Calibrating page sized Gafchromic EBT3 films.

W Crijns1, F Maes, U A van der Heide, F Van den Heuvel.   

Abstract

PURPOSE: The purpose is the development of a novel calibration method for dosimetry with Gafchromic EBT3 films. The method should be applicable for pretreatment verification of volumetric modulated arc, and intensity modulated radiotherapy. Because the exposed area on film can be large for such treatments, lateral scan errors must be taken into account. The correction for the lateral scan effect is obtained from the calibration data itself.
METHODS: In this work, the film measurements were modeled using their relative scan values (Transmittance, T). Inside the transmittance domain a linear combination and a parabolic lateral scan correction described the observed transmittance values. The linear combination model, combined a monomer transmittance state (T(0)) and a polymer transmittance state (T(∞)) of the film. The dose domain was associated with the observed effects in the transmittance domain through a rational calibration function. On the calibration film only simple static fields were applied and page sized films were used for calibration and measurements (treatment verification). Four different calibration setups were considered and compared with respect to dose estimation accuracy. The first (I) used a calibration table from 32 regions of interest (ROIs) spread on 4 calibration films, the second (II) used 16 ROIs spread on 2 calibration films, the third (III), and fourth (IV) used 8 ROIs spread on a single calibration film. The calibration tables of the setups I, II, and IV contained eight dose levels delivered to different positions on the films, while for setup III only four dose levels were applied. Validation was performed by irradiating film strips with known doses at two different time points over the course of a week. Accuracy of the dose response and the lateral effect correction was estimated using the dose difference and the root mean squared error (RMSE), respectively.
RESULTS: A calibration based on two films was the optimal balance between cost effectiveness and dosimetric accuracy. The validation resulted in dose errors of 1%-2% for the two different time points, with a maximal absolute dose error around 0.05 Gy. The lateral correction reduced the RMSE values on the sides of the film to the RMSE values at the center of the film.
CONCLUSIONS: EBT3 Gafchromic films were calibrated for large field dosimetry with a limited number of page sized films and simple static calibration fields. The transmittance was modeled as a linear combination of two transmittance states, and associated with dose using a rational calibration function. Additionally, the lateral scan effect was resolved in the calibration function itself. This allows the use of page sized films. Only two calibration films were required to estimate both the dose and the lateral response. The calibration films were used over the course of a week, with residual dose errors ≤2% or ≤0.05 Gy.

Mesh:

Year:  2013        PMID: 23298104     DOI: 10.1118/1.4771960

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Iterative metal artifact reduction improves dose calculation accuracy : Phantom study with dental implants.

Authors:  Manuel Maerz; Pia Mittermair; Andreas Krauss; Oliver Koelbl; Barbara Dobler
Journal:  Strahlenther Onkol       Date:  2016-03-11       Impact factor: 3.621

2.  In vivo dosimetry using Gafchromic films during pelvic intraoperative electron radiation therapy (IOERT).

Authors:  Filipa Costa; Sandra Sarmento; Dora Gomes; Helena Magalhães; Rosário Arrais; Graciete Moreira; Maria Fátima Cruz; José Pedro Silva; Lúcio Santos; Olga Sousa
Journal:  Br J Radiol       Date:  2016-05-18       Impact factor: 3.039

3.  Measurement of percentage dose at the surface for a 6 MV photon beam.

Authors:  O O Galván De la Cruz; M A Rodríguez-Ávila; T Rivera-Montalvo; O A García Garduño
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-18

4.  Gafchromic EBT3 film dosimetry in electron beams - energy dependence and improved film read-out.

Authors:  Petri Sipilä; Jarkko Ojala; Sampsa Kaijaluoto; Ilkka Jokelainen; Antti Kosunen
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

5.  Single-fraction spine SBRT end-to-end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom.

Authors:  John J Gallo; Isaac Kaufman; Rachel Powell; Shalini Pandya; Archana Somnay; Todd Bossenberger; Ezequiel Ramirez; Robert Reynolds; Timothy Solberg; Jay Burmeister
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

6.  Expired EBT3 Films' Sensitivity for the Measurement of X-ray and UV Radiation: An Optical Analysis.

Authors:  Hamad Yahia Abu Mhanna; Ahmad Fairuz Omar; Yasmin Md Radzi; Hanan Fawaz Akhdar; Haytham Al Ewaidat
Journal:  Materials (Basel)       Date:  2022-04-15       Impact factor: 3.748

7.  Experimental investigation of dynamic real-time rotation-including dose reconstruction during prostate tracking radiotherapy.

Authors:  Casper Gammelmark Muurholm; Thomas Ravkilde; Robin De Roover; Simon Skouboe; Rune Hansen; Wouter Crijns; Tom Depuydt; Per R Poulsen
Journal:  Med Phys       Date:  2022-04-25       Impact factor: 4.506

8.  A study on rectal dose measurement in phantom and in vivo using Gafchromic EBT3 film in IMRT and CyberKnife treatments of carcinoma of prostate.

Authors:  K Ganapathy; P G G Kurup; V Murali; M Muthukumaran; S Balaji Subramanian; J Velmurugan
Journal:  J Med Phys       Date:  2013-07

9.  Production of patient-specific electron beam aperture cut-outs using a low-cost, multi-purpose 3D printer.

Authors:  Steven Michiels; Bram Mangelschots; Robin De Roover; Cédric Devroye; Tom Depuydt
Journal:  J Appl Clin Med Phys       Date:  2018-07-26       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.