Guilhem Douysset1, Jean Gouriou, Frank Delaunay, Larry DeWerd, Kurt Stump, John Micka.
Abstract
A bilateral comparison of national dosimetric standards for high dose rate brachytherapy has been conducted between the Laboratoire National Henri Becquerel and the University of Wisconsin Accredited Dosimetry Calibration Laboratory. A complete overview of the methods that are currently in use to establish the two national standards is given. The comparison has been carried out using well-type transfer ionization chambers. Three units have been calibrated in both laboratories, and calibration coefficients have been compared. The discrepancies between the two measurements are within 0.3%.Mesh:
Substances:
Year: 2005 PMID: 15843730 DOI: 10.1088/0031-9155/50/9/003