This article proposes analytic performance goals for five quality indicators: precision, trueness, linearity, detection limits, and consistency across instruments and time. We defined our goals using methods linked to clinical practice data. Goals for desirable precision and trueness are based on biological variation. Linearity goals are related to total error recommendations. Detection limit goals are derived from 0.1 percentile of patient values. Goals for consistency are derived from the variability of distributions of patient test values. Data were collected and evaluated for each of these quality indicators for 46 chemistry tests measured on the Roche cobas 8000 analyzer.Entities:
Mesh:
Year: 2012 PMID: 23331729 DOI: 10.1016/j.cll.2012.11.005