BACKGROUND: Contrast-induced nephropathy (CIN) is an important cause of iatrogenic morbidity and mortality. The amount of contrast delivered has a major effect on CIN and is operator-dependent. A few studies suggested that the use of automated contrast injection systems is associated with reduced contrast volume. It is unknown whether this is true when smaller amounts of contrast are used and how this is affected by training.
METHODS: Volume of contrast media was measured in 1358 consecutive patients undergoing diagnostic catheterization and percutaneous coronary intervention (PCI) from January 31 to May 31, 2011. Patients were allocated to manual stopcock-manifold contrast injection (1052 patients) or automated contrast injection (306 patients).
RESULTS: No significant difference in contrast volume use was found between manual and automated contrast injection systems, respectively: diagnostic catheterization, 72 ± 40 mL vs 96 ± 63 mL (P = 0.08); diagnostic catheterization with left ventricular angiography, 98 ± 40 mL vs 95 ± 35 mL (P = 0.51); PCI, 206 ± 82 mL vs 205 ± 90 mL (P = 0.84); diagnostic catheterization and PCI, 264 ± 83 mL vs 253 ± 93 mL (P = 0.51). No significant difference in CIN incidence, according to contrast injection systems, was found among patients receiving PCI (manual 9.8% vs automated 7.4%, P = 0.43). Using smaller sized catheters during diagnostic procedures was associated with injection of smaller amounts of contrast (P < 0.0001).
CONCLUSIONS: The use of automated contrast injection for diagnostic catheterization and PCI is not associated with reduced contrast volume as compared with manual injection. The use of smaller calibre catheters might reduce contrast volume.
Copyright © 2013 Canadian Cardiovascular Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23352426     DOI: 10.1016/j.cjca.2012.11.023

Source DB:  PubMed          Journal:  Can J Cardiol        ISSN: 0828-282X            Impact factor:   5.223


  3 in total

1.  Truncus Bicaroticus With Arteria Lusoria: A Rare Combination of Aortic Root Anatomy Complicating Cardiac Catheterization.

Authors:  Capt Roy Norris; Cpt Andrew Wilson; Maj Charles Lin
Journal:  Fed Pract       Date:  2021-02

Review 2.  Meta-analysis of the effect of automated contrast injection devices versus manual injection and contrast volume on risk of contrast-induced nephropathy.

Authors:  Kristopher D Minsinger; Hayah M Kassis; Clay A Block; Mandeep Sidhu; Jeremiah R Brown
Journal:  Am J Cardiol       Date:  2013-10-03       Impact factor: 2.778

3.  A prospective study of contrast preservation using ultra-low contrast delivery technique versus standard automated contrast injector system in coronary procedures.

Authors:  Adam Stys; Maheedhar Gedela; Udit Bhatnagar; Marian Petrasko; Hazem Dawoud; Tadeusz Malinski; Tomasz Stys
Journal:  Indian Heart J       Date:  2019-09-10
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.