5-year outcome after transcatheter aortic valve implantation.

Stefan Toggweiler1, Karin H Humphries1, May Lee1, Ronald K Binder1, Robert R Moss1, Melanie Freeman1, Jian Ye1, Anson Cheung1, David A Wood1, John G Webb2.   

Abstract

OBJECTIVES: The purpose of this study was to investigate the 5-year outcome following transcatheter aortic valve implantation (TAVI).
BACKGROUND: Little is known about long-term outcomes following TAVI.
METHODS: The 5-year outcomes following successful TAVI with a balloon-expandable valve were evaluated in 88 patients. Patients who died within 30 days after TAVI were excluded.
RESULTS: Mean aortic valve gradient decreased from 46 ± 18 mm Hg to 10 ± 4.5 mm Hg after TAVI and 11.8 ± 5.7 mm Hg at 5 years (p for post-TAVI trend = 0.06). Mean aortic valve area increased from 0.62 ± 0.17 cm(2) to 1.67 ± 0.41 cm(2) after TAVI and 1.40 ± 0.25 cm(2) at 5 years (p for post-TAVI trend <0.01). At 5 years, 3 patients (3.4%) had moderate prosthetic valve dysfunction (moderate transvalvular regurgitation in 1, moderate stenosis in 1, and moderate mixed disease in 1). Survival rates at 1 to 5 years were 83%, 74%, 53%, 42%, and 35%, respectively. Median survival time after TAVI was 3.4 years (95% confidence interval [CI]: 2.6 to 4.3), and the risk of death was significantly increased in patients with chronic obstructive pulmonary disease (adjusted hazard ratio [HR]: 2.17; 95% CI: 1.28 to 3.70) and at least moderate paravalvular regurgitation (adjusted HR: 2.98; 95% CI: 1.44 to 6.17).
CONCLUSIONS: Our study demonstrated favorable long-term outcomes after TAVI. Signs of moderate prosthetic valve failure were observed in 3.4% of patients. No patients developed severe prosthetic regurgitation or stenosis. Comorbidities, notably chronic lung disease and at least moderate paravalvular regurgitation, were associated with reduced long-term survival.
Copyright © 2013 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23265333     DOI: 10.1016/j.jacc.2012.11.010

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  40 in total

Review 1.  Valve durability after transcatheter aortic valve implantation.

Authors:  Akash Kataruka; Catherine M Otto
Journal:  J Thorac Dis       Date:  2018-11       Impact factor: 2.895

Review 2.  Transcatheter aortic valve replacement: indications and beyond the clinical trials.

Authors:  Michael J Reardon
Journal:  Tex Heart Inst J       Date:  2013

Review 3.  Guidance of transcatheter aortic valve replacement by echocardiography.

Authors:  Rebecca T Hahn
Journal:  Curr Cardiol Rep       Date:  2014-01       Impact factor: 2.931

Review 4.  Pacemaker implantation after transcatheter aortic valve: why is this still happening?

Authors:  Stefan Toggweiler; Richard Kobza
Journal:  J Thorac Dis       Date:  2018-11       Impact factor: 2.895

5.  Transcatheter aortic valve replacement: favorable clinical outcomes support role in intermediate risk surgical patients.

Authors:  Ravilla Mahidhar; Jon R Resar
Journal:  J Thorac Dis       Date:  2016-09       Impact factor: 2.895

6.  The Odyssey of TAVR from concept to clinical reality.

Authors:  Alain G Cribier
Journal:  Tex Heart Inst J       Date:  2014-04-01

Review 7.  Durability of prostheses for transcatheter aortic valve implantation.

Authors:  Mani Arsalan; Thomas Walther
Journal:  Nat Rev Cardiol       Date:  2016-04-07       Impact factor: 32.419

8.  Echocardiographic outcomes of self-expandable CoreValve versus balloon-expandable Edwards SAPIEN XT valves: the comparison of two bioprosthesis implanted in a single centre.

Authors:  Dayimi Kaya; Zulkif Tanriverdi; Huseyin Dursun; Tugce Colluoglu
Journal:  Int J Cardiovasc Imaging       Date:  2016-06-09       Impact factor: 2.357

9.  Balloon- or Self-Expandable TAVI: Clinical Equipoise?

Authors:  John Jose; Gert Richardt; Mohamed Abdel-Wahab
Journal:  Interv Cardiol       Date:  2015-05

Review 10.  Aortic regurgitation after transcatheter aortic valve implantation: mechanisms and implications.

Authors:  Barbara E Stähli; Willibald Maier; Roberto Corti; Thomas F Lüscher; Rolf Jenni; Felix C Tanner
Journal:  Cardiovasc Diagn Ther       Date:  2013-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.