Reproducibility of ECG-gated ultrasound diameter assessment of small abdominal aortic aneurysms.

K Bredahl1, N Eldrup, C Meyer, J E Eiberg, H Sillesen.   

Abstract

OBJECTIVE: No standardised ultrasound procedure to obtain reliable growth estimates for abdominal aortic aneurysms (AAA) is currently available. We investigated the feasibility and reproducibility of a novel approach controlling for a combination of vessel wall delineation and cardiac cycle variation.
DESIGN: Prospective comparative study.
METHODS: Consecutive patients (N = 27) with an AAA, attending their 6-month control as part of a medical treatment trial, were scanned twice by two ultrasound operators. Then, all ultrasound recordings were transferred to a core facility and analysed by a third person. The AAA diameter was determined in four different ways: from the leading edge of adventitia on the anterior wall to either the leading edge of the adventitia (method A) or leading edge of the intima (method B) on the posterior wall, with both measurements performed in systole and diastole. RESULT: Inter-operator reproducibility was ± 3 mm for all methods applied. There was no difference in outcome between methods A and B; likewise, end-diastolic measurement did not improve reproducibility in preference to peak-systolic measurement.
CONCLUSION: The use of a standardised ultrasound protocol including ECG-gating and subsequent off-line reading with minute calliper placement reduces variability. This may be of use in developing protocols to better detect even small AAA growth rates during clinical trials.
Copyright © 2012 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23332308     DOI: 10.1016/j.ejvs.2012.12.010

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  7 in total

1.  Characterization of laser ultrasound source signals in biological tissues for imaging applications.

Authors:  Jonathan R Fincke; Charles M Wynn; Rob Haupt; Xiang Zhang; Diego Rivera; Brian Anthony
Journal:  J Biomed Opt       Date:  2018-12       Impact factor: 3.170

2.  Reliability of aortic aneurysm screening measurements.

Authors:  Rachel Weston Smith; Tony Evans; Stephen Wolstenhulme
Journal:  Ultrasound       Date:  2013-12-13

3.  Agreement between objective and subjective assessment of image quality in ultrasound abdominal aortic aneurism screening.

Authors:  S Wolstenhulme; A G Davies; C Keeble; S Moore; J A Evans
Journal:  Br J Radiol       Date:  2014-12-11       Impact factor: 3.039

4.  Ultrasound diagnostics of the abdominal aorta: English version.

Authors:  W Schäberle; L Leyerer; W Schierling; K Pfister
Journal:  Gefasschirurgie       Date:  2015

5.  Abdominal ultrasound-scanning versus non-contrast computed tomography as screening method for abdominal aortic aneurysm - a validation study from the randomized DANCAVAS study.

Authors:  Mads Liisberg; Axel C Diederichsen; Jes S Lindholt
Journal:  BMC Med Imaging       Date:  2017-02-14       Impact factor: 1.930

Review 6.  Systematic approach towards reliable estimation of abdominal aortic aneurysm size by ultrasound imaging and CT.

Authors:  S M Tomee; C A Meijer; D A Kies; S le Cessie; M N J M Wasser; J Golledge; J F Hamming; J H N Lindeman
Journal:  BJS Open       Date:  2021-01-08

7.  The reproducibility of measuring maximum abdominal aortic aneurysm diameter from ultrasound images.

Authors:  Evan O Matthews; Jenna Pinchbeck; Kylie Elmore; Rhondda E Jones; Joseph V Moxon; Jonathan Golledge
Journal:  Ultrasound J       Date:  2021-02-26
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.