Literature DB >> 23280561

Phaco-efficiency test and re-aspiration analysis of repulsed particle in phacoemulsification.

Jae-hyung Kim1, Dong-Ah Ko, Jae Yong Kim, Myoung Joon Kim, Hungwon Tchah.   

Abstract

BACKGROUND: To measure the efficiency of phacoemulsification, we have developed a new experimental model for testing phaco-efficiency and analyzed re-aspiration of repulsed particles.
METHODS: Using a Kitaro wetlab system, a piece of blood agar (BA) was placed in an artificial chamber and the phacoemulsifier was placed horizontally. The settings of the phacoemulsifier (Infiniti, Alcon Laboratories) were 26 cc/min for aspiration, 350 cc/min for vacuum, and 95 cm of bottle height. The time to remove BAs was measured using Ozil 100 %, Ozil 40 %, and longitudinal 40 % of phaco power. The angle between the re-aspirated BA particles and the axis of the phacoemulsifier (re-aspiration zone, degree) was analyzed.
RESULTS: The average time (seconds) to remove BAs was lower in the Ozil 100 % and the Ozil 40 % mode than in the longitudinal mode (0.37 ± 0.39, 0.85 ± 0.57, and 2.22 ± 1.40 respectively, P value < 0.01). Repulsion exceeding 1 mm occurred more frequently in the longitudinal mode than in the Ozil 100 % mode (100 % vs 40 %, P value = 0.01, Fisher's exact test). The average of re-aspiration zone was 25.9 ± 14.5 in the longitudinal 40 % and 54.0 ± 23.0 in the Ozil 40 % (P value = 0.016).
CONCLUSIONS: The Ozil mode was more efficient than the longitudinal mode. In addition, the Ozil mode provided less repulsion and wider aspiration zone.

Entities:  

Mesh:

Year:  2013        PMID: 23280561     DOI: 10.1007/s00417-012-2240-0

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  11 in total

1.  Bead-flow pattern: quantitation of fluid movement during torsional and longitudinal phacoemulsification.

Authors:  Luis E Fernández de Castro; Ramon C Dimalanta; Kerry D Solomon
Journal:  J Cataract Refract Surg       Date:  2010-06       Impact factor: 3.351

2.  Fluidics and heat generation of Alcon Infiniti and Legacy, Bausch & Lomb Millennium, and advanced medical optics sovereign phacoemulsification systems.

Authors:  Michael S Floyd; Jeremy R Valentine; Randall J Olson
Journal:  Am J Ophthalmol       Date:  2006-09       Impact factor: 5.258

Review 3.  Phaco fluidics and phaco ultrasound power modulations.

Authors:  Uday Devgan
Journal:  Ophthalmol Clin North Am       Date:  2006-12

4.  A fluidics comparison of Alcon Infiniti, Bausch & Lomb Stellaris, and Advanced Medical Optics Signature phacoemulsification machines.

Authors:  Dan Georgescu; Annie F Kuo; Krista I Kinard; Randall J Olson
Journal:  Am J Ophthalmol       Date:  2008-03-17       Impact factor: 5.258

5.  Comparison of torsional and longitudinal modes using phacoemulsification parameters.

Authors:  Marek Rekas; Robert Montés-Micó; Karolina Krix-Jachym; Adam Kluś; Andrzej Stankiewicz; Teresa Ferrer-Blasco
Journal:  J Cataract Refract Surg       Date:  2009-10       Impact factor: 3.351

6.  Heat production: Longitudinal versus torsional phacoemulsification.

Authors:  Young Keun Han; Kevin M Miller
Journal:  J Cataract Refract Surg       Date:  2009-10       Impact factor: 3.351

7.  Phaco-emulsification and aspiration. A new technique of cataract removal. A preliminary report.

Authors:  C D Kelman
Journal:  Am J Ophthalmol       Date:  1967-07       Impact factor: 5.258

8.  Comparison of torsional and microburst longitudinal phacoemulsification: a prospective, randomized, masked clinical trial.

Authors:  Abhay R Vasavada; Shetal M Raj; Udayan Patel; Vaishali Vasavada; Viraj Vasavada
Journal:  Ophthalmic Surg Lasers Imaging       Date:  2010 Jan-Feb

9.  The comparison between torsional and conventional mode phacoemulsification in moderate and hard cataracts.

Authors:  Dong-Hyun Kim; Won-Ryang Wee; Jin-Hak Lee; Mee-Kum Kim
Journal:  Korean J Ophthalmol       Date:  2010-11-23

10.  Risk factors for intraoperative complications in resident-performed phacoemulsification surgery.

Authors:  Tina Rutar; Travis C Porco; Ayman Naseri
Journal:  Ophthalmology       Date:  2009-01-22       Impact factor: 12.079

View more
  1 in total

1.  A novel method to compare phacoemulsification parameters in vivo: two halves of one nucleus.

Authors:  Elena Tomilova; Sergey Shukhaev
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-06-21       Impact factor: 3.117

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.