| Literature DB >> 23311454 |
XiangKui Wan1, Kanghui Yan, Dehan Luo, Yanjun Zeng.
Abstract
BACKGROUND: T-wave alternans (TWA) provides a noninvasive and clinically useful marker for the risk of sudden cardiac death (SCD). Current most widely used TWA detection algorithms work in two different domains: time and frequency. The disadvantage of the spectral analytical techniques is that they treat the alternans signal as a stationary wave with a constant amplitude and a phase. They cannot detect non-stationary characteristics of the signal. The temporal domain methods are sensitive to the alignment of the T-waves. In this study, we sought to develop a robust combined algorithm (CA) to assess T-wave alternans, which can qualitatively detect and quantitatively measure TWA in time domain.Entities:
Mesh:
Year: 2013 PMID: 23311454 PMCID: PMC3554460 DOI: 10.1186/1749-8090-8-7
Source DB: PubMed Journal: J Cardiothorac Surg ISSN: 1749-8090 Impact factor: 1.637
Figure 1Time-frequency feature extraction of T wave. (a) a rhythm; (b) wavelet temporal-frequency figure of the rhythm (c) total energy of the T wave.
Qualitative detection experiment data
| 20 | 31 | 62.0 | 10 | 37 | 74 |
| 25 | 47 | 94.0 | 20 | 42 | 84 |
| 30 | 50 | 100 | 50 | 49 | 98 |
| 35 | 50 | 100 | 100 | 50 | 100 |
| 40 | 50 | 100 | 200 | 50 | 100 |
| total | 228 | 91.2 | 228 | 91.2 |
quantitative measurement TWA (unit: μV)
| 10 | 8.01 | 7.21 | 7.53 |
| 20 | 15.95 | 14.67 | 15.35 |
| 50 | 38.30 | 37.02 | 37.59 |
| 100 | 76.19 | 74.12 | 75.21 |
| 200 | 150.71 | 149.09 | 150.04 |
TWA detection results of CA and SM
| 10 | 7.5 | 3.4 | 25 | 66 |
| 20 | 15 | 6.9 | 25 | 65.5 |
| 50 | 37.8 | 17.4 | 24.4 | 65.2 |
| 100 | 75.2 | 34.8 | 24.8 | 65.2 |
| 200 | 152 | 69.6 | 24 | 65.2 |
Test results of partial samples
| ECG1 | Subject 1 | 11(37%) | 4.92 | 2.41 |
| ECG2 | Subject 2 | 7(23%) | 3.21 | 1. 23 |
| 31 | SCDHD | 19(63%) | 18.72 | 9.09 |
| 36 | SCDHD | 10(33%) | 10.12 | 4.61 |
| 41 | SCDHD | 23(77%) | 5.12 | 2.39 |
| 45 | SCDHD | 25(83%) | 14.96 | 7.10 |
| 46 | SCDHD | 9(30%) | 9.12 | 4.71 |
| e0119 | ST-TD | 17(56%) | 16.78 | 7.31 |
| e0121 | ST-TD | 22(73%) | 21.82 | 9.83 |
| e0125 | ST-TD | 12(40%) | 17.85 | 8.11 |
| e0601 | ST-TD | 0 | 0 | 0 |
| e1302 | ST-TD | 10(33%) | 7.12 | 3.45 |
Figure 2Measured TWA amplitude of CA vs. SM.