Literature DB >> 23313184

Computer modeling for the prediction of thoracic aortic stent graft collapse.

Salvatore Pasta1, Jae-Sung Cho, Onur Dur, Kerem Pekkan, David A Vorp.   

Abstract

OBJECTIVE: To assess the biomechanical implications of excessive stent protrusion into the aortic arch in relation to thoracic aortic stent graft (TASG) collapse by simulating the structural load and quantifying the fluid dynamics on the TASG wall protrusion extended into a model arch.
METHODS: One-way coupled fluid-solid interaction analyses were performed to investigate the flow-induced hemodynamic and structural loads exerted on the proximal protrusion of the TASG and aortic wall reconstructed from a patient who underwent traumatic thoracic aortic injury repair. Mechanical properties of a Gore TAG thoracic endoprosthesis (W. L. Gore and Assoc, Flagstaff, Ariz) were assessed via experimental radial compression testing and incorporated into the computational modeling. The TASG wall protrusion geometry was characterized by the protrusion extension (PE) and by the angle (θ) between the TASG and the lesser curvature of the aorta. The effect of θ was explored with the following four models with PE fixed at 1.1 cm: θ = 10 degrees, 20 degrees, 30 degrees, and 40 degrees. The effect of PE was evaluated with the following four models with θ fixed at 10 degrees: PE = 1.1 cm, 1.4 cm, 1.7 cm and 2.0 cm.
RESULTS: The presence of TASG wall protrusion into the aortic arch resulted in the formation of swirling, complex flow regions in the proximal luminal surface of the endograft. High PE values (PE = 2.0 cm) led to a markedly reduced left subclavian flow rate (0.27 L/min), low systolic perfusion pressure (98 mm Hg), and peak systolic TASG diameter reduction (2 mm). The transmural pressure load across the TASG was maximum for the model with the highest PE and θ, 15.2 mm Hg for the model with PE = 2.0 cm and θ = 10 degrees, and 11.6 mm Hg for PE = 1.1 cm and θ = 40 degrees.
CONCLUSIONS: The findings of this study suggest that increased PE imparts an apparent risk of distal end-organ malperfusion and proximal hypertension and that both increased PE and θ lead to a markedly increased transmural pressure across the TASG wall, a load that would portend TASG collapse. Patient-specific computational modeling may allow for identification of patients with high risk of TASG collapse and guide preventive intervention. CLINICAL RELEVANCE: A potentially devastating complication that may occur after endovascular repair of traumatic thoracicaortic injuries is stent graft collapse. Although usually asymptomatic, stent graft collapse may be accompanied by adverse hemodynamic consequences. Numerous anatomic and device-related factors contribute to the development of collapse, but predictive factors have not yet been clearly defined. In the present study, we assessed the relevant hemodynamics and solid mechanics underlying stent graft collapse using a computational fluid-structure interaction framework of stent graft malapposition. Our findings suggest that both increased stent graft angle and extension into the aortic arch lead to a markedly increased transmural pressure across the stent graft wall, portending collapse. Patient-specific computational modeling may allow for identification of patients at high risk for collapse and aid in planning for an additional, prophylactic intervention to avert its occurrence.
Copyright © 2013 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23313184     DOI: 10.1016/j.jvs.2012.09.063

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  8 in total

1.  Changes of time-attenuation curve blood flow parameters in patients with and without carotid stenosis.

Authors:  C-J Lin; F-C Chang; W-Y Guo; S-C Hung; C-B Luo; J Beilner; M Kowarschik; W-F Chu
Journal:  AJNR Am J Neuroradiol       Date:  2015-02-26       Impact factor: 3.825

2.  Effects of longitudinal pre-stretch on the mechanics of human aorta before and after thoracic endovascular aortic repair (TEVAR) in trauma patients.

Authors:  Anastasia Desyatova; Jason MacTaggart; Alexey Kamenskiy
Journal:  Biomech Model Mechanobiol       Date:  2019-09-05

3.  Haemodynamic predictors of a penetrating atherosclerotic ulcer rupture using fluid-structure interaction analysis.

Authors:  Giuseppe D'Ancona; Andrea Amaducci; Antonino Rinaudo; Salvatore Pasta; Fabrizio Follis; Michele Pilato; Roberto Baglini
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-06-04

4.  Tetralogy of Fallot Surgical Repair: Shunt Configurations, Ductus Arteriosus and the Circle of Willis.

Authors:  Senol Piskin; Gozde Unal; Ahmet Arnaz; Tayyar Sarioglu; Kerem Pekkan
Journal:  Cardiovasc Eng Technol       Date:  2017-04-05       Impact factor: 2.495

Review 5.  Structural modelling of the cardiovascular system.

Authors:  Benjamin Owen; Nicholas Bojdo; Andrey Jivkov; Bernard Keavney; Alistair Revell
Journal:  Biomech Model Mechanobiol       Date:  2018-06-18

6.  Identification of geometric and mechanical factors predictive of bird-beak configuration in thoracic endovascular aortic repair using computational models of stent graft deployment.

Authors:  Negin Shahbazian; David A Romero; Thomas L Forbes; Cristina H Amon
Journal:  JVS Vasc Sci       Date:  2022-06-24

Review 7.  Thoracoabdominal aortic aneurysm repair: current endovascular perspectives.

Authors:  Nathan Orr; David Minion; Joseph L Bobadilla
Journal:  Vasc Health Risk Manag       Date:  2014-08-19

8.  Computational Pre-surgical Planning of Arterial Patch Reconstruction: Parametric Limits and In Vitro Validation.

Authors:  S Samaneh Lashkarinia; Senol Piskin; Tijen A Bozkaya; Ece Salihoglu; Can Yerebakan; Kerem Pekkan
Journal:  Ann Biomed Eng       Date:  2018-05-14       Impact factor: 3.934

  8 in total