupling to other modes in the fiber—are yet to be
investigated. Output beam mode profiles for shorter wavelengths could give an indication of the
potential mode coupling at tighter bends, and it is our intention to address this in future
work.
Fig. 4
Additional losses due to fiber bend in dB for a 1.23 m long fiber piece. The bend diameter is
given for one 180° bend.
Additional losses due to fiber bend in dB for a 1.23 m long fiber piece. The bend diameter is
given for one 180° bend.Typical bend diameters needed for endoscopy applications, such as endourology are of the order of
15 cm [24]. Such bend radii are possible with solid core
sapphire fibers but for flexibility need core diameters below 600 μm which therefore imposes
a damage threshold on the fiber [25]. State-of-the-art
multimode solid core fiber delivery systems based on fluoride glasses can achieve a bending diameter
of 20 cm with very weak dependence on the bend diameter [26]
and having an attenuation of ~0.2 dB/m. Alternatively, multimode hollow core waveguides have shown a
higher dependence between the bend diameter and losses. The attenuation for a hollow core waveguide
with a 750 µm core has been reported as 1.9 dB/m (90° bend) and 2.9 dB/m (180°
bend) when the bend diameter was 30 cm [27]. However, it is
envisaged that in order to develop novel minimally invasive procedures it may be necessary have bend
diameters in the order of 10’s mm.
3.2. Fiber output beam profile
An investigation of the fiber output beam profile was carried out by moving a second fiber
transversal to the negative curvature fiber (NCF). The “measuring” fiber used was a
previously reported HC-PCF with a band gap at 2.94 µm, core diameter of 24 µm and a
single mode core profile [15]. This measurement, using a
smaller-core fiber for signal collection, significantly increase the fidelity of the convolution
which was present in the earlier similar measurement [14]
using two identical fibers. The output power of the HC-PCF was measured with a pyroelectric detector
(Coherent P5-01). The position of the fiber was controlled with a LVDT (Linear Variable Differential
Transformer) (Tesa Tronic TTD20). The relative output power versus the fiber position is shown in
Fig. 5
. The fiber was bent during this test to a bend diameter of ~50 cm.
Fig. 5
Fiber output beam profile for a bent NCF measured by moving a HC- PCF transversally relative to
the NCF. The NCF was bent with a diameter of ~50 cm over a length of 80 cm.
Fiber output beam profile for a bent NCF measured by moving a HC- PCF transversally relative to
the NCF. The NCF was bent with a diameter of ~50 cm over a length of 80 cm.As can be seen from the measurement, the output beam profile is single mode like and close to a
Gaussian beam profile. The beam profiles are identical for straight and bent fiber configurations.
From this measurement a 1/e2 mode field diameter of 78 µm can be calculated which
is 83% of the core diameter and demonstrates good confinement in the hollow core.The M2 of the fiber-delivered beam was measured (according to ISO Standard 11146
[12]) as 1.4 which is in line with the Gaussian like output
beam profile shown in Fig. 5.Modeling was done to estimate the overlap between the laser and the fiber mode in order to
indicate the maximum coupling efficiency achievable. This was done by a calculation of the overlap
integral, I, of the amplitudes of the laser mode,
φ, and the fiber mode,
φ, whereas described in [28]. As only a 1D spatial prolife of the
fiber output was measured for this calculation the fiber mode profile at input was assumed to be a
Gaussian beam profile with the same 1/e2 beam diameter as the fiber (78 µm). The
overlap of this assumed fiber mode and the laser beam mode as calculated from Eq. (1) is 55.8%. As can be seen from the measurement, the
output beam profile is single mode like and close to a Gaussian beam profile. Based on the profile
in Fig. 5 the fiber output has a 98.7% overlap with a perfect
Gaussian of the same 1/e2 beam width.
3.3. Beam propagation
It is envisaged that the fiber could be used in contact mode or with a standoff from the tissue
in practical medical procedures. Consequently an investigation of the far field propagation was
carried out to assess the spatial beam profile at certain distances from the fiber end facet. To
assess the spatial beam profile the beam was incident on a ceramic surface and an image was captured
with a mid-infrared camera (Electrophysics PV320 L2E). The fiber used for these images had a length
of 6.48 m and was curled up in a loop with a diameter of 50 cm over a length of 5.5 m. The distances
between the fiber end and the reflective surface were 10, 20, 50 and 100 mm, respectively to
replicate what could be conceived as practical working distances for surgery. The false color images
are shown in Fig. 6
. No significant change in the general beam profile can be detected other than the expected
increase in beam diameter for distances from 10 mm to 100 mm.
