Undersensing by an ICD due to alternans of the ventricular electrogram.

Frederic E Van Heuverswyn1, Liesbeth Timmers, Roland X Stroobandt, S Serge Barold.   

Abstract

Alternans of the ventricular electrogram (VEGM) during ventricular tachycardia (VT) is a rare cause of ventricular undersensing by an implantable cardioverter-defibrillator (ICD). This report describes a patient with a St. Jude ICD who exhibited sustained monomorphic VT associated with surface QRS alternans, alternating cycle lengths, alternans of the VEGM causing intermittent undersensing of the smaller component, and intermittent 2:1 counting of ventricular intervals during 1:1 sensing in response to the ICD detection algorithm. VEGM undersensing was corrected noninvasively simply by programming the threshold start from 62.5% to 50% which increased the sensitivity based on the amplitude of the VEGM. This maneuver did not affect the satisfactory and stable defibrillation threshold. ©2012, Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 23347031      PMCID: PMC6932347          DOI: 10.1111/j.1542-474X.2012.00521.x

Source DB:  PubMed          Journal:  Ann Noninvasive Electrocardiol        ISSN: 1082-720X            Impact factor:   1.468


  9 in total

1.  Implantable cardioverter defibrillator sensing failure due to endocardial R wave electrical alternans.

Authors:  Krit Jongnarangsin; Theofanie Mela
Journal:  J Cardiovasc Electrophysiol       Date:  2002-07

2.  Alternans in QRS amplitude during ventricular tachycardia.

Authors:  Philippe Maury; Jacques Metzger
Journal:  Pacing Clin Electrophysiol       Date:  2002-02       Impact factor: 1.976

3.  Reentrant fascicular tachycardia with cycle length alternans: insights into the tachycardia mechanism and origin.

Authors:  Y T Tai; P C Fong; C P Lau; W H Chow; C H Cheng
Journal:  Pacing Clin Electrophysiol       Date:  1990-07       Impact factor: 1.976

4.  Alternation of QRS morphology and effect of radiofrequency ablation in idiopathic ventricular tachycardia.

Authors:  T Washizuka; Y Aizawa; M Chinushi; N Naitoh; T Miyajima; Y Kusano; H Kitazawa; H Uchiyama; K Takahashi; A Shibata
Journal:  Pacing Clin Electrophysiol       Date:  1995-01       Impact factor: 1.976

5.  Endocardial mapping of ventricular tachycardia in the intact human ventricle. III. Evidence of multiuse reentry with spontaneous and induced block in portions of reentrant path complex.

Authors:  E Downar; J Saito; J C Doig; T C Chen; E Sevaptsidis; S Masse; S Kimber; L Mickleborough; L Harris
Journal:  J Am Coll Cardiol       Date:  1995-06       Impact factor: 24.094

6.  Multiform ventricular tachycardia.

Authors:  D L Ross; A W Hamer; J K Vohra; J G Sloman; D Hunt
Journal:  Pacing Clin Electrophysiol       Date:  1980-01       Impact factor: 1.976

7.  QRS alternans during idiopathic ventricular tachycardia originating from the right coronary cusp of the aorta.

Authors:  Takumi Yamada; Hugh Thomas McElderry; Harish Doppalapudi; George Neal Kay
Journal:  Europace       Date:  2010-01       Impact factor: 5.214

8.  Electrical alternans in wide complex tachycardias.

Authors:  M S Kremers; J M Miller; M E Josephson
Journal:  Am J Cardiol       Date:  1985-08-01       Impact factor: 2.778

9.  Two ventricular tachycardias with cycle length and QRS alternans: insights into the mechanism from mapping and ablation of the tachycardias.

Authors:  Steven Kalbfleisch; David Hart; Raul Weiss
Journal:  Pacing Clin Electrophysiol       Date:  2009-09-10       Impact factor: 1.976

  9 in total
  1 in total

Review 1.  Sensing and detection in Medtronic implantable cardioverter defibrillators.

Authors:  Mark L Brown; Charles D Swerdlow
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.