Pre-clinical evaluation of the infant Jarvik 2000 heart in a neonate piglet model.

Xufeng Wei1, Tieluo Li, Shuying Li, Ho Sung Son, Pablo G Sanchez, Pablo Sanchez, Shuqiong Niu, A Claire Watkins, Christopher DeFilippi, Robert Jarvik, Zhongjun J Wu, Bartley P Griffith.   

Abstract

BACKGROUND: The infant Jarvik 2000 heart is a very small, hermetically sealed, intracorporeal, axial-flow ventricular assist device (VAD) designed for circulatory support in neonates and infants. The anatomic fit, short-term biocompatibility and hemodynamic performance of the device were evaluated in a neonate piglet model.
METHODS: The infant Jarvik 2000 heart with two different blade profiles (low- or high-flow blade design) was tested in 6 piglets (8.8 ± 0.9 kg). Using a median sternotomy, the pump was placed in the left ventricle through the apex without cardiopulmonary bypass. An outflow graft was anastomosed to the ascending aorta. Hemodynamics and biocompatibility were studied for 6 hours.
RESULTS: All 6 pumps were implanted without complication. Optimal anatomic positioning was found with the pump body inserted 2.4 cm into the left ventricle. Hemodynamics demonstrated stability throughout the 6-hour duration. The pump flow increased from 0.27 to 0.95 liter/min at increasing speeds from 18 to 31 krpm for the low-flow blade design, whereas the pump flow increased from 0.54 liter/min to 1.12 liters/min at increasing speeds from 16 krpm to 31 krpm for the high-flow blade design. At higher speeds, >80% of flow could be supplied by the device. Blood chemistry and final pathology demonstrated no acute organ injury or thrombosis for either blade design.
CONCLUSIONS: The infant Jarvik 2000 heart is anatomically and biologically compatible with an short-term neonate piglet model. This in vivo study demonstrates the future feasibility of this device for clinical use.
Copyright © 2012 International Society for Heart and Lung Transplantation. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23260711      PMCID: PMC3546489          DOI: 10.1016/j.healun.2012.10.011

Source DB:  PubMed          Journal:  J Heart Lung Transplant        ISSN: 1053-2498            Impact factor:   10.247


  21 in total

Review 1.  Extracorporeal life support.

Authors:  Alan M Gaffney; Stephen M Wildhirt; Michael J Griffin; Gail M Annich; Marek W Radomski
Journal:  BMJ       Date:  2010-11-02

2.  Registry of the International Society for Heart and Lung Transplantation: eleventh official pediatric heart transplantation report--2008.

Authors:  Richard Kirk; Leah B Edwards; Paul Aurora; David O Taylor; Jason Christie; Fabienne Dobbels; Anna Y Kucheryavaya; Axel O Rahmel; Marshall I Hertz
Journal:  J Heart Lung Transplant       Date:  2008-09       Impact factor: 10.247

3.  Optimism derived from 7.5 years of continuous-flow circulatory support.

Authors:  Stephen Westaby; Adrian Banning; Desley Neil; Philip Poole-Wilson; O Howard Frazier
Journal:  J Thorac Cardiovasc Surg       Date:  2009-03-17       Impact factor: 5.209

4.  Use of ventricular assist devices in children across the United States: analysis of 7.5 million pediatric hospitalizations.

Authors:  David L S Morales; Farhan Zafar; Joseph W Rossano; Jorge D Salazar; John L Jefferies; Daniel E Graves; Jeffrey S Heinle; Charles D Fraser
Journal:  Ann Thorac Surg       Date:  2010-10       Impact factor: 4.330

5.  In vivo experience of the child-size pediatric Jarvik 2000 heart: update.

Authors:  Marc Gibber; Zhongjun J Wu; Won-Bae Chang; Giacomo Bianchi; Jingping Hu; Jose Garcia; Robert Jarvik; Bartley P Griffith
Journal:  ASAIO J       Date:  2010 Jul-Aug       Impact factor: 2.872

6.  Bridging children of all sizes to cardiac transplantation: the initial multicenter North American experience with the Berlin Heart EXCOR ventricular assist device.

