Literature DB >> 23293449

Dosimetric verification of source strength for HDR afterloading units with Ir-192 and Co-60 photon sources: Comparison of three different international protocols.

Hasin A Azhari1, Frank Hensley, Wilhelm Schütte, Golam A Zakaria.   

Abstract

Before clinical use of a brachytherapy source, regulations or recommendations by medical physics societies require an independent measurement of its air kerma strength by a qualified medical physicist. Currently, in addition to Ir-192, also HDR-Co-60 sources are increasingly coming into operation. However, the existing dosimetry protocols do not provide any guidelines for Co-60 sources. The purpose of this work was therefore to compare air kerma rate measurements as recommended by different dosimetry protocols for Ir-192 HDR sources and to test their applicability to Co-60 sources. Dosimetric verification of HDR afterloading source specification was performed according to three protocols, DIN 6809-2 (1993) in combination with DGMP-Report 13 (2006), IAEA-TECDOC-1274 (2002) and AAPM Report 41 (1993) for the nuclides Ir-192 and Co-60. Measurements of the sources reference air kerma rate were performed with 3 different methods (with a cylindrical chamber both in a solid phantom and in free air, and with a well chamber) and evaluated using all three protocols for each type of source and method of measurement. The measurements with all protocols and methods show deviations from the certified specification smaller than about 1.2% for Ir-192 and 2.5% for Co-60-Sources. The measurements with the well chamber showed the lowest deviations from the certificate value. Air kerma rate measurements for Co-60 HDR sources using the existing protocols are possible with accuracy sufficient to verify source calibration as provided by the source certificate. However, extension of the protocols by correction factors for measurement with Co-60 sources would be helpful.

Entities:  

Keywords:  Air kerma rate measurement for Ir-192; Co-60; brachytherapy; dosimetry

Year:  2012        PMID: 23293449      PMCID: PMC3532746          DOI: 10.4103/0971-6203.103603

Source DB:  PubMed          Journal:  J Med Phys        ISSN: 0971-6203


Introduction

In modern brachytherapy, dose calculations are usually performed with a standardised formalism developed and published by Task Group 43 (TG 43) of the Radiation Therapy committee of the American Association of Physicists in Medicine (AAPM).[12] This formalism uses sets of standardised calculation factors which are determined makes use of dosimetry data obtained by measurement and/or Monte-Carlo calculation for each type and design of source. The adaption of the calculation to the individual source (and its activity) is achieved by its air kerma strength which is defined as the air kerma rate at distance d in vacuo due to photons of energy greater than δ, multiplied by d2[2] In the TG 43 report, for convenience, the unit of air kerma strength is called U where 1 U = 1 cGy h-1 cm2 = 1 μGy h-1 m2. In most other modern protocols and recommendations for brachytherapy dosimetry (IAEA TECDOC 1274,[3] DIN 6809-2,[45] ESTRO Physics booklet No.8[6] ICRU 38,[7] ICRU 58[8]…) the quantity used for this adaption is the reference air kerma rate in air , which can be measured in free in air. Usually, identical numerical values are used for and SK, and is specified at 1 m distance in units of μGy h-1. For commercial brachytherapy sources, air kerma strength or reference air kerma rate is specified in the source's calibration certificate provided by the manufacturer, typically with an accuracy of ±5%. Dosimetry recommendations and regulations in practically all countries require an independent measurement air kerma rate (called dosimetric verification in the following text) by a medical physicist before therapeutic applicaton of the source. In case of larger deviations between measurement and manufacturer specification, the reason for the deviation should be determined.[459] The most common nuclide used in modern HDR afterloading machines presently is Ir-192, however, use of Co-60 is increasing. The dimensions of the modern sources of these two isotopes are almost identical.[1] Due to the lower specific activities achievable for Co-60 the available sources typically have a lower activity of around 74 GBq (in comparison to 370 GBq for Ir-192). Nevertheless, treatment times required for 74 GBq Co-60 is only 1.8 times longer than that of a comparable 370 GBq Ir-192 due to the higher air kerma rate constant Γ∂ (around 0.306 μGyh -1 m2/MBq for Co-60 sources in comparison to 0.099-0.11 μGyh-1 m2/MBq for the commercially available Ir-192 sources.[10]). Dose distributions for Co-60 and Ir-192 are nearly identical,[11] so that Co-60 may be an attractive alternative to Ir-192. The advantage of Co-60 is its longer half-life of over five years in comparison to 74 days for Ir-192 so that Co-60 sources need to be replaced only at periods of several years. This may reduce effort and costs for source transport, purchase and disposal and for quality assurance. On the other hand Co-60 requires increased radiation shielding. The longer useful half-life makes Co-60 sources especially interesting for developing countries. Recommendations for dosimetric verification of air kerma strength for afterloading sources are given in IAEA-TECDOC 1274,[3] AAPM Report 41,[12] for Germany in DIN 6809-2[4] in combination with DGMP Report 13[5] and in ESTRO Physics booklet No.8.[6] While DGMP Report 13 and the ESTRO physics booklet[6] (which is not used in this work) describe measurements using a thimble ionization chamber free in air and in a PMMA solid phantom and using a well-type chamber the other two reports only give recommendations for measurements free in air and with a well chamber. Dosimetric parameters to calculate air kerma rate from instrument reading for the measurement of Ir-192 are given in DGMP Report 13, in IAEA-TECDOC 1274 and in the ESTRO physics booklet, while in AAPM Rept. 41 they are missing. None of the protocols is intended for calibration of Co-60 sources and therefore none provides these parameters for Co-60. Several publications in literature describe dosimetric verification of Ir-192 sources using DGMP Report 13 and IAEA-TECDOC 1274.[91314] First measurements with Co-60 sources using the new Bebig afterloader have been reported in 2006 and 2008.[15-17] Purpose of this work is to compare dosimetric verification of Ir-192 and Co-60 afterloading sources applying all three protocols to measurements with all three methods above. Results are discussed in comparison to each other and to recent publications. This present work is a modification of German version of current publication.[17]

