Literature DB >> 16935581

Fluidics and heat generation of Alcon Infiniti and Legacy, Bausch & Lomb Millennium, and advanced medical optics sovereign phacoemulsification systems.

Michael S Floyd1, Jeremy R Valentine, Randall J Olson.   

Abstract

PURPOSE: To study heat generation, vacuum, and flow characteristics of the Alcon Infiniti and Bausch & Lomb Millennium with results compared with the Alcon Legacy and advanced medical optics (AMO) Sovereign machines previously studied.
DESIGN: Experimental study.
METHODS: Heat generation with continuous ultrasound was determined with and without a 200-g weight. Flow and vacuum were determined from 12 to 40-ml/min in 2-ml/min steps. The impact of a STAAR Cruise Control was also tested.
RESULTS: Millennium created the most heat/20% of power (5.67 +/- 0.51 degrees C unweighted and 6.80 +/- 0.80 degrees C weighted), followed by Sovereign (4.59 +/- 0.70 degrees C unweighted and 5.65 +/- 0.72 degrees C weighted), Infiniti (2.79 +/- 0.62 degrees C unweighted and 3.96 +/- 0.31 degrees C weighted), and Legacy (1.99 +/- 0.49 degrees C unweighted and 4.27 +/- 0.76 degrees C weighted; P < .0001 for all comparisons between machines except Infiniti vs Legacy, both weighted). Flow studies revealed that Millennium Peristaltic was 17% less than indicated (P < .0001 to all other machines), and all other machines were within 3.5% of indicated. Cruise Control decreased flow by 4.1% (P < .0001 for same machine without it). Millennium Venturi had the greatest vacuum (81% more than the least Sovereign; P < .0001), and Cruise Control increased vacuum in a peristaltic machine 35% more than the Venturi system (P < .0001).
CONCLUSIONS: Percent power is not consistent in regard to heat generation, however, flow was accurate for all machines except Millennium Peristaltic. Restriction with Cruise Control elevates unoccluded vacuum to levels greater than the Venturi system tested.

Mesh:

Substances:

Year:  2006        PMID: 16935581     DOI: 10.1016/j.ajo.2006.04.042

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  3 in total

1.  Phaco-efficiency test and re-aspiration analysis of repulsed particle in phacoemulsification.

Authors:  Jae-hyung Kim; Dong-Ah Ko; Jae Yong Kim; Myoung Joon Kim; Hungwon Tchah
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-03       Impact factor: 3.117

Review 2.  Noninvasive glaucoma procedures: current options and future innovations.

Authors:  Ahmed M Abdelrahman
Journal:  Middle East Afr J Ophthalmol       Date:  2015 Jan-Mar

3.  Torsional ultrasound modality for hard nucleus phacoemulsification cataract extraction.

Authors:  M Zeng; X Liu; Y Liu; Y Xia; L Luo; Z Yuan; Y Zeng; Y Liu
Journal:  Br J Ophthalmol       Date:  2008-06-20       Impact factor: 4.638

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.