Literature DB >> 23314168

Accuracy of determining hemoglobin level using occlusion spectroscopy in patients with severe gastrointestinal bleeding.

Julien Coquin1, Amandine Bertarrex, Antoine Dewitte, Laurent Lefèvre, Olivier Joannes-Boyau, Catherine Fleureau, Stéphane Winnock, Sébastien Leuillet, Gérard Janvier, Alexandre Ouattara.   

Abstract

BACKGROUND: In patients presenting with severe hemorrhage, the authors conducted an equivalence trial that compared noninvasive occlusion spectroscopy and the capillary blood method to determine hemoglobin level.
METHODS: This prospective observational study included patients admitted to their intensive care unit for gastrointestinal bleeding. A ring-shaped sensor, connected to a NBM-200MP (OrSense, Nes Ziona, Israel), was fitted onto the patient's thumb to intermittently measure hemoglobin (SotHb). During the first 24 h after admission, venous hemoglobin level, considered as the reference method, was determined at the laboratory every 8 h and was compared to SotHb and the capillary blood method. The primary endpoint was the proportion of inaccurate measurements, defined as greater than 15% difference compared with reference values or their unavailability for any technical reason.
RESULTS: The study was scheduled to include 68 patients but was stopped prematurely after an interim analysis of 34 patients. The proportion of inaccuracies revealed that SotHb could not be considered equivalent to the capillary blood method (47% [95% CI, 43-51] and 24% [95% CI, 20-28]). Considering venous hemoglobin level as a reference method, the mean biases for SotHb (n = 133) and the capillary blood method (n = 135) were, respectively, -0.4 ± 2.0 and 0.8 ± 1.2 g/dl (P < 0.05). SotHb was associated with an increased incidence of failed transfusion. The inaccuracy of SotHb tended to be increased in patients receiving vasopressor agents.
CONCLUSIONS: Noninvasive determination of hemoglobin level based on occlusion spectroscopy lacks accuracy in patients presenting with severe gastrointestinal bleeding and cannot be considered equivalent to the capillary-based method. This inaccuracy seems to be moderately influenced by the infusion of vasopressor agents.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23314168     DOI: 10.1097/ALN.0b013e3182833fcc

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  4 in total

Review 1.  Factors affecting hemoglobin measurement.

Authors:  Lauren Berkow
Journal:  J Clin Monit Comput       Date:  2013-03-26       Impact factor: 2.502

2.  Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients.

Authors:  Werner Baulig; Burkhardt Seifert; Donat R Spahn; Oliver M Theusinger
Journal:  J Clin Monit Comput       Date:  2015-12-19       Impact factor: 2.502

3.  Non-invasive detection of severe neutropenia in chemotherapy patients by optical imaging of nailfold microcirculation.

Authors:  Aurélien Bourquard; Alberto Pablo-Trinidad; Ian Butterworth; Álvaro Sánchez-Ferro; Carolina Cerrato; Karem Humala; Marta Fabra Urdiola; Candice Del Rio; Betsy Valles; Jason M Tucker-Schwartz; Elizabeth S Lee; Benjamin J Vakoc; Timothy P Padera; María J Ledesma-Carbayo; Yi-Bin Chen; Ephraim P Hochberg; Martha L Gray; Carlos Castro-González
Journal:  Sci Rep       Date:  2018-03-28       Impact factor: 4.379

4.  Evaluation of new non-invasive & conventional invasive methods of haemoglobin estimation in blood donors.

Authors:  Diptiranjan Rout; Suchet Sachdev; Neelam Marwaha
Journal:  Indian J Med Res       Date:  2019-06       Impact factor: 2.375

  4 in total