Effective patient dose during neuroradiological C-arm CT procedures.

Mei Bai1, Xianghua Liu, Bin Liu.   

Abstract

PURPOSE: We aimed to determine the effective dose given to patients during neuroradiological C-arm computed tomography (CT) procedures.
MATERIALS AND METHODS: Measurements were performed on 48 patients using a dose-area product (DAP) meter. A PC-based Monte Carlo program (PCXMC, STUK, Helsinki, Finland) was used to calculate the effective dose from the DAP values of each patient. Organ doses were also measured with thermoluminescent dosimeters (TLDs) using a human-shaped phantom.
RESULTS: The difference between the organ doses measured using TLDs and PCXMC was not significant (P > 0.05). The mean DAP for 48 patients was 9.41±2.50 Gy·cm2; the mean effective dose for all procedures was 0.30±0.08 mSv. The coefficient for the correlation (R2) between the DAP and the effective dose was 0.97. The conversion factor between the effective dose and DAP was 0.030-0.035 mSv·Gy(-1)·cm(-2).
CONCLUSION: DAP can be used as a dose indicator to calculate the organ dose and effective dose of patients based on Monte Carlo simulation. This method can provide important information on the absorbed dose and enhance the radiation protection of patients during C-arm CT procedures.

Entities:  

Mesh:

Year:  2012        PMID: 23266966     DOI: 10.4261/1305-3825.DIR.5852-12.0

Source DB:  PubMed          Journal:  Diagn Interv Radiol        ISSN: 1305-3825            Impact factor:   2.630


  3 in total

1.  Brain radiation doses to patients in an interventional neuroradiology laboratory.

Authors:  R M Sanchez; E Vano; J M Fernández; M Moreu; L Lopez-Ibor
Journal:  AJNR Am J Neuroradiol       Date:  2014-03-13       Impact factor: 3.825

2.  Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology.

Authors:  Tobias Struffert; Michael Hauer; Rosemarie Banckwitz; Christoph Köhler; Kevin Royalty; Arnd Doerfler
Journal:  Eur Radiol       Date:  2014-04-02       Impact factor: 5.315

3.  Estimation of the effective dose of dental cone-beam computed tomography using personal computer-based Monte Carlo software.

Authors:  Eun-Kyung Kim; Won-Jeong Han; Jin-Woo Choi; Bulgan Battulga
Journal:  Imaging Sci Dent       Date:  2018-03-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.