| Literature DB >> 2325449 |
A Lozano1, J Rosell, R Pallás-Areny.
Abstract
A four-channel impedance plethysmograph has been designed. Impedance signals are obtained at two frequencies by measuring both real and imaginary parts. Particular attention has been paid to the sine wave generation circuits that provide system versatility. The required phase-sensitive demodulation is achieved by means of analogue multiplexers. Results show that there are significant variations in the thoracic equivalent capacitance related to respiration and that there is an increased sensitivity to cardiac activity at low frequencies.Mesh:
Year: 1990 PMID: 2325449 DOI: 10.1007/bf02441675
Source DB: PubMed Journal: Med Biol Eng Comput ISSN: 0140-0118 Impact factor: 2.602