Literature DB >> 23319720

Qualitative and quantitative assessment of wrist MRI at 3.0T: comparison between isotropic 3D turbo spin echo and isotropic 3D fast field echo and 2D turbo spin echo.

Jee Young Jung1, Young Cheol Yoon, Jin Young Jung, Bong-Keun Choe.   

Abstract

BACKGROUND: Isotropic three-dimensional (3D) magnetic resonance imaging (MRI) has been applied to various joints. However, comparison for image quality between isotropic 3D MRI and two-dimensional (2D) turbo spin echo (TSE) sequence of the wrist at a 3T MR system has not been investigated.
PURPOSE: To compare the image quality of isotropic 3D MRI including TSE intermediate-weighted (VISTA) sequence and fast field echo (FFE) sequence with 2D TSE intermediate-weighted sequence of the wrist joint at 3.0 T.
MATERIAL AND METHODS: MRI was performed in 10 wrists of 10 healthy volunteers with isotropic 3D sequences (VISTA and FFE) and 2D TSE intermediate-weighted sequences at 3.0 T. The signal-to-noise ratio (SNR) was obtained by imaging phantom and noise-only image. Contrast ratios (CRs) were calculated between fluid and cartilage, triangular fibrocartilage complex (TFCC), and the scapholunate ligament. Two radiologists independently assessed the visibility of TFCC, carpal ligaments, cartilage, tendons and nerves with a four-point grading scale. Statistical analysis to compare CRs (one way ANOVA with a Tukey test) and grades of visibility (Kruskal-Wallis test) between three sequences and those for inter-observer agreement (kappa analysis) were performed.
RESULTS: The SNR of 2D TSE (46.26) was higher than those of VISTA (23.34) and 3D FFE (19.41). CRs were superior in 2D TSE than VISTA (P = 0.02) for fluid-cartilage and in 2D TSE than 3D FFE (P < 0.01) for fluid-TFCC. The visibility was best in 2D TSE (P < 0.01) for TFCC and in VISTA (P = 0.01) for scapholunate ligament. The visibility was better in 2D TSE and 3D FFE (P = 0.04) for cartilage and in VISTA than 3D FFE (P < 0.01) for TFCC. The inter-observer agreement for the visibility of anatomic structures was moderate or substantial.
CONCLUSION: Image quality of 2D TSE was superior to isotropic 3D MR imaging for cartilage, and TFCC. 3D FFE has better visibility for cartilage than VISTA and VISTA has superior visibility for TFCC to 3D FFE and the visibility for scapholunate ligament was best on VISTA.

Entities:  

Mesh:

Year:  2013        PMID: 23319720     DOI: 10.1258/ar.2012.120475

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  7 in total

Review 1.  Advanced imaging of the scapholunate ligamentous complex.

Authors:  Maryam Shahabpour; Barbara Staelens; Luc Van Overstraeten; Michel De Maeseneer; Cedric Boulet; Johan De Mey; Thierry Scheerlinck
Journal:  Skeletal Radiol       Date:  2015-07-30       Impact factor: 2.199

2.  Diagnostic performance of wrist magnetic resonance (MR) arthrography: comparison of three-dimensional isotropic T1 weighted fast spin-echo MR arthrography and two-dimensional MR arthrography.

Authors:  Eunsun Oh; Young C Yoon; Min J Park
Journal:  Br J Radiol       Date:  2017-02-15       Impact factor: 3.039

3.  Comparison of two-dimensional fast spin echo T2 weighted sequences and three-dimensional volume isotropic T2 weighted fast spin echo (VISTA) MRI in the evaluation of triangular fibrocartilage of the wrist.

Authors:  Hee Jin Park; So Yeon Lee; Kyung A Kang; Eun Young Kim; Hun Kyu Shin; Se Jin Park; Jai Hyung Park; Eugene Kim
Journal:  Br J Radiol       Date:  2018-01-05       Impact factor: 3.039

4.  Using an Office-Based, Dedicated Extremity MRI Scanner for Depicting Important Structures in Common Wrist Pathologies: A Pilot Comparison with a Conventional MRI Scanner.

Authors:  Qi Yin; Radu A Manoliu; Jayant R Kichari; Marco J P F Ritt
Journal:  J Wrist Surg       Date:  2020-10-14

5.  Evaluation of diagnostic value and T2-weighted three-dimensional isotropic turbo spin-echo (3D-SPACE) image quality in comparison with T2-weighted two-dimensional turbo spin-echo (2D-TSE) sequences in lumbar spine MR imaging.

Authors:  Jomleh Hossein; Faeghi Fariborz; Rasteh Mehrnaz; Rafiei Babak
Journal:  Eur J Radiol Open       Date:  2018-12-28

6.  Visualization of wrist anatomy-a comparison between 7T and 3T MRI.

Authors:  Simon Götestrand; Anders Björkman; Isabella M Björkman-Burtscher; Ingvar Kristiansson; Elenya Aksyuk; Pawel Szaro; Karin Markenroth Bloch; Mats Geijer
Journal:  Eur Radiol       Date:  2021-08-11       Impact factor: 5.315

Review 7.  High-resolution measurements of the multilayer ultra-structure of articular cartilage and their translational potential.

Authors:  Bo He; Jian Ping Wu; Thomas Brett Kirk; John A Carrino; Chuan Xiang; Jiake Xu
Journal:  Arthritis Res Ther       Date:  2014-03-12       Impact factor: 5.156

  7 in total