Fig. 6
Far field beam profile at different distances (a) 10, (b) 20, (c) 50, and (d) 100 mm) from the
fiber end. The fiber length is 6.48 m.
Far field beam profile at different distances (a) 10, (b) 20, (c) 50, and (d) 100 mm) from the
fiber end. The fiber length is 6.48 m.
3.4. High energy microsecond pulse delivery
The maximum output energy delivered through the NCF was achieved when the full power output of
the laser was incident at the input of the fiber, resulting in an energy at the output of the fiber
of 195 ± 1 mJ for a 33 cm length of fiber and 54 ± 4 mJ for a 9.88 m length. In both
cases the fiber was bent to a diameter of 50 cm over a length of 80 cm. These pulse energies
translate to energy densities of 2300 J/cm2 for the short length and 764 J/cm2
for the long length immediately at the end of the fiber, respectively. As shown in Table 1, human dental enamel has the highest ablation threshold
of 35 J/cm2, and it is clear that even for the longer 10 m fiber the delivered energy
density exceeds it by a factor of >21. It should also be noted that the stated values for the
power delivery capability of the fiber do not represent the limits of the power handling capability
of the fiber as the experiment was limited by the available power from the laser source. Both the
input and output facets of the fiber were undamaged during the transmission experiments. It is
likely that given a laser with higher output energy and/or better beam quality, significantly higher
pulse energies could be delivered.To test the practicality of the fiber it was used free handed to ablate material. During these
tests the fiber was bent down to a diameter of <10 cm. No damage to the fiber could be detected
although the power output dropped as expected due to increased bend loss, however it was still
sufficient enough to ablate the material (porcine bone).
4. Encapsulation of the fiber with an endtip
One practical issue in using hollow-core fibers for medical applications is the possibility of
contaminating the core with debris and liquids (e.g., blood or tissue fragments) particularly if the
fiber is used in an endoscopic procedure. Therefore an encapsulation of at least one fiber end is
necessary. Our approach is the sapphire endtip and a schematic is shown in Fig. 7
. As a demonstrator this endtip was mounted onto the fiber using a heat shrinking tube (Fig. 8
).
Fig. 7
schematic and dimensions of the sapphire endtip.
Fig. 8
Endtip mounted onto the fiber using a heat shrinking tube.
schematic and dimensions of the sapphire endtip.Endtip mounted onto the fiber using a heat shrinking tube.The distance between the fiber end facet and the outer surface of the sapphire window in the
endtip (the contact point of laser irradiation and tissue) is 0.5 mm. In order to avoid damage to
the sapphire window of the endtip the energy was restricted for these tests as it still provided
sufficient energy for tissue ablation. The maximum output energy measured at this point using a 2 m
long fiber piece was 30 mJ. Using this value and the divergence half angle of 36 mrad gives an
energy density of >500 J/cm2 at the contact point. Again this energy density far
exceeds the ablation thresholds necessary for biological tissue yet is well within the operating
capability of the device.A cross section through the spatial beam profile directly at the end of the tip is shown in Fig. 9
. Although there is some change in the beam profile compared to the previous image (Fig. 5) it is not significantly different and the beam is still
single-mode-like.
Fig. 9
Beam profile at the endtip’s outer surface.
Beam profile at the endtip’s outer surface.This figure represents the beam profile incident on the tissue if the fiber is used in contact
mode, where the endtip would be in direct contact with the tissue. The far field output beam profile
at distances of 10, 20, 50 and 100 mm are also shown, again to demonstrate conditions expected for
practical surgical applications, Figs. 10(b)
and 10(c). Compared to the beam profile without an
endtip (Fig. 6) some artifacts around the central peak
position are visible and the beam has more structure to it. These are most likely due the inside
surface of the sapphire window not having an optical polish and the possibility of some
contamination during construction of the endtip.
Fig. 10
Far-field profile of the fiber with endtip. Distance from the endtip to the reflective surface:
(a) 10, (b) 20, (c) 50, and (d) 100 mm.
Far-field profile of the fiber with endtip. Distance from the endtip to the reflective surface:
(a) 10, (b) 20, (c) 50, and (d) 100 mm.