Materials and Methods

Dosimetry was performed according to all three protocols mentioned above. Ir-192 measurements were performed with the Nucletron afterloader while Co-60 measurements were performed with the Bebig Multisource afterloading machine which can operate both Ir-192 and Co-60 sources. The measurements were performed within a time period of about one year with 4 different Ir-192 and 2 Co-60 sources at three different institutions. Two different thimble chambers (M31013 and M23331, both PTW, Freiburg) and one well-type chamber (077.094-25208, PTW Freiburg) were used for the measurements. All chambers were calibrated at the PTW Freiburg secondary standard laboratory. Chambers were calibrated in terms of absorbed dose to water (Nw), in air kerma (Nk) and in exposure (Nx) following the three dosimetry protocols.[3412] Thimble chamber calibration factors Nw, were specified for Co-60 together with a radiation quality correction factor kQ for 250 kV X-rays while Nk and Nx were provided for Co-60, Cs-137 and 250 kV X-rays. The well-type chamber was calibrated directly in terms of reference air kerma rate or exposure rate both for Co-60 and Ir-192. Measurements were performed with a UNIDOS E electrometer (PTW Freiburg) in electrical units (nC or nA). The free-in air measurements were performed with the M23331 chamber (chamber volume 1 cm3), measurements in the PMMA phantom with the M31013 chamber (0.3 cm). Table 1 lists all methods of measurement for both isotopes together with the protocols used for evaluation: measurements free-in air and in the well-type chamber were performed and evaluated according all three protocols, in the PMMA phantom AL 9193 (PTW Freiburg, “Krieger” phantom) measurements followed DIN 6809-2 (1993)[4]/ DGMP rept. 13.[5] Measurements in a water phantom are not reported in this work. All protocols give recommendations only for Ir-192, correction factors for Co-60 were taken from literature.[10151618]
Table 1

Measurements performed for each protocol, (*) mark denotes: procedure/method is not available for specific protocol

Measurements performed for each protocol, (*) mark denotes: procedure/method is not available for specific protocol

Measurement of reference air kerma rate

This section describes the measurement of the reference air kerma rate at 1m distance in air. Some publications specify air kerma rate in vacuo.[210] This difference is not corrected for in this work. Integration time, i.e. the dwell time of the source in the measuring position for measurements free-in-air and in the PMMA phantom was always 3 mins. Transfer time from and back to the after loader is not accounted for. Immediately before each series of free-in-air or well chamber measurements the dwell position providing maximal dose rate was determined, and used in the following measurements. In the free-in-air and well chamber measurements the reading introduced in the evaluation of air kerma rate was determined as the average charge collected for five measurements. The reading of the well type chamber was the electrical current after arriving at a constant value. These measurements were performed only once after verifying that the maximum dose rate does not change for repeated measurements. The readings were introduced into the following formulae to calculate reference air kerma rate. In the formulae the abbreviations for factors are explained only at their first occurrence.

