Evaluation of MRI artifacts at 3 Tesla for 38 commonly used cosmetics.

Kirin Escher1, Frank G Shellock.   

Abstract

PURPOSE: To evaluate MRI artifacts at 3-Tesla for 38 commonly used cosmetics.
MATERIALS AND METHODS: Thirty-eight cosmetics (16, nail polishes; 5, eyeliners; 3, mascaras; 10, eye shadows; 1, lip gloss; 1, body lotion; 1, body glitter, and 1, hair loss concealer) underwent evaluation for MRI artifacts at 3-Tesla. The cosmetics were applied a copper-sulfate-filled, phantom and initially assessed using a "screening" gradient echo (GRE) pulse sequence. Of the 38 different cosmetics, 14 (37%) exhibited artifacts. For these 14 cosmetics, additional characterization of artifacts was performed using a GRE pulse sequence. A qualitative scale was applied to characterize the artifact size.
RESULTS: Artifacts were observed, as follows: 2, nail polishes; 5, eyeliners; 3, mascaras; 3, eye shadows; 1, hair loss concealer. Artifact size ranged from small (eye shadow) to very large (hair loss concealer) and tended to be associated with the presence of iron oxide or other metal-based ingredient.
CONCLUSIONS: Commonly used cosmetics caused artifacts that may create issues if the area of interest is the same as where the cosmetic was applied or if its presence was unknown, thus, potentially causing it to be construed as pathology. Therefore, these findings have important implications for patients referred for MRI examinations.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23290125     DOI: 10.1016/j.mri.2012.11.002

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  4 in total

1.  Unexpected MRI artifacts - experience from India.

Authors:  Anitha Sen
Journal:  Pediatr Radiol       Date:  2015-06-17

2.  The imaging appearance of crayons.

Authors:  Aaron S McAllister; Neil U Lall; Kareem O Tawfik; Blaise V Jones
Journal:  Pediatr Radiol       Date:  2017-03-07

3.  Kajal-induced Artefact Simulating a Ciliary Body Tumor on Magnetic Resonance Imaging.

Authors:  Venkatraman Indiran; L Raguram Subha; Jagannathan Kokilavani
Journal:  Turk J Ophthalmol       Date:  2018-06-28

4.  Trace Metal Lead Exposure in Typical Lip Cosmetics From Electronic Commercial Platform: Investigation, Health Risk Assessment and Blood Lead Level Analysis.

Authors:  Yanan Li; Yanyan Fang; Zehua Liu; Yahan Zhang; Kangli Liu; Luping Jiang; Boyuan Yang; Yongdie Yang; Yongwei Song; Chaoyang Liu
Journal:  Front Public Health       Date:  2021-11-17
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.