5. Tissue ablation
In order to demonstrate that the delivered power is sufficient to ablate hard and soft biological
tissue, a sample of porcine tissue (bone and muscle) was used. The fiber length used for the
ablation experiment was 2 m and the output end was sealed by the endtip shown in Fig. 8. The output power was 30 mJ which produced a fluence of
>500 J/cm2 at the output surface of the endtip. This energy density was sufficient to
ablate the porcine muscle and bone as can be seen in Fig. 11
. At these fluencies the ablation depth for a single shot on bone was approximately 200
µm and the heat-affected zone (HAZ) was about 70 µm (see Fig. 11(b). By adapting the laser parameter, the HAZ can be minimized, however the
investigation of these parameters was not the scope of these experiments. In these trials the fiber
was hand held, as opposed to being fixed, to simulate how it may be used in practice. A side effect
of this is that the pulses are not delivered perfectly normal to the tissue. The fiber was used in
contact mode and at different distances and also ablation was carried out in aqueous conditions with
the fiber and tissue completely immersed in water (Fig. 12
). These results clearly show that the fiber is capable of delivering pulses of the necessary
energy for tissue ablation and shows that the fiber and endtip configuration is robust and can be
handled in a practical manner.
Fig. 11
Tissue ablation results: (a) porcine bone; (b) cross-section through hole in porcine bone showing
ablation depth with single shot, ablation depth is 265 µm; (c) porcine muscle; (d) tissue
ablation of porcine muscle with a number of shots being distributed over the surface.
Fig. 12
Endtip immersed in water.
Tissue ablation results: (a) porcine bone; (b) cross-section through hole in porcine bone showing
ablation depth with single shot, ablation depth is 265 µm; (c) porcine muscle; (d) tissue
ablation of porcine muscle with a number of shots being distributed over the surface.Endtip immersed in water.Autoclaving was performed on the fiber. The fiber was sealed on both ends using an arc-fusion
splicer to fuse the ends by localized melting. The conditions in the autoclave were 121°C at
15 psi for 15 min. No degradation of the fiber could be determined after repeating this procedure 3
times. At these temperatures a softening of the acrylic fiber jacket can be expected which could
lead to localized weakening points. However for a medical device a polyimide jacket can be used,
which withstands much higher temperatures. The endtip was not tested in the autoclave, however
previous tests [15] have shown that the bond between the
sapphire tube and sapphire rod is hermetically sealed and stable to temperatures over
1000°C.To demonstrate the capabilities of the system of fiber and endtip the ablation of ovine bone is
presented in Fig. 13(a)
(Media 1). The
width of the cut is around 300 µm with a depth of 220 µm. The square is 2x2 mm in size
and the distance between tissue and endtip is around 5 mm. Additionally the ablation under water is
shown in Fig. 13(b) (Media
2). As can be seen the tip is fully submerged into the water.
Parameters in both experiments were kept the same.
Fig. 13
Screenshots from Media 1 and
Media 2. (a)
Ablation of ovine bone in air. Square dimensions are 2x2 mm (Media
1) (b) Ablation of ovine bone under water
(Media 2).
Screenshots from Media 1 and
Media 2. (a)
Ablation of ovine bone in air. Square dimensions are 2x2 mm (Media
1) (b) Ablation of ovine bone under water
(Media 2).
6. Conclusion
A novel delivery system for Er:YAG laser radiation is presented which has the potential to enable
new minimally invasive surgical procedures. The flexible fiber is fabricated from silica and guides
the light inside a hollow core by the ARROW principle which increases the damage threshold
significantly, compared to a solid core fiber, and allows the effect of the high absorption of
silica at this wavelength to be negated. The output beam profile is single-mode-like, leading to a
significant advantage, in terms of controllability and stability for the delivered energy, compared
to other large core fibers. The performance in terms of delivered fluence (up to 2300
J/cm2) far exceeds the thresholds needed for biological tissue. This has been practically
demonstrated by showing that hard and soft tissue could be ablated. A practical approach for
encapsulation of the fiber has been proposed which demonstrates that a practical surgical device
could be developed. This system shows a promising alternative to the existing delivery systems
already used in medicine and other high power applications at 2.94 µm and paves the way for
novel minimally invasive surgical procedures.
Authors: J Shephard; W Macpherson; R Maier; J Jones; D Hand; M Mohebbi; A George; P Roberts; J Knight Journal: Opt Express Date: 2005-09-05 Impact factor: 3.894
Authors: Stefan Stübinger; vet Brigitte von Rechenberg; Hans-Florian Zeilhofer; Robert Sader; Constantin Landes Journal: Lasers Surg Med Date: 2007-08 Impact factor: 4.025