Measurement of air kerma rate with a thimble ionization chamber in a solid phantom following DIN 6809-2 in combination with DGMP Rept. No. 13:

Figure 1 shows the set-up for the cylindrical PMMA phantom 9193 (diameter 20 cm, length 13 cm) together with the Bebig after loader during measurements for a Co-60 source. An insert in the centre accommodates a needle in which the source is positioned. Four additional inserts measuring probes are placed symmetrically with their centre at 8cm distance from the phantom axis. Four measurements are performed, in which the ionization chamber is alternately positioned in each insert. The remaining bores are filled with dummy inserts.
Figure 1

Shows the experimental set up in the Krieger phantom with the Bebig After loader and a Co-60-Source

Shows the experimental set up in the Krieger phantom with the Bebig After loader and a Co-60-Source Reference air kerma rate is determined following the equation[45]: where gw: is the fraction of energy of the electrons from the source decay liberated by photons in water that is lost to radiative processes (mostly bremsstrahlung) tenw/a: is the ratio water/air of the mean mass-energy absorption coefficients (tenw/a = 0.900 for Co-60 and Ir-192) kwp: is a correction factor accounting for the differences in scatter and distortion of the radiation field between water and PMMA (kwp = 1.000 für Co-60 und Ir-192) kzp: is a correction factor accounting for the differences in scatter and absorption in the PMMA phantom surrounding the measuring probe in comparison to free-in-air conditions. This factor is also referred to in the remaining text as the phantom calibration factor. kr = (rM/r0)2 is the correction for a measuring distance rM between probe and source in relation to the reference distance r0 for (r0= 100 cm) kT = 60/(T/min) with T = measuring time in min kA: correction for attenuation and scatter by the applicator (source holder) kP: correction factor for polarisation effect of the ionization chamber kS: correction factor for recombination losses in the ionization chamber kr: air density correction for differing temperature and air pressure from reference conditions kQ: correction factor for the different response of the ionization chamber at the measured radiation quality in comparison to the calibration quality Co-60 Nw: calibration factor of ionization chamber in terms absorbed dose to water M: reading in nC Values for the factors used in this work are given in Tables 2–4.
Table 2

Parameters for the determination of the Air kerma rate from the reading

Table 4

Methods for the determination of the factors kS and kp using different voltage value in different methods

Parameters for the determination of the Air kerma rate from the reading Calibration factors used for the different protocols Methods for the determination of the factors kS and kp using different voltage value in different methods

Measurement of reference air kerma rate with a thimble ionization chamber free in air following DIN 6809-2 in combination with DGMP Rept. 13:

Source and ionization chamber are positioned on a calibration jig shown in Figure 2. On the jig the distance source–chamber can be varied at continuous positions between 5 and 80 cm by moving the chamber holder on a rail. A mm scale allows adjusting the distance between chamber axis and source axis within an uncertainty of 0.5 mm. A second holder on the rail allows mounting a lead block between chamber and source which is used to absorb direct radiation from the source and thus measure the scatter radiation component in the air kerma measurement. The block has the same height as the sensitive volume of the ionization chamber (plus Co-60 build-up cap) and a shielding thickness of 7 cm. The block is placed immediately next to the ionization chamber.[19] All free-in-air measurements were performed with a Co-60 build-up cap.
Figure 2

Shows the experimental set up in “free in air” with the Nucletron After loader and an Ir-192-Source with shield block placed between chamber and the source

