Literature DB >> 23380412

Agreement between Pentacam and IOL master in patients undergoing toric IOL implantation.

Bradford W Lee1, Anat Galor, William J Feuer, Bozorgmehr Pouyeh, Jesse S Pelletier, Pravin K Vaddavalli, Benjamin T Lemelman, Craig See, Sonia H Yoo.   

Abstract

PURPOSE: To investigate inter-device agreement between the Pentacam (Optikgerate GmbH, Wetzlar, Germany) and IOLMaster (Carl Zeiss Meditec, Jena, Germany) in measuring corneal power, cylinder, and axis of astigmatism.
METHODS: Retrospective case series reviewing the status of eyes after toric IOL implantation to evaluate inter-device agreement in measuring corneal power, cylinder, and axis of astigmatism.
RESULTS: Forty-nine eyes from 41 patients were evaluated. Agreement for corneal power and cylinder were deemed good and moderate, with 95% limits of agreement of -1.02 to +1.13 and -1.37 to +1.09 diopters (D), respectively. Agreement for axis of astigmatism was only fair, with a mean absolute difference of 8.9 degrees (range: 0.3 to 37.7 degrees, standard deviation: 8.79 degrees). Thirty percent of eyes had a 10-degree or greater difference in axis measurements, and 13% of eyes had a 20-degree or greater difference in measurements. Analysis by multivariate logistic regression showed higher cylinder powers associated with discrepant corneal power measurements (odds ratio: 3.46, P = .02), and increasing age was associated with discrepant axis measurements (odds ratio: 1.09, P = .03). Overall, there was no significant difference in accuracy predicting postoperative spherical equivalent refraction.
CONCLUSIONS: Pentacam and IOLMaster agree well for corneal power but less so for cylinder and axis of astigmatism and thus cannot be used interchangeably. Copyright 2013, SLACK Incorporated.

Entities:  

Mesh:

Year:  2013        PMID: 23380412     DOI: 10.3928/1081597X-20130117-06

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  7 in total

1.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

2.  Comparison of Total Corneal Astigmatism between IOLMaster and Pentacam.

Authors:  Xiaochun Li; Xiaoguang Cao; Yongzhen Bao
Journal:  Biomed Res Int       Date:  2022-07-08       Impact factor: 3.246

Review 3.  Optimizing outcomes with toric intraocular lenses.

Authors:  Manpreet Kaur; Farin Shaikh; Ruchita Falera; Jeewan S Titiyal
Journal:  Indian J Ophthalmol       Date:  2017-12       Impact factor: 1.848

4.  Astigmatic Outcomes of Single, Non-Paired Intrastromal Limbal Relaxing Incisions During Femtosecond Laser-Assisted Cataract Surgery Based on a Custom Nomogram.

Authors:  Carter W Lim; Sohel Somani; Hannah H Chiu; Raj Maini; Eric S Tam
Journal:  Clin Ophthalmol       Date:  2020-04-22

5.  Corneal Astigmatism Measurements Comparison among Ray-Tracing Aberrometry, Partial Coherence Interferometry, and Scheimpflug Imaging System.

Authors:  Yaqin Zhang; Jing Dong; Suhua Zhang; Bin Sun; Xiaoliang Wang; Maolong Tang; Xiaogang Wang
Journal:  J Ophthalmol       Date:  2020-04-01       Impact factor: 1.909

6.  Calculation of the Real Corneal Refractive Power after Photorefractive Keratectomy Using Pentacam, When Only the Preoperative Refractive Error is Known.

Authors:  Maddalena De Bernardo; Maria Borrelli; Roberto Imparato; Nicola Rosa
Journal:  J Ophthalmol       Date:  2020-04-01       Impact factor: 1.909

7.  Prediction of residual astigmatism in cataract surgery at different diameter zones using optical biometry measurement.

Authors:  Yin-Hsi Chang; Christy Pu; Ken-Kuo Lin; Jiahn-Shing Lee; Chiun-Ho Hou
Journal:  Sci Rep       Date:  2022-03-11       Impact factor: 4.379

  7 in total