| Literature DB >> 23249477 |
Jesús Hernández-Saz1, Miriam Herrera, Diego Alonso-Álvarez, Sergio I Molina.
Abstract
The 3D distribution of self-assembled stacked quantum dots (QDs) is a key parameter to obtain the highest performance in a variety of optoelectronic devices. In this work, we have measured this distribution in 3D using a combined procedure of needle-shaped specimen preparation and electron tomography. We show that conventional 2D measurements of the distribution of QDs are not reliable, and only 3D analysis allows an accurate correlation between the growth design and the structural characteristics.Entities:
Year: 2012 PMID: 23249477 PMCID: PMC3534247 DOI: 10.1186/1556-276X-7-681
Source DB: PubMed Journal: Nanoscale Res Lett ISSN: 1556-276X Impact factor: 4.703
Parameters used in each step of the annular milling process to fabricate GaAs needles with a reduced diameter along a large range
| 1 | 1,000 | 1,500 | 100 | 30 |
| 2 | 800 | 1,400 | 81 | 20 |
| 3 | 700 | 1,200 | 23 | 20 |
| 4 | 600 | 1,000 | 23 | 20 |
| 5 | 500 | 850 | 23 | 20 |
| 6 | 400 | 700 | 4 | 20 |
| 7 | 300 | 600 | 4 | 20 |
| 8 | 150 | 400 | 4 | 20 |
| 9 | - | - | 70 | 5 |
The last step is to clean the amorphous layer around the needle.
Figure 1Cross-sectional HAADF images of the needle-shaped specimen taken at different rotation angles. Note that the angles between the stacking of QDs and the growth direction are different for the three images: (a) 0°, (b) 5°, and (c) 11°.
Figure 2The surfaces render of the reconstructed volume and an axial slice through the needle. (a) Semi-transparent external surface of the tomogram of the needle with opaque surfaces for the QDs below the platinum deposition. A projection of one of the central QDs is included in the base of the needle. (b) Segmentation of the QDs in the tomogram, showing that the stacking of QDs follows a straight line that deviates 10° from the growth direction. (c) Slice through the upper QD of the reconstructed tomogram where we have superimposed a circle to evidence the elongation in the direction of the optical axis of the microscope. The upper and lower QDs of the Figure 2b have been included with a white and black dotted line respectively.