Shows the experimental set up in “free in air” with the Nucletron After loader and an Ir-192-Source with shield block placed between chamber and the source Reference air kerma rate is then determined following the equation[45]: where: kair: Correction for losses of primary radiation due to scatter and attenuation in the air between source and ionization chamber (kair(rM) = exp[μ*rM] with μ = 0.00011/cm for Ir-192 and Co-60) kw→a: Is a correction factor accounting for the differences in scatter and radiation field distortion in air surrounding the measuring probe in comparison to water (kw→a = 1.000 for Ir-192 and Co-60) ga: Is the fraction of energy of the electrons from the source decay liberated by photons in air that is lost to radiative processes (mostly bremsstrahlung) kAK: Is a correction factor for attenuation and scatter from a Co-60 build-up cap (kAK = 1.005 was used both for Ir-192 and Co-60) kV: Correction for the finite volume of the measuring probe (following the recommendations by Kondo and Randolph[20] the 1 cm3 thimble chamber 23331 (1.0 cm3) has a kV = 1.0010 for source-chamber distances between15 cm and 25 cm and kV = 1.0000 for source-chamber distances ≥ 30 cm) kscatt: Correction for scatter from surrounding objects (floor, walls, set-up, etc.) kscatt = (M - Mscatt)/M with M = reading without scatter absorber, Mscatt reading with absorber NW: Calibration factor of ionization chamber in terms of absorbed dose to water M: reading in nC

Measurement of reference air kerma rate with a thimble ionization chamber free in air following IAEA-TECDOC-1274:

The set-up for the measurements following IAEA-TECDOC 1274 and AAPM Rept.41 is identical to that according to DIN-6809. Reference air kerma rate is determined following the equation[3]: with: kV: Correction for the finite volume of the measuring probe (named “non-uniformity correction factor” kn in IAEA-TECDOC 1274) Nk: Calibration factor of ionization chamber in terms of air kerma. Nk for Ir-192 was calculated by the equation given in Table 3 (as recommended by IAEA-TECDOC 1274).
Table 3

Calibration factors used for the different protocols

M: Reading in nC

Measurement of reference air kerma rate with a thimble ionization chamber free in air following AAPM Report 41:

Reference air kerma rate is determined following the equation[12]: where (W/e)air = 33.97 J/C = 8.76 mGy/R is the ionization energy of dry air kS: Correction factor for recombination losses in the ionization chamber (named “correction for the collection efficiency at calibration Aion” or “correction for the collection efficiency at the time of the study Pion” in AAPM Report 41) kscatt: Correction for room scatter (in AAPM Report 41: “room scatter correction PRS”) Nx: Calibration factor of ionization chamber in terms of exposure. Nx for Ir-192 was calculated by the equation given in Table 3 (as recommended by IAEA-TECDOC 1274). M: Reading in nC The “exposure gradient (displacement) correction Pgrad” recommended in AAPM Report 41 is set to 1.0 in this work.

Measurement of reference air kerma rate with a well-type chamber following DIN 6809-2 and IAEA-TECDOC1274:

Set-up for the well chamber measurements is identical in both protocols [Figure 3]. The chamber was positioned at large distance from any wall or large object in order to reduce scatter radiation as far as possible. The measurements were performed in a dedicated source holder for each type of source (for the Nucletron Ir-192 source: T33002.1 and for the Bebig Co-60 source: T3304.1, both PTW Freiburg) at the dwell position in the chamber producing the maximal ionization current (the “sweet spot”). In the measurements the identical source holder (type and serial no.) must be used as for calibration of the well chamber. Measurements were performed in current mode of the electrometer after achieving a constant current.
Figure 3

Shows the measurements with the well chamber with the Nucletron After loader and an Ir-192-Source

Shows the measurements with the well chamber with the Nucletron After loader and an Ir-192-Source Reference air kerma rate is determined following the equation[3-5]: with kS = 1/Aion:Correction factor for recombination losses in the well-type chamber Nk: Calibration factor of well-type chamber in terms of absorbed dose to water M: Reading in nA

Measurement of reference air kerma rate with a well-type chamber AAPM Report 41:

