Central corneal thickness measurements obtained with anterior segment spectral domain optical coherence tomography compared to ultrasound pachymetry in healthy subjects.

Lori Vollmer1, Joseph Sowka, Joseph Pizzimenti, Xinha Yu.   

Abstract

INTRODUCTION: Central corneal thickness (CCT) imparts information about an individual's risk of conversion to glaucoma from ocular hypertension, progression of established glaucoma, and the likelihood of developing structural and functional abnormalities in patients with ocular hypertension. Most typically, CCT is measured through ultrasound (US) pachymetry. Currently, optical coherence tomography (OCT) has the ability to image the anterior segment, cornea, and anterior chamber angle. With this ability comes the option of determining CCT. The purpose of this study is to ascertain any significant difference in CCT measurement results as well as quantify the reproducibility of measurements of the two technologies. In addition, by measuring CCT both with traditional US pachymetry as well as spectral domain (SD) OCT technology, we sought to determine if CCT measurement by SD-OCT is an accurate, comparable and viable option;
METHODS: Eighty eyes of forty healthy volunteers were used to determine CCT with both SD-OCT and US pachymetry. Three consecutive measurements were collected with each method on every eye.
RESULTS: CCT measurements made by US pachymetry and SD-OCT were similar and consistent (r=0.99 for both methods). CCT measurements made by SD-OCT were consistently thinner by approximately 12 micrometers than measurements made by US pachymetry. Repeated measurements of CCT obtained by SD-OCT were more reproducible and had less variability than measurements obtained by US pachymetry. The mean within-subject standard deviation among SDOCT was significantly smaller than that in US pachymetry (1.92 in SD-OCT vs. 2.04 in US pachymetry, p=0.036);
CONCLUSIONS: Measurement of CCT by SD-OCT compares favorably with and is at least as accurate as measurements made by US pachymetry. Repeat measurements of CCT by SD-OCT have less variability than those obtained by US pachymetry, are more reproducible, possibly more reliable, and may better represent actual CCT. American Optometric Association.

Entities:  

Mesh:

Year:  2012        PMID: 23249119

Source DB:  PubMed          Journal:  Optometry        ISSN: 1558-1527


  6 in total

1.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

2.  Comparison of Central Corneal Thickness using Anterior Segment Optical Coherence Tomography Versus Ultrasound Pachymetry.

Authors:  Prasanna Venkatesh Ramesh; Kirti Nath Jha; Krishnagopal Srikanth
Journal:  J Clin Diagn Res       Date:  2017-08-01

3.  In vivo evaluation of the cornea and conjunctiva of the normal laboratory beagle using time- and Fourier-domain optical coherence tomography and ultrasound pachymetry.

Authors:  Ann R Strom; Dennis E Cortés; Carol A Rasmussen; Sara M Thomasy; Kim McIntyre; Shwu-Fei Lee; Philip H Kass; Mark J Mannis; Christopher J Murphy
Journal:  Vet Ophthalmol       Date:  2015-02-10       Impact factor: 1.644

4.  Accuracy of optical coherence tomography (OCT) in pachymetry for glaucoma patients.

Authors:  Marcelo Ayala; Robert Strandås
Journal:  BMC Ophthalmol       Date:  2015-09-29       Impact factor: 2.209

5.  Correlation of the anterior ocular segment biometry with HbA1c level in type 2 diabetes mellitus patients.

Authors:  Abd-Rashid Suraida; Mohtar Ibrahim; Embong Zunaina
Journal:  PLoS One       Date:  2018-01-11       Impact factor: 3.240

6.  Central Corneal Thickness in Spectral-Domain OCT and Associations with Ocular and Systemic Parameters.

Authors:  Alexander Karl-Georg Schuster; Joachim Ernst Fischer; Urs Vossmerbaeumer
Journal:  J Ophthalmol       Date:  2016-06-02       Impact factor: 1.909

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.