A phantom-based evaluation of three commercially available patient organ shields for computed tomography X-ray examinations in diagnostic radiology.

Jamie Huggett1, William Mukonoweshuro, Robert Loader.   

Abstract

Three commercially available in-plane patient organ shields (barium eye, bismuth eye and bismuth breast) for computed tomography (CT) examinations were evaluated to determine their effectiveness for dose reduction. Absorbed doses were measured using metal oxide semiconductor field effect transistor dosemeters fastened to a Kyoto CT Torso phantom. Resultant images were visually compared with those minus shielding by an experienced radiologist. Approximate dose reductions of 21, 38 and 50 % were achieved by the barium eye, bismuth eye and bismuth breast shields, respectively, at a cost of increased image noise and streak artefacts. Shielded images produced varied levels of image artefact, particularly those resulting from the eye shields. Measured dose reductions were not consistent with the potential dose savings stated by the manufacturers of the shields. When evaluating the breast shield, similar dose reduction was achieved without shield-induced artefact by simply reducing the X-ray tube current.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23222552     DOI: 10.1093/rpd/ncs327

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  6 in total

1.  Radiation dose reduction using a neck detection algorithm for single spiral brain and cervical spine CT acquisition in the trauma setting.

Authors:  Nicholas D Ardley; Ken K Lau; Kevin Buchan
Journal:  Emerg Radiol       Date:  2013-07-20

2.  Sample size requirements for estimating effective dose from computed tomography using solid-state metal-oxide-semiconductor field-effect transistor dosimetry.

Authors:  Sigal Trattner; Bin Cheng; Radoslaw L Pieniazek; Udo Hoffmann; Pamela S Douglas; Andrew J Einstein
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

3.  Calibration and error analysis of metal-oxide-semiconductor field-effect transistor dosimeters for computed tomography radiation dosimetry.

Authors:  Sigal Trattner; Peter Prinsen; Jens Wiegert; Elazar-Lars Gerland; Efrat Shefer; Tom Morton; Carla M Thompson; Yoad Yagil; Bin Cheng; Sachin Jambawalikar; Rani Al-Senan; Maxwell Amurao; Sandra S Halliburton; Andrew J Einstein
Journal:  Med Phys       Date:  2017-10-26       Impact factor: 4.071

4.  Shielding effect of radiation dose reduction fiber during the use of C-arm fluoroscopy: a phantom study.

Authors:  Hyemi Cha; Kisung Lee; Moon Seok Park; Kyoung Min Lee; Kyeyoung Cho; Ki Hyuk Sung
Journal:  J Radiat Res       Date:  2020-09-08       Impact factor: 2.724

5.  An evaluation of organ dose modulation on a GE optima CT660-computed tomography scanner.

Authors:  Matthew T Dixon; Robert J Loader; Gregory C Stevens; Nick P Rowles
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

6.  Bismuth shield affecting CT image quality and radiation dose in adjacent and distant zones relative to shielding surface: A phantom study.

Authors:  Ying-Lan Liao; Nan-Ku Lai; Yeu-Sheng Tyan; Hui-Yu Tsai
Journal:  Biomed J       Date:  2019-11-04       Impact factor: 4.910

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.