In contrast to DIN 6809-2 and IAEA-TECDOC 1274, AAPM Rept. 41 requires a well-type chamber which is calibrated in terms of exposure. Reference air kerma rate is determined following the equation[12]: where: Nx: Calibration factor of well-type chamber in terms of exposure M: Reading in nA Table 2 shows parameters and correction factors used in the calculation of air kerma rate from the readings. Additional factors have been described previously. The values of all factors were taken from literature.[3-61012141517] Usually, ionization chambers are not calibrated directly for Ir-192. The different protocols solve this problem by different methods DIN 6809-2 uses the calibration factor for Co-60 for both isotopes and corrects the Ir-measurements for differing radiation quality with a factor k = 1.0. IAEA[3] and AAPM[12] recommend interpolating the Ir-192 calibration from calibration factors for 250 kV x-rays and Co-60 gamma radiation as shown in Table 3. A number of corrections must be determined for the individual ionization chamber. Their calculation and values are listed in Table 4. The well-type chambers are individually calibrated for each isotope. While DIN 6809-2 (Rept. 13) and IAEA-TECDOC 1274 give recommendations for all correction factors for Ir-192, some are missing in AAPM Rept. 41. None of the protocols gives any correction factors for Co-60. These were therefore taken from scientific literature[151618] . Where no published factors were found, either the values recommended for Ir-192 by DIN 6809-2 or IAEA-TECDOC 1274, or where applicable, were set equal to 1 (AAPM Rept. 41).

Results

The sources were repeatedly measured on different days. On each day the measured reference air kerma rates were recalculated to the equivalent rate at the time of calibration as indicated in the source certificate. Then the differences between measured and certified air kerma rates were averaged separately for each protocol. The variance of the measurements additionally indicates the reproducibility of each method. Mean deviations from the certified calibration and the variation of the measurements are shown as fat bars in Figure 4, the error bars indicate the variation of repeated measurements.
Figure 4

Deviations from the certified calibration of Co-60 and Ir 192 sources in three protocols. The error bars indicate the variation of repeated measurements

Deviations from the certified calibration of Co-60 and Ir 192 sources in three protocols. The error bars indicate the variation of repeated measurements All protocols and methods of measurement show agreement with the certified air kerma rates within maximally 1.2% for Ir-192 and 2.5% for Co-60. Table 5 shows that generally the deviations of the measured from the certified air kerma rate are larger for Co-60 (maximum deviation –2.5%) than for Ir-192 (maximum deviation 1.2%). The smallest deviations (<1%) are found in the measurements with the well-type chamber. The Ir-192 results of this work agree well with results in literature.[1421] Deviations are far below the uncertainties of ±6% for Co-60 (Amersham-Buchler/Bebig) or ±5% for Ir-192 (Nucletron) specified in the source certificates provided by the suppliers.[9]
Table 5

Deviations of measured Air kerma rate from the source certificate value and measured relative uncertainties (in brackets) for different protocols

Deviations of measured Air kerma rate from the source certificate value and measured relative uncertainties (in brackets) for different protocols First measurements on Bebig Co-60 sources were published by Andrassy and Grundei in 2006[15] and 2008[16] after introduction of the new Bebig afterloader. In this work the authors also determined the Co-60 phantom calibration factor kzp for the Krieger PMMA phantom. These measurements also deviated from the certified values by ±2% and therefore confirm our measurements in the Krieger phantom where we obtained deviations of 1.6%. Our Co-60 measurements free in air deviate from the certificate by 2.5%, the well chamber measurements by 0.8%.

Discussion

In the experiments, free in air the scatter radiation component was determined by a second measurement using a scattering block which was then subtracted from the measurement without absorber. The scatter contribution was dependent on the position of source and ionization chamber and their distance from any wall or massive object. For Ir-192 it amounted to around 2% (corresponding to a correction factor k = 0.98) and for Co-60 around 4% (k = 0.96). For the measurements in the PMMA phantom and using the well-type chamber, scatter contributions can be neglected provided the distances to walls and massive objects are large enough (generally >1.5 m). Free in air measurements require a minimum ionization chamber volume of 1 cm3.[4] Whether or not a build-up cap should be used is discussed controversial in literature. DGMP Rept.13 and IAEA-TECDOC 1274 recommend the use of a Co-60 build up cap both for Co-60 and for Ir-192. Different protocols recommend source-chamber distances between 10-30cm[45] and 1m.[312] At a distance in the region of 1m the chamber signal can be very small and measuring times very large, so that statistical noise and dark currents can increase measurement uncertainty. On the other hand, at this distance positioning errors and the uncertainty of the effective measuring position of the probe contribute less to overall uncertainty. Generally, the positioning uncertainties must be reduced to a minimum. This can be achieved e.g., by use of a precise measuring jig (as in this work) or also by varying the position and determining the exact source chamber distance from the distance dependence of the reading using the method described by Goetsch.[22] The uncertainties in these measurements are summarized in Table 6. The uncertainties of any quantity were either taken from the specifications of the respective manufacturer, the recommendations in the protocols or calculated as standard deviations of repeated measurements. The total uncertainty is calculated as the root square sum of the individual uncertainties. These independently calculated total uncertainties for all three methods agree well with the best practice uncertainties (k = 1) as stated by AAPM Task Group No. 138[23] for the measurement of S in the clinic using a well-type chamber.
Table 6

