A novel method to position an endotracheal tube at the correct depth using an infrared sensor stylet.

Yeongtak Song1, Jaehoon Oh, Youngjoon Chee, Taeho Lim, Hyunggoo Kang, Youngsuk Cho.   

Abstract

PURPOSE: This study is a feasibility assessment to determine the ability of novice users to utilize an infrared (IR) sensor stylet as a guide to position the tip of the endotracheal tube (ETT) 40 mm proximal to the carina in the swine trachea.
METHODS: We developed a stylet system using an IR sensor attached to the tip of a stylet to facilitate measuring the distance of the ETT from the carina. The indicator lamp of the IR sensor system turns on through calibration when the ETT tip arrives at a point 20 mm proximal to the carina. In order to place the ETT tip 40 mm (middle of 20-60 mm) from the carina after the indicator lamp turns on, the operator uses the ETT marker to withdraw the ETT 20 mm. For this study, five fresh swine airways were used in random order after drawing lots, and ten novices were recruited to attempt the procedure ten times on each swine's airway.
RESULTS: Five hundred endotracheal intubations were performed. For the target distance of 40 mm from the ETT tip to the carina, the mean (standard deviation) of the total data set was 37.9 (3.5) mm; all data were within a 20-60 mm range (500/500), and 98.2% (491/500) of the attempts fell within the 30-50 mm range.
CONCLUSIONS: The IR sensor stylet system can facilitate correct positioning of the ETT tip at an appropriate depth above the carina in the swine trachea. Evaluation of the IR sensor stylet methodology in human subjects is warranted.

Entities:  

Mesh:

Year:  2013        PMID: 23370979     DOI: 10.1007/s12630-013-9898-6

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  2 in total

1.  Design and Control of a Mechatronic Tracheostomy Tube for Automated Tracheal Suctioning.

Authors:  Thanh Nho Do; Tian En Timothy Seah; Soo Jay Phee
Journal:  IEEE Trans Biomed Eng       Date:  2015-10-15       Impact factor: 4.538

2.  Continuous monitoring of endotracheal tube positioning: closer to the sangreal?

Authors:  Ahmed El Kalioubie; Saad Nseir
Journal:  J Clin Monit Comput       Date:  2015-02       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.