Scatter radiation intensities around full-field digital mammography units.

M A Judge1, E Keavey, N Phelan.   

Abstract

The aim of this study was to investigate the scatter radiation intensity around digital mammography systems and apply these data to standard shielding calculations to reveal whether shielding design of existing breast screening rooms is adequate for the use of digital mammography systems. Three digital mammography systems from GE Healthcare, Hologic and Philips were employed in the study. A breast-equivalent phantom was imaged under clinical workload conditions and scatter radiation intensities around the digital mammography systems were measured for a range of angles in three planes using an ionisation chamber. The results were compared with those from previous studies of film-screen systems. It may be deduced from the results that scattering in the backward direction is significant for all three systems, while scattering in the forward direction can be significant for some planes around the GE and Hologic systems. Measurements at typical clinical settings on each system revealed the Philips system to have markedly lower scatter radiation intensities than the other systems. Substituting the measured scattered radiation intensity into shielding calculations yielded barrier requirements similar to those already in place at the screening centres operating these systems. Current radiation protection requirements based on film-screen technology remain sufficient when applied to rooms with digital mammography installations and no alteration is required to the structural shielding.

Mesh:

Year:  2013        PMID: 23239693      PMCID: PMC3615403          DOI: 10.1259/bjr.20120130

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  9 in total

1.  Comprehensive dose survey of breast screening in Ireland.

Authors:  P Baldelli; J McCullagh; N Phelan; F Flanagan
Journal:  Radiat Prot Dosimetry       Date:  2010-11-18       Impact factor: 0.972

2.  Clinical dose performance of full field digital mammography in a breast screening programme.

Authors:  J B McCullagh; P Baldelli; N Phelan
Journal:  Br J Radiol       Date:  2011-05-17       Impact factor: 3.039

3.  Transmission data for shielding diagnostic x-ray facilities.

Authors:  D J Simpkin
Journal:  Health Phys       Date:  1995-05       Impact factor: 1.316

4.  Scatter radiation in mammography.

Authors:  R J Simeoni; D L Thiele
Journal:  Australas Phys Eng Sci Med       Date:  1993-03       Impact factor: 1.430

5.  Technical note: the measurement of mammographic room protection.

Authors:  C H Johnson; A Walker
Journal:  Br J Radiol       Date:  1994-05       Impact factor: 3.039

6.  Scatter radiation intensities about mammography units.

Authors:  D J Simpkin
Journal:  Health Phys       Date:  1996-02       Impact factor: 1.316

7.  Diagnostic X-ray shielding design based on an empirical model of photon attenuation.

Authors:  B R Archer; J I Thornby; S C Bushong
Journal:  Health Phys       Date:  1983-05       Impact factor: 1.316

8.  Measurements of diagnostic x-ray backscatter by a novel ion chamber method.

Authors:  L Stanton; S D Brattelli; J L Day; R E Stanton; T Villafana
Journal:  Med Phys       Date:  1982 Jan-Feb       Impact factor: 4.071

9.  Comparison of digital mammography and screen-film mammography in breast cancer screening: a review in the Irish breast screening program.

Authors:  Niamh M Hambly; Michelle M McNicholas; Niall Phelan; Gormlaith C Hargaden; Ann O'Doherty; Fidelma L Flanagan
Journal:  AJR Am J Roentgenol       Date:  2009-10       Impact factor: 3.959

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.