Overview of influence quantities and their contributions to total uncertainty. Where not stated otherwise, uncertainties are taken from[10]

Overview of influence quantities and their contributions to total uncertainty. Where not stated otherwise, uncertainties are taken from[10] In the equations recommended by some protocols to calculate air kerma rate a number of corrections are missing. E.g. DGMP Rept.13 uses no corrections ks for recombination losses or kA for attenuation by the applicator. Especially applicator attenuation can require a considerable factor when measuring in a steel or titanium applicator. In this work attenuation factors kA = 1.0165 and 1.0267 for Ir-192 and Co-60, respectively were used. These factors were taken from recommendations by Bebig TPS HDRplus version 2.2[24] and Baltas.[10] For calibration measurements on Ir-192 using an ionization chamber some protocols recommend to obtain two chamber calibrations (DIN 6809-2[4]: Co-60 and 250 kV X-rays; AAPM TG 56[25]: Cs-137 and 250 kV X-rays) in order to calculate the interpolated calibration factor for Ir-192. Since determination of Nk by interpolation will add some increased uncertainty in the air kerma calibration factor for Ir-192, a direct calibration for Ir-192 is desirable as it is provided by the Britisch NPL[26] and the German PTB.[27] These standard laboratories will provide Ir-192 calibrations in terms of reference air kerma rate for well-type chambers, and for free complete measurement arrangements for free-in air measurement (consisting of ionization chamber, source holder and measurement jig) or complete PMMA phantom arrangements (consisting of ionization chamber, source holder and phantom). Calibrations of other types of ionization chambers in terms of air kerma rate for Ir-192 are so far not available. Well-type chambers have proven to be fast and precise measuring instruments. They are also used for source calibration by the manufacturers. In clinical use the PMMA phantom may have some advantages since it can also be used e.g., to check the calibration of in-vivo probes or for QA tests of the after loading machine. If one has no direct calibration of the phantom arrangement (as provided by PTB[27]), one needs to know the phantom calibration factor kzp. In some cases kzp can be obtained from the manufacturer (or from literature), otherwise one needs to determine it by cross-calibration from a measurement of reference air kerma rate using one of the other methods. Calibration factors for the Krieger type PMMA phantom used in this work are available both for Ir-192 and for Co-60. DGMP Report 13 provides guidance for dosimetry of Ir-192 sources for all three methods of measurement (including measurements in the Krieger type PMMA phantom) following the German standard protocol DIN 6809-2 which is dated from 1993 and is outdated in some parts. The international protocol IAEA-TECDOC 1274[3] and AAPM Report 41[12] give recommendations only for measurements free in air and for the well-type chamber. AAPM Report 41 report (from 1992) is outdated since it still recommends using an ionization chamber calibrated in exposure but is in principle still applicable provided one can obtain a calibration in exposure. This type of calibration has been replaced by calibration in terms of absorbed dose to water in most protocols, including the American AAPM TG 51 protocol for high energy photon and electron beams.[28] AAPM TG 56[25] recommends measuring Ir-192 source strength using instruments calibrated directly in terms of air kerma however gives no correction factors. It is desirable to agree on an internationally accepted common protocol which describes all methods of calibration using the best available correction factors, and also giving recommendations for Co-60 sources, and possibly also for a number of other nuclides which may be available for high dose rate after loading in the near future. A direct calibration of the chambers for Ir-192 compared to the interpolative method (uncertainty about 2%) showed an uncertainty of 0.8%.[2930] According to AAPM TG 138,[23] the propagation of uncertainties from the various well-chamber measurements involved in the transfer of the source-strength standard to the clinic results in a minimum expanded uncertainty (k = 2) in SK,CLINIC of 2.56%. Therefore, in clinical practice it is recommendable to use the more precise direct calibration. In the international co-operation EURAMET[31] several national standard laboratories have collaborated in the project JRP06 brachytherapy in which new standards for the calibration of brachytherapy sources are under development. Within this project methods are being developed to directly measure the absorbed dose to water surrounding brachytherapy isotopes including Ir-192 at distances of a few cm.[31-37] As a major result the project has obtained an improved value for the dose rate constant Λ reducing the uncertainty in source specification (in reference air kerma rate) from previously ±5% to around ±2% (k = 2).[37] Source specification directly in terms of absorbed dose to water is not planned. If this were the case the same methods of source verification could be used. With these standards it would however be necessary to determine new calibration factors for the instruments in terms of absorbed dose to water. Dose calculations with such a specification could be performed in the same manner as today, since in the AAPM TG-43 formalism[12] the product is equal to the dose rate in water under reference conditions. The remaining formalism could be used without changes since all factors in the calculation are defined already in terms of absorbed dose to water.

