Reliability of the Dinamap non-invasive monitor in the measurement of blood pressure of ill Asian newborns.

F Chia1, A T Ang, T W Wong, K W Tan, K P Fung, J Lee, K Khin.   

Abstract

Four hundred thirty-one paired sets of readings of systolic and diastolic blood pressure and 438 paired sets of readings of mean arterial BP from 49 ill newborns, including 21 very low birth weight infants, were analyzed for the extent and pattern of agreement and the linear relationship between the Dinamap oscillometric monitor and the direct intraarterial blood pressure readings. Agreement between the two methods was measured by the intraclass correlation, whereas the linear relationship was assessed by the product-moment correlation. The intraclass correlations for systolic, diastolic and mean blood pressures were 0.696, 0.766, and 0.781, respectively. The product-moment correlations for systolic, diastolic and mean blood pressures were 0.706, 0.768, and 0.786, respectively. BP measurements by the Dinamap monitor showed reasonably close agreement to those obtained by the intraarterial mean arterial pressure ranges above 40 mmHg. For mean arterial pressure of 40 mmHg and lower, BP readings by the Dinamap monitor tended to be higher than those obtained by the intraarterial method. These findings appeared to be consistent regardless of the birth weight of the newborn.

Entities:  

Mesh:

Year:  1990        PMID: 2340688     DOI: 10.1177/000992289002900502

Source DB:  PubMed          Journal:  Clin Pediatr (Phila)        ISSN: 0009-9228            Impact factor:   1.168


  5 in total

1.  Accuracy of non-invasive blood pressure monitoring in very preterm infants.

Authors:  Kai König; Dan M Casalaz; Emily J Burke; Andrew Watkins
Journal:  Intensive Care Med       Date:  2012-03-06       Impact factor: 17.440

2.  A proposal for scientific validation of instruments for indirect blood pressure measurement at rest, during exercise, and in critical care.

Authors:  Y Iyriboz; C M Hearon
Journal:  J Clin Monit       Date:  1994-05

3.  Systolic blood pressure in babies of less than 32 weeks gestation in the first year of life. Northern Neonatal Nursing Initiative.

Authors: 
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-01       Impact factor: 5.747

4.  Use of pulse oximetry for blood pressure measurement after cardiac surgery.

Authors:  A J Movius; S L Bratton; G K Sorensen
Journal:  Arch Dis Child       Date:  1998-05       Impact factor: 3.791

5.  The accuracy of blood pressure measured by arterial line and non-invasive cuff in critically ill children.

Authors:  Rachel Joffe; Jonathan Duff; Gonzalo Garcia Guerra; Jodie Pugh; Ari R Joffe
Journal:  Crit Care       Date:  2016-06-08       Impact factor: 9.097

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.