Authors:  David L S Morales; Christopher S D Almond; Robert D B Jaquiss; David N Rosenthal; David C Naftel; M Patricia Massicotte; Tilman Humpl; Mark W Turrentine; James S Tweddell; Gordon A Cohen; Robert Kroslowitz; Eric J Devaney; Charles E Canter; Francis Fynn-Thompson; Olaf Reinhartz; Michiaki Imamura; Nancy S Ghanayem; Holger Buchholz; Sarah Furness; Robert Mazor; Sanjiv K Gandhi; Charles D Fraser
Journal:  J Heart Lung Transplant       Date:  2011-01       Impact factor: 10.247

7.  Current use of the EXCOR pediatric ventricular assist device.

Authors:  Roosevelt Bryant; Marie Steiner; James D St Louis
Journal:  J Cardiovasc Transl Res       Date:  2010-09-04       Impact factor: 4.132

8.  Mechanical support and medical therapy reverse heart failure in infants and children.

Authors:  Hannah Zimmerman; Diane Covington; Richard Smith; Jack Copeland
Journal:  Artif Organs       Date:  2010-11       Impact factor: 3.094

9.  Incidence and risk factors for mortality in infants awaiting heart transplantation in the USA.

Authors:  Douglas Mah; Tajinder P Singh; Ravi R Thiagarajan; Kimberlee Gauvreau; Gary E Piercey; Elizabeth D Blume; Francis Fynn-Thompson; Christopher S D Almond
Journal:  J Heart Lung Transplant       Date:  2009-09-26       Impact factor: 10.247

Review 10.  Mechanical circulatory support as a bridge to pediatric cardiac transplantation.

Authors:  James K Kirklin
Journal:  Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu       Date:  2008
View more
  5 in total

1.  Modern Outcomes of Mechanical Circulatory Support as a Bridge to Pediatric Heart Transplantation.

Authors:  Brody Wehman; Kristen A Stafford; Gregory J Bittle; Zachary N Kon; Charles F Evans; Keshava Rajagopal; Nicholas Pietris; Sunjay Kaushal; Bartley P Griffith
Journal:  Ann Thorac Surg       Date:  2016-02-22       Impact factor: 4.330

Review 2.  Technology landscape of pediatric mechanical circulatory support devices: A systematic review 2010-2021.

Authors:  Thomas Palazzolo; Matthew Hirschhorn; Ellen Garven; Steven Day; Randy M Stevens; Joseph Rossano; Vakhtang Tchantchaleishvili; Amy L Throckmorton
Journal:  Artif Organs       Date:  2022-04-14       Impact factor: 2.663

3.  A transapical-to-aorta double lumen cannula-based neonate left ventricular assist device efficiently unloads the left ventricle in neonate lambs.

Authors:  Cheng Zhou; Dongfang Wang; Cherry Ballard-Croft; Guangfeng Zhao; Hassan K Reda; Stephen Topaz; Joseph Zwischenberger
Journal:  J Thorac Cardiovasc Surg       Date:  2016-08-31       Impact factor: 5.209

4.  Fontan Liver Disease: Review of an Emerging Epidemic and Management Options.

Authors:  Elisa Bradley; Benjamin Hendrickson; Curt Daniels
Journal:  Curr Treat Options Cardiovasc Med       Date:  2015-11

Review 5.  Sizing up models of heart failure: Proteomics from flies to humans.

Authors:  Viola Kooij; Vidya Venkatraman; John Tra; Jonathan A Kirk; Janelle Rowell; Anna Blice-Baum; Anthony Cammarato; Jennifer E Van Eyk
Journal:  Proteomics Clin Appl       Date:  2014-06-25       Impact factor: 3.494

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.