Conclusions

Verification of source calibration of Co-60 HDR brachytherapy sources by the clinical user is feasible. Source strength for Co-60 sources can be measured with comparable uncertainty using the same instruments with the same dosimetry protocols as used for the more common Ir-192 sources. The correction factors to calculate air kerma rate from the reading of calibrated ionization chambers for Co-60 can be taken from literature. Since dose distributions for Co-60 are very similar to those for Ir-192,[11] Co-60 brachytherapy sources may be an attractive and cheaper alternative to the existing Ir-192 sources.
  13 in total

1.  Comparison of calibration procedures for 192Ir high-dose-rate brachytherapy sources.

Authors:  D Baltas; K Geramani; G T Ioannidis; K Hierholz; B Rogge; C Kolotas; K Müller-Sievers; N Milickovic; B Kober; N Zamboglou
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-02-01       Impact factor: 7.038

2.  [Dosimetry of HDR afterloading machines with Ir-192- und Co-60-sources: comparison of different international protocols].

Authors:  Golam Abu Zakaria; Wilhelm Schütte; Hasin Anupama Azhari
Journal:  Z Med Phys       Date:  2010-07-16       Impact factor: 4.820

3.  Direct measurement of absorbed dose to water in HDR 192Ir brachytherapy: water calorimetry, ionization chamber, Gafchromic film, and TG-43.

Authors:  Arman Sarfehnia; Iwan Kawrakow; Jan Seuntjens
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

4.  Comparison of dosimetric standards of USA and France for HDR brachytherapy.

Authors:  Guilhem Douysset; Jean Gouriou; Frank Delaunay; Larry DeWerd; Kurt Stump; John Micka
Journal:  Phys Med Biol       Date:  2005-04-13       Impact factor: 3.609

5.  Comparison of 60cobalt and 192iridium sources in high dose rate afterloading brachytherapy.

Authors:  Jürgen Richter; Kurt Baier; Michael Flentje
Journal:  Strahlenther Onkol       Date:  2008-04       Impact factor: 3.621

6.  Comparison of air kerma standards of LNE-LNHB and NPL for 192Ir HDR brachytherapy sources: EUROMET project no 814.

Authors:  Guilhem Douysset; Thorsten Sander; Jean Gouriou; Rebecca Nutbrown
Journal:  Phys Med Biol       Date:  2008-03-06       Impact factor: 3.609

Review 7.  The evolution of brachytherapy treatment planning.

Authors:  Mark J Rivard; Jack L M Venselaar; Luc Beaulieu
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

8.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

Review 9.  Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; J A Meli; A J Olch; J A Stitt; J F Williamson
Journal:  Med Phys       Date:  1997-10       Impact factor: 4.071

10.  Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; G Luxton; K A Weaver; J F Williamson; A S Meigooni
Journal:  Med Phys       Date:  1995-02       Impact factor: 4.071

View more
  1 in total

1.  Evaluation of BEBIG HDR (60)Co system for non-invasive image-guided breast brachytherapy.

Authors:  Mehdi Zehtabian; Sedigheh Sina; Mark J Rivard; Ali S Meigooni
Journal:  J Contemp Brachytherapy       Date:  2015-12-30
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.