| Literature DB >> 23293449 |
Hasin A Azhari1, Frank Hensley, Wilhelm Schütte, Golam A Zakaria.
Abstract
Before clinical use of a brachytherapy source, regulations or recommendations by medical physics societies require an independent measurement of its air kerma strength by a qualified medical physicist. Currently, in addition to Ir-192, also HDR-Co-60 sources are increasingly coming into operation. However, the existing dosimetry protocols do not provide any guidelines for Co-60 sources. The purpose of this work was therefore to compare air kerma rate measurements as recommended by different dosimetry protocols for Ir-192 HDR sources and to test their applicability to Co-60 sources. Dosimetric verification of HDR afterloading source specification was performed according to three protocols, DIN 6809-2 (1993) in combination with DGMP-Report 13 (2006), IAEA-TECDOC-1274 (2002) and AAPM Report 41 (1993) for the nuclides Ir-192 and Co-60. Measurements of the sources reference air kerma rate were performed with 3 different methods (with a cylindrical chamber both in a solid phantom and in free air, and with a well chamber) and evaluated using all three protocols for each type of source and method of measurement. The measurements with all protocols and methods show deviations from the certified specification smaller than about 1.2% for Ir-192 and 2.5% for Co-60-Sources. The measurements with the well chamber showed the lowest deviations from the certificate value. Air kerma rate measurements for Co-60 HDR sources using the existing protocols are possible with accuracy sufficient to verify source calibration as provided by the source certificate. However, extension of the protocols by correction factors for measurement with Co-60 sources would be helpful.Entities:
Keywords: Air kerma rate measurement for Ir-192; Co-60; brachytherapy; dosimetry
Year: 2012 PMID: 23293449 PMCID: PMC3532746 DOI: 10.4103/0971-6203.103603
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Measurements performed for each protocol, (*) mark denotes: procedure/method is not available for specific protocol
Figure 1Shows the experimental set up in the Krieger phantom with the Bebig After loader and a Co-60-Source
Parameters for the determination of the Air kerma rate from the reading
Methods for the determination of the factors kS and kp using different voltage value in different methods
Figure 2Shows the experimental set up in “free in air” with the Nucletron After loader and an Ir-192-Source with shield block placed between chamber and the source
Calibration factors used for the different protocols
Figure 3Shows the measurements with the well chamber with the Nucletron After loader and an Ir-192-Source
Figure 4Deviations from the certified calibration of Co-60 and Ir 192 sources in three protocols. The error bars indicate the variation of repeated measurements
Deviations of measured Air kerma rate from the source certificate value and measured relative uncertainties (in brackets) for different protocols
Overview of influence quantities and their contributions to total uncertainty. Where not stated otherwise, uncertainties are taken from[10]